Retrospective assessment of low-dose and low-contrast agent protocols for photon-counting CT angiography of the aorta

Our objective was to assess dedicated low-dose (LD) and low-contrast (LC) scan protocols for aortic angiography, compared to standard protocols in photon-counting CT (PCD-CT) and energy-integrating CT (EID-CT). 57 patients (23 women) who underwent aortic PCD-CT angiography (Nov. 23 - July 24) were assessed retrospectively. 19 received a low-contrast protocol (LC; 14 mg iodine, image quality level IQ64), 18 a low-dose protocol (LD; 21 mg iodine, IQ42), and 20 a combined low-dose/low-contrast protocol (LD/LC) according to defined patient characteristics. In addition, two groups of 20 BMI-matched patients were created, one of which underwent CTA using the standard PCD-CT protocol and the other using EID-CT, and these were compared with the new protocols. CNR was calculated. Overall quality, luminal contrast, vessel sharpness, and artifacts were rated by four radiologists. LD enabled ~ 40% dose-reduction compared to EID-CT (3.3 ± 0.8 versus 5.3 ± 0.4mGy; p < 0.01) but higher CNR (33.5 ± 6.1 vs. 25.1 ± 6.4; p < 0.01). CNR of LC (26.3 ± 5.6) and LD/LC (25.6 ± 9.1) did not differ significantly to EID-CT (25.1 ± 6.4; p > 0.99). Image quality of LC was deemed superior to EID-CT (5 [5-5] vs. 4 [4-4]; p < 0.01). LD (4 [4-5]) and LD/LC (4 [4-4]) demonstrated comparable image quality to EID-CT (p = 0.52/0.94). In conclusion even the LD/LC protocol provided image quality comparable to EID-CT despite substantial reduction of radiation and contrast agent dose. Special dose-reduced and contrast-reduced protocols are feasible without compromising diagnostic image quality.

Authors

Institutions

Publication Details

Journal
The International Journal of Cardiovascular Imaging
Published
2026-06-19
DOI
https://doi.org/10.1007/s10554-026-03757-y
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Retrospective assessment of low-dose and low-contrast agent protocols for photon-counting CT angiography of the aorta

Pauline Pannenbecker, Henner Huflage, Philipp Gruschwitz, Jan‐Peter Grunz et al.
The International Journal of Cardiovascular Imaging
Advanced X-ray and CT Imaging
article

Retrospective assessment of low-dose and low-contrast agent protocols for photon-counting CT angiography of the aorta

Pauline Pannenbecker, Henner Huflage, Philipp Gruschwitz, Jan‐Peter Grunz, Thorsten Alexander Bley, Jan-Lucca Hennes, Viktor Hartung, Kristina Krompaß, Ramin Saam Dazeh
article en

Abstract

Our objective was to assess dedicated low-dose (LD) and low-contrast (LC) scan protocols for aortic angiography, compared to standard protocols in photon-counting CT (PCD-CT) and energy-integrating CT (EID-CT). 57 patients (23 women) who underwent aortic PCD-CT angiography (Nov. 23 - July 24) were assessed retrospectively. 19 received a low-contrast protocol (LC; 14 mg iodine, image quality level IQ64), 18 a low-dose protocol (LD; 21 mg iodine, IQ42), and 20 a combined low-dose/low-contrast protocol (LD/LC) according to defined patient characteristics. In addition, two groups of 20 BMI-matched patients were created, one of which underwent CTA using the standard PCD-CT protocol and the other using EID-CT, and these were compared with the new protocols. CNR was calculated. Overall quality, luminal contrast, vessel sharpness, and artifacts were rated by four radiologists. LD enabled ~ 40% dose-reduction compared to EID-CT (3.3 ± 0.8 versus 5.3 ± 0.4mGy; p < 0.01) but higher CNR (33.5 ± 6.1 vs. 25.1 ± 6.4; p < 0.01). CNR of LC (26.3 ± 5.6) and LD/LC (25.6 ± 9.1) did not differ significantly to EID-CT (25.1 ± 6.4; p > 0.99). Image quality of LC was deemed superior to EID-CT (5 [5-5] vs. 4 [4-4]; p < 0.01). LD (4 [4-5]) and LD/LC (4 [4-4]) demonstrated comparable image quality to EID-CT (p = 0.52/0.94). In conclusion even the LD/LC protocol provided image quality comparable to EID-CT despite substantial reduction of radiation and contrast agent dose. Special dose-reduced and contrast-reduced protocols are feasible without compromising diagnostic image quality.

The International Journal of Cardiovascular Imaging
Universitätsklinikum Würzburg (DE)
Openalex Percentile: Top 12%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.