Reverberant wavefield characterization and directional filtering for penile vibro-elastography: a pilot feasibility study in post-prostatectomy erectile dysfunction

OBJECTIVE: Penile vibro-elastography is complicated by reverberant shear wavefields from the cylindrical geometry of the corpus cavernosum (CC), which bias shear wave speed (SWS) estimates. This study evaluated local phase velocity imaging with wavefield band-pass filtering (LPVI-WBF) for CC stiffness assessment in post-prostatectomy erectile dysfunction (ED), compared against phase gradient (PG) and time-of-flight (TOF) methods. APPROACH: Six patients with post-prostatectomy ED underwent penile vibroelastography at four longitudinal visits (902 acquisitions). SWS was estimated in flaccid and erect states at 100, 150, and 200 Hz using LPVI-WBF, PG, and TOF. Wavefield composition was quantified via the standing wave index. Simulations evaluated accuracy and frequency stability under clean and reverberant conditions. MAIN RESULTS: Reverberant components dominated penile acquisitions, with traveling wave energy absent in 59% of acquisitions and mean standing wave fraction of 57 ± 17%. LPVI-WBF was the only method producing physiologically consistent estimates, with erect SWS exceeding flaccid (3.59 ± 0.48 vs 3.25 ± 0.58 m/s, p < 0.0001). PG and TOF yielded reversed results (E < F, p < 0.0001). The dispersion exponent was near-zero for LPVI-WBF (α = 0.048, p = 0.21), consistent with weakly viscoelastic tissue, while PG (α = 0.184) and TOF (α = 0.321) showed significant frequency dependence from reverberant contamination. Simulations confirm PG accuracy in clean wavefields but collapse under reverberant ones. SIGNIFICANCE: LPVI-WBF produced SWS estimates consistent with expected CC biomechanics across all patients and frequencies, a criterion PG and TOF failed to satisfy. Phantom validation and multi-directional wavefield simulation are needed to establish accuracy.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-06-12
DOI
https://doi.org/10.1088/1361-6560/ae7cd7
Primary Topic
Sexual function and dysfunction studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reverberant wavefield characterization and directional filtering for penile vibro-elastography: a pilot feasibility study in post-prostatectomy erectile dysfunction

Xiaoming Zhang, Ngoc Thang Bui, Matt Ziegelmann, Tobias S Kohler
Physics in Medicine and Biology
Sexual function and dysfunction studies
article

Reverberant wavefield characterization and directional filtering for penile vibro-elastography: a pilot feasibility study in post-prostatectomy erectile dysfunction

Xiaoming Zhang, Ngoc Thang Bui, Matt Ziegelmann, Tobias S Kohler
article en

Abstract

OBJECTIVE: Penile vibro-elastography is complicated by reverberant shear wavefields from the cylindrical geometry of the corpus cavernosum (CC), which bias shear wave speed (SWS) estimates. This study evaluated local phase velocity imaging with wavefield band-pass filtering (LPVI-WBF) for CC stiffness assessment in post-prostatectomy erectile dysfunction (ED), compared against phase gradient (PG) and time-of-flight (TOF) methods. APPROACH: Six patients with post-prostatectomy ED underwent penile vibroelastography at four longitudinal visits (902 acquisitions). SWS was estimated in flaccid and erect states at 100, 150, and 200 Hz using LPVI-WBF, PG, and TOF. Wavefield composition was quantified via the standing wave index. Simulations evaluated accuracy and frequency stability under clean and reverberant conditions. MAIN RESULTS: Reverberant components dominated penile acquisitions, with traveling wave energy absent in 59% of acquisitions and mean standing wave fraction of 57 ± 17%. LPVI-WBF was the only method producing physiologically consistent estimates, with erect SWS exceeding flaccid (3.59 ± 0.48 vs 3.25 ± 0.58 m/s, p < 0.0001). PG and TOF yielded reversed results (E < F, p < 0.0001). The dispersion exponent was near-zero for LPVI-WBF (α = 0.048, p = 0.21), consistent with weakly viscoelastic tissue, while PG (α = 0.184) and TOF (α = 0.321) showed significant frequency dependence from reverberant contamination. Simulations confirm PG accuracy in clean wavefields but collapse under reverberant ones. SIGNIFICANCE: LPVI-WBF produced SWS estimates consistent with expected CC biomechanics across all patients and frequencies, a criterion PG and TOF failed to satisfy. Phantom validation and multi-directional wavefield simulation are needed to establish accuracy.

Physics in Medicine and Biology
WinnMed (US), Mayo Clinic in Florida (US)
Sexual Medicine Society of North America
Openalex Percentile: Top 6%
Sexual function and dysfunction studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.