Device–anatomy interactions in the dural venous sinuses: angioscopic insights from a perfused human cadaveric model

OBJECTIVE: Neuroendovascular venous interventions are increasingly performed using technologies originally developed for arterial procedures and indications. However, the major dural venous sinuses possess a unique intraluminal anatomy that is not present in arteries, raising concerns about device-anatomy interactions that may affect procedural performance. We used a perfused human cadaveric model with direct intraluminal angioscopic visualization to evaluate currently available endovascular devices within the dural venous sinuses and to characterize mechanisms of device-anatomy interactions associated with technical difficulty and failure. METHODS: Six fresh human head-and-neck cadaveric specimens were perfused with 0.9% saline solution via bilateral internal jugular vein catheterization using a peristaltic pump. Direct intraluminal angioscopic visualization was achieved through transcranial access to the major dural venous sinuses, allowing real-time observation of target segments during device manipulation. Standard endovascular maneuvers were performed within the dural venous sinuses, including guidewire and microcatheter navigation, catheter advancement, venous stent deployment, stent retriever deployment, aspiration thrombectomy, and balloon angioplasty. Angioscopic and fluoroscopic recordings were independently reviewed by experienced neurointerventionists to identify and categorize technical challenges and failure mechanisms. RESULTS: Angioscopy revealed multiple device-intraluminal interactions that were not fully appreciated on fluoroscopy alone. Several representative technical challenge and failure scenarios were identified and grouped into four principal mechanisms: (1) catheterization of venous channels parallel to the main sinus lumen, resulting in catheter entrapment and incomplete expansion of venous stents and stent retrievers; (2) device deformation or incomplete expansion due to intraluminal bands, including stent deformation, malposition, and constrained balloon angioplasty; (3) arrested or impaired device advancement caused by intraluminal bands, frequently necessitating microcatheter-assisted support to overcome ledge effects; and (4) interaction with arachnoid granulations leading to occlusion of aspiration catheter inlets and impeded intraluminal navigation. CONCLUSION: The venous system differs fundamentally from arteries in luminal geometry and internal architecture. Our findings demonstrate that arterial-derived devices incompletely accommodate these differences, resulting in parallel channel navigation, constrained expansion and deformation of stents, and occlusion of suction catheters. These findings highlight the fact that veins are not arteries and underscore the need for venous-specific techniques and technologies.

Authors

Institutions

Publication Details

Journal
Journal of NeuroInterventional Surgery
Published
2026-06-12
DOI
https://doi.org/10.1136/jnis-2026-025277
Primary Topic
Vascular Malformations Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Device–anatomy interactions in the dural venous sinuses: angioscopic insights from a perfused human cadaveric model

Ali Haider, Yiğit Can Şenol, Nishanth Krishnan, Matthew R. Amans et al.
Journal of NeuroInterventional Surgery
Vascular Malformations Diagnosis and Treatment
article

Device–anatomy interactions in the dural venous sinuses: angioscopic insights from a perfused human cadaveric model

Ali Haider, Yiğit Can Şenol, Nishanth Krishnan, Matthew R. Amans, L Savastano, Nanditha Krishnan, Prabhat Kumar, Adrian Liu, Panos Pardalidis, Benjamin Cho, Daniel Leong
article en

Abstract

OBJECTIVE: Neuroendovascular venous interventions are increasingly performed using technologies originally developed for arterial procedures and indications. However, the major dural venous sinuses possess a unique intraluminal anatomy that is not present in arteries, raising concerns about device-anatomy interactions that may affect procedural performance. We used a perfused human cadaveric model with direct intraluminal angioscopic visualization to evaluate currently available endovascular devices within the dural venous sinuses and to characterize mechanisms of device-anatomy interactions associated with technical difficulty and failure. METHODS: Six fresh human head-and-neck cadaveric specimens were perfused with 0.9% saline solution via bilateral internal jugular vein catheterization using a peristaltic pump. Direct intraluminal angioscopic visualization was achieved through transcranial access to the major dural venous sinuses, allowing real-time observation of target segments during device manipulation. Standard endovascular maneuvers were performed within the dural venous sinuses, including guidewire and microcatheter navigation, catheter advancement, venous stent deployment, stent retriever deployment, aspiration thrombectomy, and balloon angioplasty. Angioscopic and fluoroscopic recordings were independently reviewed by experienced neurointerventionists to identify and categorize technical challenges and failure mechanisms. RESULTS: Angioscopy revealed multiple device-intraluminal interactions that were not fully appreciated on fluoroscopy alone. Several representative technical challenge and failure scenarios were identified and grouped into four principal mechanisms: (1) catheterization of venous channels parallel to the main sinus lumen, resulting in catheter entrapment and incomplete expansion of venous stents and stent retrievers; (2) device deformation or incomplete expansion due to intraluminal bands, including stent deformation, malposition, and constrained balloon angioplasty; (3) arrested or impaired device advancement caused by intraluminal bands, frequently necessitating microcatheter-assisted support to overcome ledge effects; and (4) interaction with arachnoid granulations leading to occlusion of aspiration catheter inlets and impeded intraluminal navigation. CONCLUSION: The venous system differs fundamentally from arteries in luminal geometry and internal architecture. Our findings demonstrate that arterial-derived devices incompletely accommodate these differences, resulting in parallel channel navigation, constrained expansion and deformation of stents, and occlusion of suction catheters. These findings highlight the fact that veins are not arteries and underscore the need for venous-specific techniques and technologies.

Journal of NeuroInterventional Surgery
University of California, Los Angeles (US), University of California, San Francisco (US), Neurological Surgery (US)
Good health and well-being
Openalex Percentile: Top 6%
Vascular Malformations Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.