Multicenter Interlaboratory Study of Thrombin Generation Assay in Healthy Adults Using the ST Genesia Analyzer With the STG‐ThromboScreen Reagent (With and Without Thrombomodulin)

INTRODUCTION: Fully automated thrombin generation (TG) assays facilitate the transition from research to clinical application. We conducted a multicenter inter-laboratory study using the ST Genesia analyzer to evaluate assay performance and establish reference intervals in healthy adults from Spain and Portugal. METHODS: In this multicenter study, plasma samples from healthy adult volunteers were analyzed across 12 laboratories following a harmonized protocol. TG was measured using the STG-ThromboScreen reagent in the presence and absence of thrombomodulin (TM). Analytical performance and reference intervals were assessed. RESULTS: Within-run and intra-day precision were excellent (coefficient of variation < 10%) for all TG parameters except velocity index. Normalization reduced interlaboratory variability for amplitude-based parameters such as endogenous thrombin potential (ETP) and peak height, whereas its effect was less pronounced for velocity index. Satisfactory interlaboratory performance (|z| ≤ 2) was achieved in 96%-98% of results, and normalized parameters showed minimal bias across centers. In 252 healthy adults, median normalized values were: ETP 131% (IQR 118-142), peak height 116% (IQR 104-127), lag time 2.6 min (IQR 2.3-2.8), and time-to-peak 5.5 min (IQR 5.0-5.9). Median ETP inhibition in the presence of TM was 47.2% (IQR 35.1-61.8). Age had a modest influence on TG parameters, whereas sex differences were observed. CONCLUSIONS: This multicenter interlaboratory study in the Iberian Peninsula demonstrates high reproducibility of TG measurement using the ST Genesia platform with the STG-ThromboScreen reagent. These findings support its use for standardized TG assessment and provide reference intervals for clinical and research settings.

Authors

Institutions

Publication Details

Journal
International Journal of Laboratory Hematology
Published
2026-06-09
DOI
https://doi.org/10.1111/ijlh.70168
Primary Topic
Blood Coagulation and Thrombosis Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multicenter Interlaboratory Study of Thrombin Generation Assay in Healthy Adults Using the ST Genesia Analyzer With the STG‐ThromboScreen Reagent (With and Without Thrombomodulin)

Laura Martos, Susanna Gassiot, Diego Velasco‐Rodríguez, Antonio Moscardó et al.
International Journal of Laboratory Hematology
Blood Coagulation and Thrombosis Mechanisms
article

Multicenter Interlaboratory Study of Thrombin Generation Assay in Healthy Adults Using the ST Genesia Analyzer With the STG‐ThromboScreen Reagent (With and Without Thrombomodulin)

Laura Martos, Susanna Gassiot, Diego Velasco‐Rodríguez, Antonio Moscardó, Joan Carles Reverter, Isabel Gutierrez‐Jomarrón, N. Vilalta, Santiago Bonanad, Pascual Marco, Inés Martínez-Alfonzo, Dolors Llobet, D Santos Ferreira, José Antonio Páramo, Carlos Aguilar, Sara Lopes, Pilar Llamas-Sillero, Ana Marco-Rico, Ángel Bernardo, Luciana Ricca, María Marcos‐Jubilar, João Mariano‐Pego, David Aragoneses
article en

Abstract

INTRODUCTION: Fully automated thrombin generation (TG) assays facilitate the transition from research to clinical application. We conducted a multicenter inter-laboratory study using the ST Genesia analyzer to evaluate assay performance and establish reference intervals in healthy adults from Spain and Portugal. METHODS: In this multicenter study, plasma samples from healthy adult volunteers were analyzed across 12 laboratories following a harmonized protocol. TG was measured using the STG-ThromboScreen reagent in the presence and absence of thrombomodulin (TM). Analytical performance and reference intervals were assessed. RESULTS: Within-run and intra-day precision were excellent (coefficient of variation < 10%) for all TG parameters except velocity index. Normalization reduced interlaboratory variability for amplitude-based parameters such as endogenous thrombin potential (ETP) and peak height, whereas its effect was less pronounced for velocity index. Satisfactory interlaboratory performance (|z| ≤ 2) was achieved in 96%-98% of results, and normalized parameters showed minimal bias across centers. In 252 healthy adults, median normalized values were: ETP 131% (IQR 118-142), peak height 116% (IQR 104-127), lag time 2.6 min (IQR 2.3-2.8), and time-to-peak 5.5 min (IQR 5.0-5.9). Median ETP inhibition in the presence of TM was 47.2% (IQR 35.1-61.8). Age had a modest influence on TG parameters, whereas sex differences were observed. CONCLUSIONS: This multicenter interlaboratory study in the Iberian Peninsula demonstrates high reproducibility of TG measurement using the ST Genesia platform with the STG-ThromboScreen reagent. These findings support its use for standardized TG assessment and provide reference intervals for clinical and research settings.

International Journal of Laboratory Hematology
Hospital Sant Joan de Déu Barcelona (ES), Hospital de Sant Pau (ES), Hospitais da Universidade de Coimbra (PT), Hospital de São João (PT), Hospital Universitari i Politècnic La Fe (ES), Stago (France) (FR), Hospital Universitario Príncipe de Asturias (ES), Hospital Clínic de Barcelona (ES), Hospital Universitario Fundación Jiménez Díaz (ES), Hospital Universitario Central de Asturias (ES), Clinica Universidad de Navarra (ES), Hospital General Universitario de Alicante Doctor Balmis (ES), Centro Hospitalar de Vila Nova de Gaia (PT)
Openalex Percentile: Top 7%
Blood Coagulation and Thrombosis Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.