Evaluation of Point‐of‐Care Coagulation Measurement Systems for Activated Clotting Time and Partial Thromboplastin Time in Heparinized and Non‐Heparinized Equine Blood

OBJECTIVES: To compare the performance of one point-of-care activated clotting time (ACT) machine (Vet-Tube), one point-of-care activated partial thromboplastin time (aPTT) machine (Coagulation DX), and one bench-top ACT machine (Medtronic ACT II plus) with previously published measurements from stationary devices in whole blood from healthy horses. DESIGN: Prospective, observational study SETTING: University veterinary teaching hospital, from August 2023 to September 2023. ANIMALS: Thirty healthy university-owned horses. METHODS: ACT and aPTT were measured via point-of-care analyzers (ACT POC, aPTT POC) and by conventional stationary analyzers (ACT Stationary, aPTT Stationary) in non-heparinized and heparinized blood collected from healthy horses. Samples were analyzed within 30 min of collection. RESULTS: Moderate correlation (r = 0.5), acceptable bias, and a low intraassay coefficient of variation (<5%) were seen for the ACT stationery device and ACT POC coagulation tubes. The ACT stationery and ACT POC showed no significant difference for both non-heparinized (n = 30, p = 0.64) and heparinized samples (n = 15, p = 0.43). Strong correlation but high bias and a high intraassay coefficient of variation (>5%) were seen for the aPTT POC device, with the aPTT stationary device being within previously published ranges. The aPTT stationary chemistry analyzer and aPTT POC showed a significant difference for non-heparinized samples (n = 30, p = 0.0002) but had no significant difference for heparinized samples (n = 15, p = 0.03). CLINICAL RELEVANCE: The results of this study provide proof of concept for the use of low-cost POC monitoring of ACT in equine patients with coagulopathy or undergoing heparin therapy. Rigorous evaluation in patients with clinical disease is still necessary.

Authors

Institutions

Publication Details

Journal
Journal of Veterinary Emergency and Critical Care
Published
2026-06-09
DOI
https://doi.org/10.1111/vec.70127
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Point‐of‐Care Coagulation Measurement Systems for Activated Clotting Time and Partial Thromboplastin Time in Heparinized and Non‐Heparinized Equine Blood

Bethanie L. Cooper, Megan Burke, Kallie J. Hobbs, Yu Ueda et al.
Journal of Veterinary Emergency and Critical Care
Clinical Laboratory Practices and Quality Control
article

Evaluation of Point‐of‐Care Coagulation Measurement Systems for Activated Clotting Time and Partial Thromboplastin Time in Heparinized and Non‐Heparinized Equine Blood

Bethanie L. Cooper, Megan Burke, Kallie J. Hobbs, Yu Ueda, M. Katie Sheats, Yang Qu
article en

Abstract

OBJECTIVES: To compare the performance of one point-of-care activated clotting time (ACT) machine (Vet-Tube), one point-of-care activated partial thromboplastin time (aPTT) machine (Coagulation DX), and one bench-top ACT machine (Medtronic ACT II plus) with previously published measurements from stationary devices in whole blood from healthy horses. DESIGN: Prospective, observational study SETTING: University veterinary teaching hospital, from August 2023 to September 2023. ANIMALS: Thirty healthy university-owned horses. METHODS: ACT and aPTT were measured via point-of-care analyzers (ACT POC, aPTT POC) and by conventional stationary analyzers (ACT Stationary, aPTT Stationary) in non-heparinized and heparinized blood collected from healthy horses. Samples were analyzed within 30 min of collection. RESULTS: Moderate correlation (r = 0.5), acceptable bias, and a low intraassay coefficient of variation (<5%) were seen for the ACT stationery device and ACT POC coagulation tubes. The ACT stationery and ACT POC showed no significant difference for both non-heparinized (n = 30, p = 0.64) and heparinized samples (n = 15, p = 0.43). Strong correlation but high bias and a high intraassay coefficient of variation (>5%) were seen for the aPTT POC device, with the aPTT stationary device being within previously published ranges. The aPTT stationary chemistry analyzer and aPTT POC showed a significant difference for non-heparinized samples (n = 30, p = 0.0002) but had no significant difference for heparinized samples (n = 15, p = 0.03). CLINICAL RELEVANCE: The results of this study provide proof of concept for the use of low-cost POC monitoring of ACT in equine patients with coagulopathy or undergoing heparin therapy. Rigorous evaluation in patients with clinical disease is still necessary.

Journal of Veterinary Emergency and Critical Care
North Carolina State University (US), Texas A&M University (US)
Good health and well-being
Openalex Percentile: Top 7%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.