Strengthening STD Screening Programs: Comprehensive Evaluation of High-Throughput Immunoassays for HIV and Syphilis Detection

Fourth-generation immunoassays are widely used for HIV and syphilis screening; however, false-reactive results may increase confirmatory testing and operational burden in high-throughput laboratories. This study evaluated the comparative performance of automated chemiluminescent immunoassays (MAGLUMI® HIV Ab/Ag Combi (Snibe Diagnostics Co. Ltd., Shenzhen, China), VITROS® ECiQ HIV Combo (Ortho Clinical Diagnostics, Raritan, NJ, USA), MAGLUMI® Syphilis (Snibe Diagnostics Co. Ltd., Shenzhen, China), and ARCHITECT® Syphilis TP (Abbott Diagnostics, Abbott Park, IL, USA) within a routine diagnostic algorithm, incorporating antibody differentiation immunoassays (INNO-LIA® HIV I/II Score (Fujirebio Europe N.V., Ghent, Belgium) and HIV-1 RNA PCR where applicable. A total of 240 archived serum samples for HIV testing and 180 for syphilis testing were analyzed. Agreement-based performance measures including sensitivity, specificity, overall percent agreement (OPA), and Cohen’s kappa (κ) were calculated as comparator-based estimates reflecting concordance within the routine diagnostic algorithm rather than absolute diagnostic accuracy against a universal reference standard. For comparisons with HIV-1 RNA PCR, positive and negative concordance rates are reported to reflect agreement between assays detecting different biological targets. Among samples with definitive (positive or negative) results, the MAGLUMI® HIV Ab/Ag Combi assay showed complete agreement with INNO-LIA® HIV I/II Score (κ = 1.00) and high agreement with PCR within the ARCHITECT® HIV Ag/Ab Combo-reactive subset (κ = 0.90). The VITROS® ECiQ HIV Combo assay demonstrated high agreement with INNO-LIA® HIV I/II Score (κ = 0.916) and substantial agreement with PCR (κ = 0.715), with a lower negative concordance rate with PCR observed in the ARCHITECT-reactive subset. A parallel five-modality analysis of 11 discordant samples applying the CDC 2014 algorithm demonstrated that all three immunoassay platforms successfully detected confirmed HIV-seropositive individuals with controlled viremia despite negative PCR, while MAGLUMI® HIV Ab/Ag Combi produced fewer false-reactive results than both ARCHITECT® and VITROS® in this discordant subset. Additionally, two cases showed INNO-LIA® indeterminate results with positive PCR, consistent with acute HIV infection during the early seroconversion stage; all three immunoassay platforms produced signals above the non-reactive threshold in both cases. For syphilis testing, both MAGLUMI® Syphilis and ARCHITECT® Syphilis TP assays showed complete agreement with INNO-LIA® Syphilis Score among samples with definitive results (κ = 1.00). In contrast, the RPR assay showed reduced positive predictive value (49.4%) and moderate agreement with INNO-LIA® Syphilis Score (κ = 0.52). Automated chemiluminescent immunoassay (CLIA) platforms demonstrated high agreement within a structured diagnostic algorithm in a high-throughput screening setting. Differences in assay performance were observed across platforms, particularly with respect to discordant results in the ARCHITECT-reactive PCR-evaluated subset for HIV and non-treponemal concordance for syphilis. These platforms may support more efficient laboratory workflows; however, findings should be interpreted within the context of comparator-based classification rather than absolute diagnostic accuracy.

Authors

Institutions

Publication Details

Journal
Microorganisms
Published
2026-06-09
DOI
https://doi.org/10.3390/microorganisms14061302
Primary Topic
Syphilis Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Strengthening STD Screening Programs: Comprehensive Evaluation of High-Throughput Immunoassays for HIV and Syphilis Detection

Gheyath K. Nasrallah, Mutaz Mohamed Ali, Ibrahim Al Shaar, Parveen Banu Nizamuddin et al.
Microorganisms
Syphilis Diagnosis and Treatment
article

Strengthening STD Screening Programs: Comprehensive Evaluation of High-Throughput Immunoassays for HIV and Syphilis Detection

Gheyath K. Nasrallah, Mutaz Mohamed Ali, Ibrahim Al Shaar, Parveen Banu Nizamuddin, Shaden Abunasser, Ahmed Ismail, Mazen Najib Abouassali, Ibrahim Wissam Karimeh, Mohammed Abdelfatah Ibrahim, Israa M. Salameh, Salma Younes, Manal Elshaikh, Nadin Younes, Laith J. Abu-Raddad, Hadi M. Yassine
article en

Abstract

Fourth-generation immunoassays are widely used for HIV and syphilis screening; however, false-reactive results may increase confirmatory testing and operational burden in high-throughput laboratories. This study evaluated the comparative performance of automated chemiluminescent immunoassays (MAGLUMI® HIV Ab/Ag Combi (Snibe Diagnostics Co. Ltd., Shenzhen, China), VITROS® ECiQ HIV Combo (Ortho Clinical Diagnostics, Raritan, NJ, USA), MAGLUMI® Syphilis (Snibe Diagnostics Co. Ltd., Shenzhen, China), and ARCHITECT® Syphilis TP (Abbott Diagnostics, Abbott Park, IL, USA) within a routine diagnostic algorithm, incorporating antibody differentiation immunoassays (INNO-LIA® HIV I/II Score (Fujirebio Europe N.V., Ghent, Belgium) and HIV-1 RNA PCR where applicable. A total of 240 archived serum samples for HIV testing and 180 for syphilis testing were analyzed. Agreement-based performance measures including sensitivity, specificity, overall percent agreement (OPA), and Cohen’s kappa (κ) were calculated as comparator-based estimates reflecting concordance within the routine diagnostic algorithm rather than absolute diagnostic accuracy against a universal reference standard. For comparisons with HIV-1 RNA PCR, positive and negative concordance rates are reported to reflect agreement between assays detecting different biological targets. Among samples with definitive (positive or negative) results, the MAGLUMI® HIV Ab/Ag Combi assay showed complete agreement with INNO-LIA® HIV I/II Score (κ = 1.00) and high agreement with PCR within the ARCHITECT® HIV Ag/Ab Combo-reactive subset (κ = 0.90). The VITROS® ECiQ HIV Combo assay demonstrated high agreement with INNO-LIA® HIV I/II Score (κ = 0.916) and substantial agreement with PCR (κ = 0.715), with a lower negative concordance rate with PCR observed in the ARCHITECT-reactive subset. A parallel five-modality analysis of 11 discordant samples applying the CDC 2014 algorithm demonstrated that all three immunoassay platforms successfully detected confirmed HIV-seropositive individuals with controlled viremia despite negative PCR, while MAGLUMI® HIV Ab/Ag Combi produced fewer false-reactive results than both ARCHITECT® and VITROS® in this discordant subset. Additionally, two cases showed INNO-LIA® indeterminate results with positive PCR, consistent with acute HIV infection during the early seroconversion stage; all three immunoassay platforms produced signals above the non-reactive threshold in both cases. For syphilis testing, both MAGLUMI® Syphilis and ARCHITECT® Syphilis TP assays showed complete agreement with INNO-LIA® Syphilis Score among samples with definitive results (κ = 1.00). In contrast, the RPR assay showed reduced positive predictive value (49.4%) and moderate agreement with INNO-LIA® Syphilis Score (κ = 0.52). Automated chemiluminescent immunoassay (CLIA) platforms demonstrated high agreement within a structured diagnostic algorithm in a high-throughput screening setting. Differences in assay performance were observed across platforms, particularly with respect to discordant results in the ARCHITECT-reactive PCR-evaluated subset for HIV and non-treponemal concordance for syphilis. These platforms may support more efficient laboratory workflows; however, findings should be interpreted within the context of comparator-based classification rather than absolute diagnostic accuracy.

MicroorganismsVol. 14(6)
Cornell University (US), Ministry of Public Health (QA), Weill Cornell Medical College in Qatar (QA), Qatar University (QA)
Qatar National Research Fund
Good health and well-being
Openalex Percentile: Top 7%
Syphilis Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.