Feasibility Study of a High-Flow Air-Cooled Metal-Tip Microwave Thermal Ablation Needle
Microwave (MW) ablation is a minimally invasive technique used to destroy pathological tissues through localized heating generated by a needle applicator. Internally cooled applicators using water circulation have long been the standard for high-power applications; however, water cooling introduces significant mechanical complexity. This work investigates the feasibility of a novel air-cooled coaxial thermal-ablation needle operating at 2.45 GHz up to 70 W. The system uses two concentric metal tubes—an outer 14 G stainless steel shaft (OD 2.1 mm) and an inner copper capillary (OD 1 mm, ID 0.7 mm)—serving simultaneously as the MW transmission line and cooling conduit, with dry air at room temperature (25 °C) flowing at 11 L/min under 5 bar input pressure. Experimental cooling efficiency tests demonstrated 78% efficiency for the shaft section in air and 32% for the section embedded in tissue. Electromagnetic and thermal simulations predicted ablation dimensions in a non-perfused liver of 35 mm short axis with ellipticity of 0.65 for the basic applicator, improving to 0.88 with an advanced PEEK-shaft design featuring a cancelling slot. A prototype was built and tested on exvivo bovine liver, achieving input matching better than −24 dB at 2.44 GHz and ablation dimensions (average of 5 tests) of 31 mm short axis and 45 mm long axis. Results confirm the feasibility of air cooling as a simpler, safer, and lower-cost alternative to water cooling for medium-power MW ablation.
Authors
- M Ricci (ORCID: https://orcid.org/0009-0001-9982-0441)
- Mattia Dimitri (ORCID: https://orcid.org/0000-0001-9719-9498)
- Guido Biffi Gentili
Institutions
- University of Florence (IT)
Publication Details
- Journal
- AppliedPhys
- Published
- 2026-06-09
- DOI
- https://doi.org/10.3390/appliedphys2020005
- Primary Topic
- Hepatocellular Carcinoma Treatment and Prognosis
- Type
- article
- Field-Weighted Citation Impact
- 0.00