Minimising pacemaker exposure in thoracic radiotherapy: a comparative study of IMRT, VMAT, and helical tomotherapy

Abstract Purpose Patients with cardiac implantable electronic devices (CIEDs) increasingly require thoracic radiotherapy, but comparative evidence guiding technique selection remains limited. This study evaluated CIED dose, target coverage, organ-at-risk (OAR) sparing, and delivery efficiency across IMRT, VMAT, and helical tomotherapy in matched thoracic cases. Methods Eighty patients with left-sided subclavicular pacemakers were planned using five techniques: IMRT, VMAT, and tomotherapy with 1 cm, 2.5 cm, and 5 cm field widths. All plans used 6 MV FFF photons with harmonised dose constraints. CIED endpoints included Dmax, D0.03 cc, and Dmean. Secondary endpoints included PTV coverage, homogeneity, conformity, lung and heart dose metrics, monitor units, and beam-on time. Statistical comparisons were performed using repeated-measures ANOVA with Holm-Bonferroni correction.Out-of-field dose was validated using OSLDs in a RANDO phantom. Results Target coverage was comparable across techniques (PTV D95 ≈ 98.5–99.2%). CIED dose was significantly reduced with tomotherapy, with mean Dmax decreasing from 1.86 ± 0.43 Gy (IMRT) and 1.68 ± 0.38 Gy (VMAT)to 1.10 ± 0.25 Gy for the 1 cm field width ( p < 0.01). Mean heart dose decreased from 8.2 Gy to 3.8 Gy, and mean lung dose from 9.4 Gy to 7.6 Gy. VMAT required fewer monitor units (500 vs. 910 for IMRT) and shorter delivery (1–1.5 min), whereas tomotherapy ranged from 140 to 500 s. Conclusion Helical tomotherapy improves CIED sparing and selected OAR metrics while maintaining target coverage. However, VMAT offers superior delivery efficiency. Technique selection should therefore be individualised, balancing dosimetric benefit against intrafraction motion risk, target geometry, and patient-specific clinical priorities.

Authors

Institutions

Publication Details

Journal
Journal of the Egyptian National Cancer Institute
Published
2026-06-13
DOI
https://doi.org/10.1186/s43046-026-00369-4
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Minimising pacemaker exposure in thoracic radiotherapy: a comparative study of IMRT, VMAT, and helical tomotherapy

Munish Gairola, Abhay Kumar Singh, Manindra Bhushan, Sandeep Singh et al.
Journal of the Egyptian National Cancer Institute
Advanced Radiotherapy Techniques
article

Minimising pacemaker exposure in thoracic radiotherapy: a comparative study of IMRT, VMAT, and helical tomotherapy

Munish Gairola, Abhay Kumar Singh, Manindra Bhushan, Sandeep Singh, Raj Pal Singh, Anuj Vijay, Supratik Sen, Benoy Kumar Singh, Dipesh .
article en

Abstract

Abstract Purpose Patients with cardiac implantable electronic devices (CIEDs) increasingly require thoracic radiotherapy, but comparative evidence guiding technique selection remains limited. This study evaluated CIED dose, target coverage, organ-at-risk (OAR) sparing, and delivery efficiency across IMRT, VMAT, and helical tomotherapy in matched thoracic cases. Methods Eighty patients with left-sided subclavicular pacemakers were planned using five techniques: IMRT, VMAT, and tomotherapy with 1 cm, 2.5 cm, and 5 cm field widths. All plans used 6 MV FFF photons with harmonised dose constraints. CIED endpoints included Dmax, D0.03 cc, and Dmean. Secondary endpoints included PTV coverage, homogeneity, conformity, lung and heart dose metrics, monitor units, and beam-on time. Statistical comparisons were performed using repeated-measures ANOVA with Holm-Bonferroni correction.Out-of-field dose was validated using OSLDs in a RANDO phantom. Results Target coverage was comparable across techniques (PTV D95 ≈ 98.5–99.2%). CIED dose was significantly reduced with tomotherapy, with mean Dmax decreasing from 1.86 ± 0.43 Gy (IMRT) and 1.68 ± 0.38 Gy (VMAT)to 1.10 ± 0.25 Gy for the 1 cm field width ( p < 0.01). Mean heart dose decreased from 8.2 Gy to 3.8 Gy, and mean lung dose from 9.4 Gy to 7.6 Gy. VMAT required fewer monitor units (500 vs. 910 for IMRT) and shorter delivery (1–1.5 min), whereas tomotherapy ranged from 140 to 500 s. Conclusion Helical tomotherapy improves CIED sparing and selected OAR metrics while maintaining target coverage. However, VMAT offers superior delivery efficiency. Technique selection should therefore be individualised, balancing dosimetric benefit against intrafraction motion risk, target geometry, and patient-specific clinical priorities.

Journal of the Egyptian National Cancer InstituteVol. 38(1)
Bhubaneswar Borooah Cancer Institute (IN), Rajiv Gandhi Cancer Institute and Research Centre (IN), Cancer Hospital and Research Institute (IN), BLK Super Speciality Hospital (IN), Amrita Vishwa Vidyapeetham (IN), GLA University (IN)
Openalex Percentile: Top 5%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.