A structured productivity challenge study of a fully automated mass spectrometry system

Abstract Objectives The Roche Cobas ® i601 is a fully automated mass spectrometry analyzer designed for 24 h random-access operation in routine medical laboratories. We aimed to systematically assess its throughput, robustness, and precision – along with its ease of use and walk-away time – within a high-volume clinical laboratory setting using the first commercially available reagents. Methods A panel of 13 measurands, implemented across three different reagent configurations (i pack Ionify ® reagents: Ionify steroids 1, Ionify steroids 2, Ionify vitamin D 25 and 24,25), was quantified using aliquots of three human serum pools. Identical series layouts were applied across four study days. After loading reagents, consumables, and samples and initiating the respective series, no further operator intervention was required until completion of the run. Results Across the four series, a total of 3,016 valid quantitative results were generated. Each series was completed within 6 h without operator intervention, corresponding to the minimum observed walk-away time. This equates to an average throughput of 119 analyses per hour. Analysis of the human serum pools demonstrated coefficients of variation (CVs) below 4 % (with few exceptions). Conclusions Under a high-throughput evaluation protocol in a routine clinical laboratory setting, the fully automated MS methods demonstrated performance and productivity characteristics consistent with the manufacturer’s representative data, particularly with regard to calibration stability, precision, analytical throughput, and walk-away time.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-06-12
DOI
https://doi.org/10.1515/cclm-2026-0163
Primary Topic
Hormonal and reproductive studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A structured productivity challenge study of a fully automated mass spectrometry system

Michael Vogeser, Bernd H. Northoff, Katharina Habler, Arber Rexhaj et al.
Clinical Chemistry and Laboratory Medicine (CCLM)
Hormonal and reproductive studies
article

A structured productivity challenge study of a fully automated mass spectrometry system

Michael Vogeser, Bernd H. Northoff, Katharina Habler, Arber Rexhaj, Florian Arend
article en

Abstract

Abstract Objectives The Roche Cobas ® i601 is a fully automated mass spectrometry analyzer designed for 24 h random-access operation in routine medical laboratories. We aimed to systematically assess its throughput, robustness, and precision – along with its ease of use and walk-away time – within a high-volume clinical laboratory setting using the first commercially available reagents. Methods A panel of 13 measurands, implemented across three different reagent configurations (i pack Ionify ® reagents: Ionify steroids 1, Ionify steroids 2, Ionify vitamin D 25 and 24,25), was quantified using aliquots of three human serum pools. Identical series layouts were applied across four study days. After loading reagents, consumables, and samples and initiating the respective series, no further operator intervention was required until completion of the run. Results Across the four series, a total of 3,016 valid quantitative results were generated. Each series was completed within 6 h without operator intervention, corresponding to the minimum observed walk-away time. This equates to an average throughput of 119 analyses per hour. Analysis of the human serum pools demonstrated coefficients of variation (CVs) below 4 % (with few exceptions). Conclusions Under a high-throughput evaluation protocol in a routine clinical laboratory setting, the fully automated MS methods demonstrated performance and productivity characteristics consistent with the manufacturer’s representative data, particularly with regard to calibration stability, precision, analytical throughput, and walk-away time.

Clinical Chemistry and Laboratory Medicine (CCLM)
Ludwig-Maximilians-Universität München (DE)
Openalex Percentile: Top 6%
Hormonal and reproductive studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A structured productivity challenge study of a fully automated mass spectrometry system — Michael Vogeser, Bernd H. Northoff, et al. · Clinical Chemistry and Laboratory Medicine (CCLM) (2026) | TGRS Research Map | TGRS