Clinical Evaluation of Doppler Blood Pressure Measurement in Continuous-Flow LVAD Patients: Implications for Postoperative Management

Background: Continuous-flow left ventricular assist devices (LVADs) generate non-pulsatile circulation, rendering conventional oscillometric blood pressure measurements unreliable. Accurate monitoring is critical to prevent complications including stroke, pump thrombosis, and aortic regurgitation. Doppler-based measurement is widely used as a non-invasive alternative, yet its accuracy relative to invasive arterial pressure remains insufficiently characterized. Methods: In this prospective single-centre study, 32 adult continuous-flow LVAD patients underwent simultaneous invasive radial artery and Doppler blood pressure measurements twice daily over three consecutive days (192 paired readings; Day 3: n = 27 due to technical recording issues). Pulsatility was assessed by means of peripheral pulse palpation and transthoracic echocardiography. Spearman’s rho, Wilcoxon signed-rank test, and Bland–Altman analysis were applied. Results: Median invasive MAP was 73.0 [IQR 66–80] mmHg and median Doppler pressure was 75.0 [IQR 70–80] mmHg. Doppler measurements demonstrated strong-to-excellent correlation with invasive MAP across all time points (r = 0.78–0.91, p < 0.001), with no significant paired differences (all p > 0.05). Bland–Altman analysis revealed a bias of −0.35 mmHg with limits of agreement of −9.10 to +8.40 mmHg, within the accepted ±10 mmHg threshold. Correlation with systolic pressure was lower (r = 0.66–0.89, p < 0.001), with a positive bias of +13.47 mmHg and wide limits of agreement (+1.28 to +25.67 mmHg), indicating clinically unacceptable agreement. Conclusions: Doppler-derived blood pressure may provide a reliable estimate of invasive MAP in continuous-flow LVAD patients, whereas its utility for systolic pressure estimation appears limited. Doppler measurement represents a practical, non-invasive tool for routine MAP monitoring in both inpatient and outpatient settings.

Authors

Institutions

Publication Details

Journal
Journal of Cardiovascular Development and Disease
Published
2026-06-18
DOI
https://doi.org/10.3390/jcdd13060276
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Evaluation of Doppler Blood Pressure Measurement in Continuous-Flow LVAD Patients: Implications for Postoperative Management

Ahmet Daylan, Çağatay Engin, Ümit Kahraman, Mustafa Özbaran et al.
Journal of Cardiovascular Development and Disease
Mechanical Circulatory Support Devices
article

Clinical Evaluation of Doppler Blood Pressure Measurement in Continuous-Flow LVAD Patients: Implications for Postoperative Management

Ahmet Daylan, Çağatay Engin, Ümit Kahraman, Mustafa Özbaran, Serkan Ertugay, Tahir Yağdı, Emrah Oğuz, Ayşen Yaprak Kapkın, Sanem Nalbantgil, Vusali Kasumovi
article en

Abstract

Background: Continuous-flow left ventricular assist devices (LVADs) generate non-pulsatile circulation, rendering conventional oscillometric blood pressure measurements unreliable. Accurate monitoring is critical to prevent complications including stroke, pump thrombosis, and aortic regurgitation. Doppler-based measurement is widely used as a non-invasive alternative, yet its accuracy relative to invasive arterial pressure remains insufficiently characterized. Methods: In this prospective single-centre study, 32 adult continuous-flow LVAD patients underwent simultaneous invasive radial artery and Doppler blood pressure measurements twice daily over three consecutive days (192 paired readings; Day 3: n = 27 due to technical recording issues). Pulsatility was assessed by means of peripheral pulse palpation and transthoracic echocardiography. Spearman’s rho, Wilcoxon signed-rank test, and Bland–Altman analysis were applied. Results: Median invasive MAP was 73.0 [IQR 66–80] mmHg and median Doppler pressure was 75.0 [IQR 70–80] mmHg. Doppler measurements demonstrated strong-to-excellent correlation with invasive MAP across all time points (r = 0.78–0.91, p < 0.001), with no significant paired differences (all p > 0.05). Bland–Altman analysis revealed a bias of −0.35 mmHg with limits of agreement of −9.10 to +8.40 mmHg, within the accepted ±10 mmHg threshold. Correlation with systolic pressure was lower (r = 0.66–0.89, p < 0.001), with a positive bias of +13.47 mmHg and wide limits of agreement (+1.28 to +25.67 mmHg), indicating clinically unacceptable agreement. Conclusions: Doppler-derived blood pressure may provide a reliable estimate of invasive MAP in continuous-flow LVAD patients, whereas its utility for systolic pressure estimation appears limited. Doppler measurement represents a practical, non-invasive tool for routine MAP monitoring in both inpatient and outpatient settings.

Journal of Cardiovascular Development and DiseaseVol. 13(6)
Ege University (TR)
Good health and well-being
Openalex Percentile: Top 11%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.