Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard flowmeters under hyperbaric conditions: a technical report

Introduction: During clinical use of extracorporeal membrane oxygenation (ECMO) in hyperbaric conditions at our centre, upward titration of indicated sweep gas flow rates is required to maintain adequate CO₂ clearance. This project measured the impact of hyperbaric pressure on oxygen flow rates delivered by the Comweld Ezi-Flow flowmeters used in our centre. Methods: Oxygen flow rates through Comweld Ezi-flow standard and low oxygen gas flowmeters were set at 101.3 kPa (1 atmosphere absolute [atm abs]) and then measured at intervals up to 284 kPa (2.8 atm abs) using a calibrated gas flow analyser, with cross verification against a small Douglas bag test type apparatus. During testing, the chamber was compressed and decompressed at a rate of 10 kPa·min⁻¹. Flow rates during chamber compression and decompression were compared. Results: The indicated rate of oxygen gas flow through the unadjusted flowmeters changed minimally - typically rising by a maximum of half of the diameter of the indicator ball. The actual (volumetric) flow, tested across indicated flow rates from 3 to 12 L·min⁻¹, was consistently reduced by approximately 50% as the chamber pressure increased from 101.3 to 284 kPa (1 to 2.8 atm abs). A slightly smaller reduction was observed when assessing the low flowmeter across the same pressure range; reductions of 40.0 and 43.3% were demonstrated at 0.3 to 0.6 L·min⁻¹ respectively. Differences in flow rates between compression and decompression were minor except at the very lowest flows. Conclusions: At 284 kPa (2.8 atm abs), actual volumetric flow of oxygen through Comweld Ezi-Flow flowmeters is dramatically reduced and this needs appropriate compensation to ensure therapeutic aims are achieved.

Authors

Institutions

Publication Details

Journal
Diving and Hyperbaric Medicine Journal
Published
2026-06-17
DOI
https://doi.org/10.28920/dhm56.2.154-160
Primary Topic
Intravenous Infusion Technology and Safety
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard flowmeters under hyperbaric conditions: a technical report

Theo Tsouras, Yoav Aufgang, Jason Chan, Bridget Devaney et al.
Diving and Hyperbaric Medicine Journal
Intravenous Infusion Technology and Safety
article

Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard flowmeters under hyperbaric conditions: a technical report

Theo Tsouras, Yoav Aufgang, Jason Chan, Bridget Devaney, Ian Millar
article en

Abstract

Introduction: During clinical use of extracorporeal membrane oxygenation (ECMO) in hyperbaric conditions at our centre, upward titration of indicated sweep gas flow rates is required to maintain adequate CO₂ clearance. This project measured the impact of hyperbaric pressure on oxygen flow rates delivered by the Comweld Ezi-Flow flowmeters used in our centre. Methods: Oxygen flow rates through Comweld Ezi-flow standard and low oxygen gas flowmeters were set at 101.3 kPa (1 atmosphere absolute [atm abs]) and then measured at intervals up to 284 kPa (2.8 atm abs) using a calibrated gas flow analyser, with cross verification against a small Douglas bag test type apparatus. During testing, the chamber was compressed and decompressed at a rate of 10 kPa·min⁻¹. Flow rates during chamber compression and decompression were compared. Results: The indicated rate of oxygen gas flow through the unadjusted flowmeters changed minimally - typically rising by a maximum of half of the diameter of the indicator ball. The actual (volumetric) flow, tested across indicated flow rates from 3 to 12 L·min⁻¹, was consistently reduced by approximately 50% as the chamber pressure increased from 101.3 to 284 kPa (1 to 2.8 atm abs). A slightly smaller reduction was observed when assessing the low flowmeter across the same pressure range; reductions of 40.0 and 43.3% were demonstrated at 0.3 to 0.6 L·min⁻¹ respectively. Differences in flow rates between compression and decompression were minor except at the very lowest flows. Conclusions: At 284 kPa (2.8 atm abs), actual volumetric flow of oxygen through Comweld Ezi-Flow flowmeters is dramatically reduced and this needs appropriate compensation to ensure therapeutic aims are achieved.

Diving and Hyperbaric Medicine JournalVol. 56(2)
Alfred Health (AU), Monash University (AU)
Openalex Percentile: Top 14%
Intravenous Infusion Technology and Safety
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.