Effect of Implant Design on Fracture Resistance, Mode, and Crystal Structure of One‐ and Two‐Piece Zirconia Implants With Ceramic or Titanium Abutment Screws: An In Vitro Study

ABSTRACT Objectives To evaluate the fracture resistance, fracture modes, and crystalline phase changes of zirconia one‐piece implants (OPI) and bone‐level (BLTPI) or tissue‐level, two‐piece implants (TLTPI), with ceramic (C) or titanium (T) abutment screws after dynamic preloading in a moist environment. Materials and Methods Dynamic preloading under moist conditions and subsequent static load‐to‐fracture testing were performed to evaluate fracture resistance and fracture mode. Scanning electron microscopy (SEM) analysis of fracture surfaces was employed to supplement the observations of macroscopic fracture modes. Crystalline phase changes were assessed using micro‐Raman spectroscopy. Specimen preparations were performed in accordance with ISO 14801:2016. Five different ceramic (zirconia) implant designs were divided into the following groups: OPI, BLTPI (BLC, BLT), TLTPI (TLC, TLT). One‐way and two‐way ANOVA tests were used for comparisons and factor effect analysis. Results Fracture resistance (N) and bending moment (N cm) differed significantly among the implant groups ( p < 0.001), following the order TLTPI (775.2 ± 71 N) > OPI (605.7 ± 37 N) > BLTPI (498.3 ± 93 N). Within the two‐piece implants (TPI), tissue‐level designs (TL) outperformed bone‐level designs (BL), while designs with different abutment screw materials showed no significant difference ( p = 0.982). Conclusions Within the limitations of this in vitro study, the evaluated zirconia‐based OPI, BLTPI, and TLTPI systems demonstrate sufficient fracture resistance to support further clinical investigation. Implant‐level design (TL vs. BL) influences mechanical performance, whereas the influence of abutment screw material (ceramic vs. titanium) is less pronounced and varies according to the implant‐level design.

Authors

Institutions

Publication Details

Journal
Clinical Oral Implants Research
Published
2026-06-13
DOI
https://doi.org/10.1111/clr.70149
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of Implant Design on Fracture Resistance, Mode, and Crystal Structure of One‐ and Two‐Piece Zirconia Implants With Ceramic or Titanium Abutment Screws: An In Vitro Study

Evaggelia Papia, Jonas P. Becktor, Siamak Eqtesadi, A. Gül et al.
Clinical Oral Implants Research
Dental Implant Techniques and Outcomes
article

Effect of Implant Design on Fracture Resistance, Mode, and Crystal Structure of One‐ and Two‐Piece Zirconia Implants With Ceramic or Titanium Abutment Screws: An In Vitro Study

Evaggelia Papia, Jonas P. Becktor, Siamak Eqtesadi, A. Gül, Per Vult von Steyern, Magnus Anderson
article en

Abstract

ABSTRACT Objectives To evaluate the fracture resistance, fracture modes, and crystalline phase changes of zirconia one‐piece implants (OPI) and bone‐level (BLTPI) or tissue‐level, two‐piece implants (TLTPI), with ceramic (C) or titanium (T) abutment screws after dynamic preloading in a moist environment. Materials and Methods Dynamic preloading under moist conditions and subsequent static load‐to‐fracture testing were performed to evaluate fracture resistance and fracture mode. Scanning electron microscopy (SEM) analysis of fracture surfaces was employed to supplement the observations of macroscopic fracture modes. Crystalline phase changes were assessed using micro‐Raman spectroscopy. Specimen preparations were performed in accordance with ISO 14801:2016. Five different ceramic (zirconia) implant designs were divided into the following groups: OPI, BLTPI (BLC, BLT), TLTPI (TLC, TLT). One‐way and two‐way ANOVA tests were used for comparisons and factor effect analysis. Results Fracture resistance (N) and bending moment (N cm) differed significantly among the implant groups ( p < 0.001), following the order TLTPI (775.2 ± 71 N) > OPI (605.7 ± 37 N) > BLTPI (498.3 ± 93 N). Within the two‐piece implants (TPI), tissue‐level designs (TL) outperformed bone‐level designs (BL), while designs with different abutment screw materials showed no significant difference ( p = 0.982). Conclusions Within the limitations of this in vitro study, the evaluated zirconia‐based OPI, BLTPI, and TLTPI systems demonstrate sufficient fracture resistance to support further clinical investigation. Implant‐level design (TL vs. BL) influences mechanical performance, whereas the influence of abutment screw material (ceramic vs. titanium) is less pronounced and varies according to the implant‐level design.

Clinical Oral Implants Research
Malmö University (SE), University of Oslo (NO), Umm al-Qura University (SA), Nordic Institute of Dental Materials (NO)
Openalex Percentile: Top 4%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.