A novel robot for CT-guided bone needle insertion with a rotational drilling and force-feedback speed control mechanism: preliminary evaluation in swine

Abstract Purpose To evaluate the feasibility and accuracy of computed tomography (CT)-guided needle insertion into swine bones by using a novel robotic system capable of rotational drilling. Materials and methods This was an animal experiment using three swine. A remote-controlled robot equipped with a rotational drilling and force-feedback insertion speed control mechanism was developed for bone needle insertion. Using the robot, CT-guided insertion of a 10-gauge bone access needle was attempted in the lumbar vertebrae, ilia, and femora six times each. Needle insertion accuracy was evaluated using the angle error, which is defined as the difference between the predetermined and post-insertion needle angles on axial and sagittal CT images. The time required for needle insertion was measured. The angle error and time required for needle insertion were compared among the bones using the Kruskal–Wallis test. Adverse events were also evaluated. Results Robotic bone needle insertion was successful in all attempts. The median axial and sagittal angle errors were 0.21 and 0.21° for the lumbar vertebrae, 0.32 and 0.13° for the ilia, and 0.65 and 0.25° for the femora, respectively. Axial angle errors were significantly different among the bone types ( p = 0.038). The time required for needle insertion was 23.6, 21.3, and 59.7 s for the lumbar vertebrae, ilia, and femora, respectively. Time was significantly different among bone types ( p = 0.017). No adverse events were observed. Conclusion CT-guided bone needle insertion using a robot equipped with a rotational drilling and force-feedback insertion speed control mechanism was feasible and accurate in swine.

Authors

Institutions

Publication Details

Journal
Japanese Journal of Radiology
Published
2026-06-13
DOI
https://doi.org/10.1007/s11604-026-02030-5
Primary Topic
Soft Robotics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel robot for CT-guided bone needle insertion with a rotational drilling and force-feedback speed control mechanism: preliminary evaluation in swine

Takao Hiraki, Yusuke Matsui, Koji Tomita, Yuta Kimura et al.
Japanese Journal of Radiology
Soft Robotics and Applications
article

A novel robot for CT-guided bone needle insertion with a rotational drilling and force-feedback speed control mechanism: preliminary evaluation in swine

Takao Hiraki, Yusuke Matsui, Koji Tomita, Yuta Kimura, Kiyoshi Matsumiya, Kazuaki Munetomo, Takanori Sasaki, Tetsushi Kamegawa, Fumiyo Higaki, Toshihiro Iguchi, Soichiro Okamoto, Noriyuki Umakoshi, Ryutaro Matsuura, Jun Sakùrai, Atsushi Nakazawa, Takayuki Matsuno
article en

Abstract

Abstract Purpose To evaluate the feasibility and accuracy of computed tomography (CT)-guided needle insertion into swine bones by using a novel robotic system capable of rotational drilling. Materials and methods This was an animal experiment using three swine. A remote-controlled robot equipped with a rotational drilling and force-feedback insertion speed control mechanism was developed for bone needle insertion. Using the robot, CT-guided insertion of a 10-gauge bone access needle was attempted in the lumbar vertebrae, ilia, and femora six times each. Needle insertion accuracy was evaluated using the angle error, which is defined as the difference between the predetermined and post-insertion needle angles on axial and sagittal CT images. The time required for needle insertion was measured. The angle error and time required for needle insertion were compared among the bones using the Kruskal–Wallis test. Adverse events were also evaluated. Results Robotic bone needle insertion was successful in all attempts. The median axial and sagittal angle errors were 0.21 and 0.21° for the lumbar vertebrae, 0.32 and 0.13° for the ilia, and 0.65 and 0.25° for the femora, respectively. Axial angle errors were significantly different among the bone types ( p = 0.038). The time required for needle insertion was 23.6, 21.3, and 59.7 s for the lumbar vertebrae, ilia, and femora, respectively. Time was significantly different among bone types ( p = 0.017). No adverse events were observed. Conclusion CT-guided bone needle insertion using a robot equipped with a rotational drilling and force-feedback insertion speed control mechanism was feasible and accurate in swine.

Japanese Journal of Radiology
Okayama University of Science (JP), Okayama University (JP), Okayama University Hospital (JP)
Openalex Percentile: Top 10%
Soft Robotics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.