AeroCoil: Pneumatically Controlled Dual‐Tunable RF Coil for Proton ( <sup>1</sup> H) and Deuterium ( <sup>2</sup> H)

ABSTRACT In specific MRI applications, RF coils must operate at multiple resonance frequencies to support multinuclear imaging and spectroscopy. Designing such coils that can either resonate at or switch between multiple frequencies presents considerable technical challenges. Traditional multituned coils typically suffer from strong electromagnetic (EM) coupling due to limited physical space and the interaction between multiple resonating structures. Frequency‐switchable coils offer an alternative by adjusting circuit capacitance or inductance to tune for different nuclei, typically relying on PIN diodes or similar electronic switches. However, these designs require frequency‐sensitive bias circuitry, dedicated drivers, and long DC control lines, all of which are prone to RF interference within the MRI environment. Furthermore, PIN diodes introduce inherent RF leakage and nonlinearities that can significantly degrade image quality. This study introduces a pneumatically actuated frequency‐switchable RF coil, termed AeroCoil, that enables resonance switching without electrical biasing. The AeroCoil design utilizes air pressure to mechanically adjust capacitance, enabling transitions between the 1 H and 2 H Larmor frequencies at 9.4 T. This pneumatic mechanism eliminates the need for local electrical components near the imaging volume, thereby reducing RF noise, eliminating EM interference, and simplifying the coil's circuit architecture. The AeroCoil's performance was benchmarked against single‐tuned coils, representing optimal performance at each frequency, and a PIN diode‐based switchable coil. Bench measurements evaluated return losses and quality factors, whereas MRI experiments assessed SNR and image artifacts. The AeroCoil retained 88.2%–91.9% of the SNR compared with a single‐tuned 1 H coil, whereas the PIN diode‐based 1 H coil achieved only 68.3%–75%. At the 2 H frequency, AeroCoil performance was further emphasized, retaining 71.7%–74.6% SNR, whereas the PIN diode‐based design dropped sharply to 27.1%–28.8%.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2026-06-18
DOI
https://doi.org/10.1002/nbm.70337
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

AeroCoil: Pneumatically Controlled Dual‐Tunable RF Coil for Proton ( 1 H) and Deuterium ( 2 H)

Sung‐Min Sohn, Vikram D. Kodibagkar, Chavalchart Herabut, Samantha McClendon et al.
NMR in Biomedicine
Advanced MRI Techniques and Applications
article

AeroCoil: Pneumatically Controlled Dual‐Tunable RF Coil for Proton ( 1 H) and Deuterium ( 2 H)

Sung‐Min Sohn, Vikram D. Kodibagkar, Chavalchart Herabut, Samantha McClendon, Dakota Ortega, Scott Beeman, Bryan Rangel Valle
article en

Abstract

ABSTRACT In specific MRI applications, RF coils must operate at multiple resonance frequencies to support multinuclear imaging and spectroscopy. Designing such coils that can either resonate at or switch between multiple frequencies presents considerable technical challenges. Traditional multituned coils typically suffer from strong electromagnetic (EM) coupling due to limited physical space and the interaction between multiple resonating structures. Frequency‐switchable coils offer an alternative by adjusting circuit capacitance or inductance to tune for different nuclei, typically relying on PIN diodes or similar electronic switches. However, these designs require frequency‐sensitive bias circuitry, dedicated drivers, and long DC control lines, all of which are prone to RF interference within the MRI environment. Furthermore, PIN diodes introduce inherent RF leakage and nonlinearities that can significantly degrade image quality. This study introduces a pneumatically actuated frequency‐switchable RF coil, termed AeroCoil, that enables resonance switching without electrical biasing. The AeroCoil design utilizes air pressure to mechanically adjust capacitance, enabling transitions between the 1 H and 2 H Larmor frequencies at 9.4 T. This pneumatic mechanism eliminates the need for local electrical components near the imaging volume, thereby reducing RF noise, eliminating EM interference, and simplifying the coil's circuit architecture. The AeroCoil's performance was benchmarked against single‐tuned coils, representing optimal performance at each frequency, and a PIN diode‐based switchable coil. Bench measurements evaluated return losses and quality factors, whereas MRI experiments assessed SNR and image artifacts. The AeroCoil retained 88.2%–91.9% of the SNR compared with a single‐tuned 1 H coil, whereas the PIN diode‐based 1 H coil achieved only 68.3%–75%. At the 2 H frequency, AeroCoil performance was further emphasized, retaining 71.7%–74.6% SNR, whereas the PIN diode‐based design dropped sharply to 27.1%–28.8%.

NMR in BiomedicineVol. 39(8)
Arizona State University (US)
Affordable and clean energy
Openalex Percentile: Top 6%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.