Subluxation-induced liner adhesion and the effect of impaction to prevent liner dislocation in ceramic hip arthroplasty

Aims: Ceramic-on-ceramic bearings are commonly used in total hip arthroplasty. The long-term clinical results show low revision rates. However, there might be the possibility of liner dislocation during clinical use, due to adhesion forces occurring in the bearing. Methods: In this study, the adhesion forces during subluxation of the ceramic head from the liner were measured. Two different implant systems (Mathys SeleXys and DePuy Synthes Pinnacle) with two different head sizes (28 mm and 36 mm) were studied. Additionally, the liner-to-cup fixation forces with three different liner sizes (44 mm, 52 mm, and 64 mm) were analyzed to correlate the adhesion forces with the liner fixation force. Three different assembly forces (1 kN, 2 kN, and 4 kN) were considered. Results: There was a significant difference in the adhesion forces between Mathys SeleXys and Depuy Synthes Pinnacle bearings at a head size of 28 mm. However, no significant difference was found for the 36 mm head size. The maximum adhesion force reached 265 N. The liner fixation forces varied depending on the assembly force. For an assembly force of 1 kN, the fixation forces varied between 95 and 385 N, which is partly below the corresponding adhesion force. For assembly forces of 2 kN or more, the liner fixation force was always above the adhesion forces. Conclusion: A low liner impaction force can lead to liner dislocation due to possible adhesion forces between the ceramic head and the ceramic liner. Sufficient liner impaction forces of more than 2 kN should be applied to prevent the risk of adhesion-related liner extractions.

Authors

Institutions

Publication Details

Journal
Bone and Joint Research
Published
2026-06-17
DOI
https://doi.org/10.1302/2046-3758.156.bjr-2025-0615.r1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Subluxation-induced liner adhesion and the effect of impaction to prevent liner dislocation in ceramic hip arthroplasty

Jan Philippe Kretzer, Mareike Schonhoff, Maximilian Uhler, Sebastian Jaeger et al.
Bone and Joint Research
Orthopaedic implants and arthroplasty
article

Subluxation-induced liner adhesion and the effect of impaction to prevent liner dislocation in ceramic hip arthroplasty

Jan Philippe Kretzer, Mareike Schonhoff, Maximilian Uhler, Sebastian Jaeger, Therese Bormann
article en

Abstract

Aims: Ceramic-on-ceramic bearings are commonly used in total hip arthroplasty. The long-term clinical results show low revision rates. However, there might be the possibility of liner dislocation during clinical use, due to adhesion forces occurring in the bearing. Methods: In this study, the adhesion forces during subluxation of the ceramic head from the liner were measured. Two different implant systems (Mathys SeleXys and DePuy Synthes Pinnacle) with two different head sizes (28 mm and 36 mm) were studied. Additionally, the liner-to-cup fixation forces with three different liner sizes (44 mm, 52 mm, and 64 mm) were analyzed to correlate the adhesion forces with the liner fixation force. Three different assembly forces (1 kN, 2 kN, and 4 kN) were considered. Results: There was a significant difference in the adhesion forces between Mathys SeleXys and Depuy Synthes Pinnacle bearings at a head size of 28 mm. However, no significant difference was found for the 36 mm head size. The maximum adhesion force reached 265 N. The liner fixation forces varied depending on the assembly force. For an assembly force of 1 kN, the fixation forces varied between 95 and 385 N, which is partly below the corresponding adhesion force. For assembly forces of 2 kN or more, the liner fixation force was always above the adhesion forces. Conclusion: A low liner impaction force can lead to liner dislocation due to possible adhesion forces between the ceramic head and the ceramic liner. Sufficient liner impaction forces of more than 2 kN should be applied to prevent the risk of adhesion-related liner extractions.

Bone and Joint ResearchVol. 15(6)
Heidelberg University (DE), University Hospital Heidelberg (DE)
Openalex Percentile: Top 6%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.