Variability of Scheimpflug Densitometry Measurements in Clear and Opaque Corneas Under Different Lighting Settings

PURPOSE: To evaluate whether variability in ambient light levels affects Scheimpflug corneal optical densitometry measurements in clear and opaque corneas. METHODS: Eyes with clear corneas (N = 60) or infectious corneal opacity (N = 49) underwent Scheimpflug densitometry (Pentacam, Oculus) under 3 ambient lighting conditions: dark room (1.8 ± 0.9 lux), mild ambient light (7.9 ± 0.9 lux), and bright light (>1000 lux). Densitometry measurements in grayscale units (GSUs) were obtained for the anterior, middle, posterior, and entire cornea and concentric corneal zones. The Wilcoxon signed-rank test and generalized estimating equation models were used to compare measurements across lighting settings. RESULTS: Under bright light, densitometry measurements could not be obtained in 90% of clear corneas and 92% of opaque corneas because of machine error, while measurements that were obtained were significantly elevated (P < 0.001). Comparing dark room versus mild ambient light settings, clear corneas showed significantly higher readings under mild ambient light for anterior (1.0 GSUs, P = 0.004), middle (1.5 GSUs, P < 0.001), and posterior (1.5 GSUs, P < 0.001) layers, all concentric corneal zones, and the entire cornea (all P < 0.05). Among opaque corneas, all regions except the anterior and far peripheral cornea (P > 0.05) showed significant differences (1.3-2.4 GSUs, all P < 0.05). Overall, 88.3% of clear corneas and 73.5% of opaque corneas demonstrated higher readings under mild ambient light compared with dark room conditions. CONCLUSIONS: Scheimpflug densitometry should not be performed under regular room lighting because of frequent errors and measurement overestimation. Dark room conditions are recommended for optimal measurement, particularly when quantifying subtle changes in opacity.

Authors

Institutions

Publication Details

Journal
Cornea
Published
2026-06-16
DOI
https://doi.org/10.1097/ico.0000000000004222
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Variability of Scheimpflug Densitometry Measurements in Clear and Opaque Corneas Under Different Lighting Settings

Subeesh Kuyyadiyil, Kunal S. Parikh, Gautam S. Parmar, Nakul Singh Shekhawat et al.
Cornea
Corneal surgery and disorders
article

Variability of Scheimpflug Densitometry Measurements in Clear and Opaque Corneas Under Different Lighting Settings

Subeesh Kuyyadiyil, Kunal S. Parikh, Gautam S. Parmar, Nakul Singh Shekhawat, Elesh Jain, Kamini N. Reddy, Jordan M. Shuff
article en

Abstract

PURPOSE: To evaluate whether variability in ambient light levels affects Scheimpflug corneal optical densitometry measurements in clear and opaque corneas. METHODS: Eyes with clear corneas (N = 60) or infectious corneal opacity (N = 49) underwent Scheimpflug densitometry (Pentacam, Oculus) under 3 ambient lighting conditions: dark room (1.8 ± 0.9 lux), mild ambient light (7.9 ± 0.9 lux), and bright light (>1000 lux). Densitometry measurements in grayscale units (GSUs) were obtained for the anterior, middle, posterior, and entire cornea and concentric corneal zones. The Wilcoxon signed-rank test and generalized estimating equation models were used to compare measurements across lighting settings. RESULTS: Under bright light, densitometry measurements could not be obtained in 90% of clear corneas and 92% of opaque corneas because of machine error, while measurements that were obtained were significantly elevated (P < 0.001). Comparing dark room versus mild ambient light settings, clear corneas showed significantly higher readings under mild ambient light for anterior (1.0 GSUs, P = 0.004), middle (1.5 GSUs, P < 0.001), and posterior (1.5 GSUs, P < 0.001) layers, all concentric corneal zones, and the entire cornea (all P < 0.05). Among opaque corneas, all regions except the anterior and far peripheral cornea (P > 0.05) showed significant differences (1.3-2.4 GSUs, all P < 0.05). Overall, 88.3% of clear corneas and 73.5% of opaque corneas demonstrated higher readings under mild ambient light compared with dark room conditions. CONCLUSIONS: Scheimpflug densitometry should not be performed under regular room lighting because of frequent errors and measurement overestimation. Dark room conditions are recommended for optimal measurement, particularly when quantifying subtle changes in opacity.

Cornea
Johns Hopkins University (US), Johns Hopkins Medicine (US), Mahatma Gandhi Chitrakoot Gramodaya Vishwavidyalaya (IN)
National Eye Institute
Openalex Percentile: Top 7%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.