Validity and Reliability of At-Home Sweat Rate and Sodium Patches

Introduction: Hydration recommendations encourage athletes to know their sweat rate and sweat electrolyte concentrations, and novel devices are now available for at-home testing. This study aims to determine the validity and reliability of an at-home sweat rate and Na + loss device. Methods: Eleven males (78.5 ± 8.7 kg, 55.0 ± 13.7 mL·kg −1 ·min −1 ) and 11 females (62.4 ± 7.3 kg, 37.6 ± 10.3 mL·kg −1 ·min −1 ) completed two matched cycling trials (~60% ) in an environmental chamber set to 30°C and 55% relative humidity while wearing a standard lab-based absorbent sweat patch and an at-home microfluidic sweat patch (Gatorade, Barrington, IL) on the ipsilateral forearm. The lab-based absorbent patch was assessed using standard calculations, and at-home microfluidic patch data were obtained through the associated smartphone application. Spearman ρ correlations were used for reliability between trials. Paired-samples t tests compared the lab-based patch versus the microfluidic patch sweat rate and Na + losses. Results: There were no significant differences in wet bulb globe temperature ( P = 0.41), power output ( P = 0.97), rectal temperature ( P = 0.86), or heart rate ( P = 0.61) between trials. The microfluidic patch failed to provide a sweat rate for 9 of 44 trials (20%). Trial 1 sweat rate was not significantly correlated with Trial 2 for the microfluidic patch (ρ = −0.060; P = 0.84) but was for lab-based sweat rate (ρ = 0.698; P < 0.01). Sweat rate assessed by the microfluidic patch (0.53 ± 0.28 L·h −1 ) was significantly less than lab-based measures (0.96 ± 0.37 L·h −1 ; P < 0.01). The microfluidic patch failure rate for Na + losses was 43%. Na + losses were significantly greater for the lab (1053 ± 203 mg) compared with the microfluidic patch (729 ± 143 mg; P < 0.01). Sweat rate data for the microfluidic patch were within 0.5 L·h −1 for 16/34 trials (47%), and Na + losses were within 200 mg for 36% of trials compared with the lab-based patch. Conclusions: Athletes and sports scientists should take caution when assessing sweat rate and Na + losses using at-home microfluidic patches.

Authors

Institutions

Publication Details

Journal
Translational Journal of the American College of Sports Medicine
Published
2026-06-18
DOI
https://doi.org/10.1249/tjx.0000000000000381
Primary Topic
Thermoregulation and physiological responses
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity and Reliability of At-Home Sweat Rate and Sodium Patches

Whitley C. Atkins, Brendon P. McDermott, Abigail Chopelas
Translational Journal of the American College of Sports Medicine
Thermoregulation and physiological responses
article

Validity and Reliability of At-Home Sweat Rate and Sodium Patches

Whitley C. Atkins, Brendon P. McDermott, Abigail Chopelas
article en

Abstract

Introduction: Hydration recommendations encourage athletes to know their sweat rate and sweat electrolyte concentrations, and novel devices are now available for at-home testing. This study aims to determine the validity and reliability of an at-home sweat rate and Na + loss device. Methods: Eleven males (78.5 ± 8.7 kg, 55.0 ± 13.7 mL·kg −1 ·min −1 ) and 11 females (62.4 ± 7.3 kg, 37.6 ± 10.3 mL·kg −1 ·min −1 ) completed two matched cycling trials (~60% ) in an environmental chamber set to 30°C and 55% relative humidity while wearing a standard lab-based absorbent sweat patch and an at-home microfluidic sweat patch (Gatorade, Barrington, IL) on the ipsilateral forearm. The lab-based absorbent patch was assessed using standard calculations, and at-home microfluidic patch data were obtained through the associated smartphone application. Spearman ρ correlations were used for reliability between trials. Paired-samples t tests compared the lab-based patch versus the microfluidic patch sweat rate and Na + losses. Results: There were no significant differences in wet bulb globe temperature ( P = 0.41), power output ( P = 0.97), rectal temperature ( P = 0.86), or heart rate ( P = 0.61) between trials. The microfluidic patch failed to provide a sweat rate for 9 of 44 trials (20%). Trial 1 sweat rate was not significantly correlated with Trial 2 for the microfluidic patch (ρ = −0.060; P = 0.84) but was for lab-based sweat rate (ρ = 0.698; P < 0.01). Sweat rate assessed by the microfluidic patch (0.53 ± 0.28 L·h −1 ) was significantly less than lab-based measures (0.96 ± 0.37 L·h −1 ; P < 0.01). The microfluidic patch failure rate for Na + losses was 43%. Na + losses were significantly greater for the lab (1053 ± 203 mg) compared with the microfluidic patch (729 ± 143 mg; P < 0.01). Sweat rate data for the microfluidic patch were within 0.5 L·h −1 for 16/34 trials (47%), and Na + losses were within 200 mg for 36% of trials compared with the lab-based patch. Conclusions: Athletes and sports scientists should take caution when assessing sweat rate and Na + losses using at-home microfluidic patches.

Translational Journal of the American College of Sports MedicineVol. 11(3)
New Mexico Department of Health (US), University of New Mexico (US), University of Arkansas at Fayetteville (US)
Openalex Percentile: Top 7%
Thermoregulation and physiological responses
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.