Is statistical evaluation sufficient? New external quality assessment performance metrics through clinical performance

Abstract Objectives External Quality Assessment (EQA) programmes conventionally use the Standard Deviation Index (SDI) with an analyte-independent acceptance criterion of −2<SDI<+2, which does not account for biological variation or clinical relevance. We developed two novel metrics, the biologically calibrated tolerance index (SDI RCV ) and percentage SDI (pSDI), that integrate biological variation into EQA interpretation and quantified the sources of analytical imprecision through variance component analysis (VCA). Methods Monthly EQA results from 12 consecutive survey cycles were analysed for 45 analytes (26 biochemistry, 19 immunoassay) in a single clinical laboratory. SDIRCV was defined as the ratio of the reference change value (RCV) to the peer-group coefficient of variation. The pSDI expressed EQA bias as a percentage of the RCV. A five-zone concordance framework was constructed by cross-tabulating conventional SDI and pSDI criteria. A three-level VCA decomposed total imprecision into within-month repeatability, between-month variation, and reagent lot-to-lot variation. Results Of 540 observations, nine (1.67 %) met the Biologically Significant, Statistically Undetected Deviation criterion (−2<SDI<+2 yet pSDI≥+100 % or ≤−100 %), indicating clinically significant bias despite acceptable SDI. Four observations (0.74 %) showed Clinically Inconsequential SDI Exceedance (SDI≥+3 or ≤−3 yet −100 %<pSDI<+100 %). Eleven analytes (24 %) had median SDI RCV <2.0, defining a persistent vulnerability zone. Lot-to-lot variation accounted for a median of 23.6 % of the total variance, exceeding 50 % for six analytes. Conclusions SDI RCV and pSDI detect both false acceptance and false rejection inherent in conventional EQA criteria, offering a biologically grounded framework for aligning laboratory quality assessment with clinical needs.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-06-18
DOI
https://doi.org/10.1515/cclm-2026-0702
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Is statistical evaluation sufficient? New external quality assessment performance metrics through clinical performance

Hamit Hakan, Halil İbrahim Akbay, Elvar Theodorsson
Clinical Chemistry and Laboratory Medicine (CCLM)
Clinical Laboratory Practices and Quality Control
article

Is statistical evaluation sufficient? New external quality assessment performance metrics through clinical performance

Hamit Hakan, Halil İbrahim Akbay, Elvar Theodorsson
article en

Abstract

Abstract Objectives External Quality Assessment (EQA) programmes conventionally use the Standard Deviation Index (SDI) with an analyte-independent acceptance criterion of −2<SDI<+2, which does not account for biological variation or clinical relevance. We developed two novel metrics, the biologically calibrated tolerance index (SDI RCV ) and percentage SDI (pSDI), that integrate biological variation into EQA interpretation and quantified the sources of analytical imprecision through variance component analysis (VCA). Methods Monthly EQA results from 12 consecutive survey cycles were analysed for 45 analytes (26 biochemistry, 19 immunoassay) in a single clinical laboratory. SDIRCV was defined as the ratio of the reference change value (RCV) to the peer-group coefficient of variation. The pSDI expressed EQA bias as a percentage of the RCV. A five-zone concordance framework was constructed by cross-tabulating conventional SDI and pSDI criteria. A three-level VCA decomposed total imprecision into within-month repeatability, between-month variation, and reagent lot-to-lot variation. Results Of 540 observations, nine (1.67 %) met the Biologically Significant, Statistically Undetected Deviation criterion (−2<SDI<+2 yet pSDI≥+100 % or ≤−100 %), indicating clinically significant bias despite acceptable SDI. Four observations (0.74 %) showed Clinically Inconsequential SDI Exceedance (SDI≥+3 or ≤−3 yet −100 %<pSDI<+100 %). Eleven analytes (24 %) had median SDI RCV <2.0, defining a persistent vulnerability zone. Lot-to-lot variation accounted for a median of 23.6 % of the total variance, exceeding 50 % for six analytes. Conclusions SDI RCV and pSDI detect both false acceptance and false rejection inherent in conventional EQA criteria, offering a biologically grounded framework for aligning laboratory quality assessment with clinical needs.

Clinical Chemistry and Laboratory Medicine (CCLM)
Linköping University (SE), Van Yüzüncü Yıl Üniversitesi (TR)
Openalex Percentile: Top 7%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.