Comparison of TFL and high-power Ho: YAG lasers with multiple pulse modulations: an in vitro study

Abstract Purpose To compare the stone ablation volume (AV) achieved by thulium fiber laser (TFL) and high-power Holmium: YAG (Ho: YAG) lasers with MOSES™, Virtual Basket™ (VB™), Vapor Tunnel™ (VT™), and Magneto pulse-modulation technologies in an artificial stone model. Methods BegoStone phantoms (15:6 “powder-to-water” ratio) were used. A motorized arm applied the laser at a constant speed (0.7 mm/s). Laser settings included 1.5 J × 5 Hz, 1 J × 20 Hz, and 0.3 J × 50 Hz, with five 21-mm cuts per setting. AV was calculated from width and depth measurements using an optical microscope. ANOVA was performed. Results Among 750 measurements, TFL-200 μm achieved the highest AV in most settings. At 1.5 J x 5 Hz, TFL-200 μm (15.9 mm³) significantly outperformed MOSES™ (LP: 11 mm³, SP: 10.4 mm³, MD: 9.7 mm³) and Magneto (9.1 mm³) ( p < 0.05). At 1 J × 20 Hz, TFL-200 μm (21.2 mm³) exceeded all technologies ( p < 0.001). At 0.3 J x 50 Hz, TFL-200 μm (8.9 mm³) outperformed MOSES™ MD (3.8 mm³) and Magneto (4.9 mm³), p < 0.05. TFL-150 μm showed comparable AV to Ho: YAG lasers across settings, except at 1 J x 20 Hz, where it was lower than TFL-200 μm and CyberHo150 SP and VT™ ( p < 0.05), primarily due to lower depth of fissure (DOF) while width of fissure (WOF) remained comparable. At 0.3 J × 50 Hz, TFL-150 μm (6.7 mm³) demonstrated lower AV than CyberHo150 SP (10.9 mm³; p = 0.003), but higher AV than MOSES™ MD (3.8 mm³; p = 0.034). Conclusions TFL-200 μm achieved higher AV than Ho: YAG in most settings, but statistical significance was observed in one-third of comparisons. TFL-150 μm demonstrated similar AV to Ho: YAG, except in cases where Ho: YAG had higher DOF. TFL-150 μm has known advantages in terms of irrigation, visibility, ureteroscope tip deflection, and potential for miniaturization of flexible ureteroscopy devices and aspiration methods, supporting its potential applicability for dusting strategies while maintaining comparable AV.

Authors

Institutions

Publication Details

Journal
World Journal of Urology
Published
2026-06-18
DOI
https://doi.org/10.1007/s00345-026-06514-x
Primary Topic
Kidney Stones and Urolithiasis Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of TFL and high-power Ho: YAG lasers with multiple pulse modulations: an in vitro study

Oriol Angerri, Damiano Stracci, Pietro Diana, Paola Arena et al.
World Journal of Urology
Kidney Stones and Urolithiasis Treatments
article

Comparison of TFL and high-power Ho: YAG lasers with multiple pulse modulations: an in vitro study

Oriol Angerri, Damiano Stracci, Pietro Diana, Paola Arena, Esteban Emiliani, Alejandra Bravo‐Balado, A. Sánchez-Puy, Alberto Piana, Sílvia Gràcia Garcia, Cristina Esquinas, Paula Izquierdo, Félix Millán, A. Kanashiro, Joan Palou, Olivier Traxer, Francisco Sánchez-Martín
article en

Abstract

Abstract Purpose To compare the stone ablation volume (AV) achieved by thulium fiber laser (TFL) and high-power Holmium: YAG (Ho: YAG) lasers with MOSES™, Virtual Basket™ (VB™), Vapor Tunnel™ (VT™), and Magneto pulse-modulation technologies in an artificial stone model. Methods BegoStone phantoms (15:6 “powder-to-water” ratio) were used. A motorized arm applied the laser at a constant speed (0.7 mm/s). Laser settings included 1.5 J × 5 Hz, 1 J × 20 Hz, and 0.3 J × 50 Hz, with five 21-mm cuts per setting. AV was calculated from width and depth measurements using an optical microscope. ANOVA was performed. Results Among 750 measurements, TFL-200 μm achieved the highest AV in most settings. At 1.5 J x 5 Hz, TFL-200 μm (15.9 mm³) significantly outperformed MOSES™ (LP: 11 mm³, SP: 10.4 mm³, MD: 9.7 mm³) and Magneto (9.1 mm³) ( p < 0.05). At 1 J × 20 Hz, TFL-200 μm (21.2 mm³) exceeded all technologies ( p < 0.001). At 0.3 J x 50 Hz, TFL-200 μm (8.9 mm³) outperformed MOSES™ MD (3.8 mm³) and Magneto (4.9 mm³), p < 0.05. TFL-150 μm showed comparable AV to Ho: YAG lasers across settings, except at 1 J x 20 Hz, where it was lower than TFL-200 μm and CyberHo150 SP and VT™ ( p < 0.05), primarily due to lower depth of fissure (DOF) while width of fissure (WOF) remained comparable. At 0.3 J × 50 Hz, TFL-150 μm (6.7 mm³) demonstrated lower AV than CyberHo150 SP (10.9 mm³; p = 0.003), but higher AV than MOSES™ MD (3.8 mm³; p = 0.034). Conclusions TFL-200 μm achieved higher AV than Ho: YAG in most settings, but statistical significance was observed in one-third of comparisons. TFL-150 μm demonstrated similar AV to Ho: YAG, except in cases where Ho: YAG had higher DOF. TFL-150 μm has known advantages in terms of irrigation, visibility, ureteroscope tip deflection, and potential for miniaturization of flexible ureteroscopy devices and aspiration methods, supporting its potential applicability for dusting strategies while maintaining comparable AV.

World Journal of UrologyVol. 44(1)
Universitat Autònoma de Barcelona (ES), ParisTech (FR), Sorbonne Université (FR), NYU Langone Health (US), HESAM Université (FR), Puigvert Foundation (ES), Hôpital Tenon (FR), Analyse, Recherche, Développement et Evaluation en Endourologie et Lithiase Urinaire (FR), Universitat de Barcelona (ES)
Openalex Percentile: Top 7%
Kidney Stones and Urolithiasis Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.