Development of a High-Performance Liquid Chromatography Assay Using Relative Molar Sensitivity for the Simultaneous Quantification of Clozapine and <i>N</i>-Desmethylclozapine Concentrations in Human Plasma

The LM1010 high-performance liquid chromatography (HPLC) system determines the plasma concentrations of clozapine using an absolute calibration method. In this study, the LM1010 HPLC method using relative molar sensitivity (RMS) was developed to quantify the plasma concentrations of clozapine and N-desmethylclozapine. The RMS values were calculated from the ratio of the peak area for clozapine or N-desmethylclozapine to that of highly pure and stable carbamazepine used as an RMS reference. The relative standard deviation of the RMS determined by three different laboratories was 1.723% for clozapine and 1.821% for N-desmethylclozapine, suggesting that the mean RMS values of 1.843 and 1.639, respectively, can be used for quantitative analysis in different laboratories. In a retrospective analysis of data obtained from a previous study using a conventional LM1010 method, Deming regression analyses showed a high correlation between the data calculated based on RMS values and the data obtained by HPLC using an internal standard method (slope = 0.942 and 0.974 for clozapine and N-desmethylclozapine, respectively). This result was equivalent to the regression between the results obtained from the LM1010 method using absolute calibration and the HPLC method. A prospective analysis of the RMS method showed strong correlations between RMS and HPLC results (slope = 0.922 and 1.037 for clozapine and N-desmethylclozapine, respectively) and was equivalent to correlations between results obtained from the LM1010 using absolute calibration and the HPLC methods. These findings suggest that the RMS method is suitable for the quantitative assay of plasma concentrations of clozapine and N-desmethylclozapine and offers an alternative to the current absolute calibration method.

Authors

Institutions

Publication Details

Journal
Iryo Yakugaku (Japanese Journal of Pharmaceutical Health Care and Sciences)
Published
2026-06-09
DOI
https://doi.org/10.5649/jjphcs.52.324
Primary Topic
Forensic Toxicology and Drug Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a High-Performance Liquid Chromatography Assay Using Relative Molar Sensitivity for the Simultaneous Quantification of Clozapine and N-Desmethylclozapine Concentrations in Human Plasma

Takashi Ohtsuki, Yumiko Akamine, Miyuki Matsushita, Masatomo Miura et al.
Iryo Yakugaku (Japanese Journal of Pharmaceutical Health Care and Sciences)
Forensic Toxicology and Drug Analysis
article

Development of a High-Performance Liquid Chromatography Assay Using Relative Molar Sensitivity for the Simultaneous Quantification of Clozapine and N-Desmethylclozapine Concentrations in Human Plasma

Takashi Ohtsuki, Yumiko Akamine, Miyuki Matsushita, Masatomo Miura, Hiroshi Matsufuji, Satoru Morikawa
article en

Abstract

The LM1010 high-performance liquid chromatography (HPLC) system determines the plasma concentrations of clozapine using an absolute calibration method. In this study, the LM1010 HPLC method using relative molar sensitivity (RMS) was developed to quantify the plasma concentrations of clozapine and N-desmethylclozapine. The RMS values were calculated from the ratio of the peak area for clozapine or N-desmethylclozapine to that of highly pure and stable carbamazepine used as an RMS reference. The relative standard deviation of the RMS determined by three different laboratories was 1.723% for clozapine and 1.821% for N-desmethylclozapine, suggesting that the mean RMS values of 1.843 and 1.639, respectively, can be used for quantitative analysis in different laboratories. In a retrospective analysis of data obtained from a previous study using a conventional LM1010 method, Deming regression analyses showed a high correlation between the data calculated based on RMS values and the data obtained by HPLC using an internal standard method (slope = 0.942 and 0.974 for clozapine and N-desmethylclozapine, respectively). This result was equivalent to the regression between the results obtained from the LM1010 method using absolute calibration and the HPLC method. A prospective analysis of the RMS method showed strong correlations between RMS and HPLC results (slope = 0.922 and 1.037 for clozapine and N-desmethylclozapine, respectively) and was equivalent to correlations between results obtained from the LM1010 using absolute calibration and the HPLC methods. These findings suggest that the RMS method is suitable for the quantitative assay of plasma concentrations of clozapine and N-desmethylclozapine and offers an alternative to the current absolute calibration method.

Iryo Yakugaku (Japanese Journal of Pharmaceutical Health Care and Sciences)Vol. 52(6)
Nihon University (JP), Akita University (JP), Akita University Hospital (JP), Hitachi (Japan) (JP)
Industry, innovation and infrastructure
Openalex Percentile: Top 7%
Forensic Toxicology and Drug Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.