Development and performance evaluation of a phantom for dose measurement in radiation blood irradiators

This study developed a specialized alanine-based PMMA phantom replicating the geometry of a 300 mL blood bag to improve dose distribution measurement in blood irradiators. Using precision CNC machining, ten dosimeter slots were positioned at 35 mm intervals to evaluate internal and external dose gradients. Experiments using a Cs-137 irradiator at 25 Gy showed that outside doses (C1–C5) ranged from 25.82 to 31.38 Gy, while inside doses (C6–C10) ranged from 19.93 to 25.71 Gy. The central inside position (C8) showed high accuracy (25.71 ± 0.73 Gy), and the mean error remained within ±3% over repeated trials. These results demonstrate that the developed phantom significantly reduces positioning uncertainty and ensures high reproducibility, satisfying international QA guidelines for blood irradiation.

Authors

Institutions

Publication Details

Journal
Nuclear Engineering and Technology
Published
2026-06-09
DOI
https://doi.org/10.1016/j.net.2026.104465
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and performance evaluation of a phantom for dose measurement in radiation blood irradiators

Jae-Hwan Cho, Jeong-Yeon Han, Gyu-Ha Ryu, Hyeon-Jun Hwang et al.
Nuclear Engineering and Technology
Advanced Radiotherapy Techniques
article

Development and performance evaluation of a phantom for dose measurement in radiation blood irradiators

Jae-Hwan Cho, Jeong-Yeon Han, Gyu-Ha Ryu, Hyeon-Jun Hwang, Seung-Wook Baek, Yong-Min Kim, Sang-Hwii Lee, Jung-Lye Ko
article en

Abstract

This study developed a specialized alanine-based PMMA phantom replicating the geometry of a 300 mL blood bag to improve dose distribution measurement in blood irradiators. Using precision CNC machining, ten dosimeter slots were positioned at 35 mm intervals to evaluate internal and external dose gradients. Experiments using a Cs-137 irradiator at 25 Gy showed that outside doses (C1–C5) ranged from 25.82 to 31.38 Gy, while inside doses (C6–C10) ranged from 19.93 to 25.71 Gy. The central inside position (C8) showed high accuracy (25.71 ± 0.73 Gy), and the mean error remained within ±3% over repeated trials. These results demonstrate that the developed phantom significantly reduces positioning uncertainty and ensures high reproducibility, satisfying international QA guidelines for blood irradiation.

Nuclear Engineering and TechnologyVol. 58(10)
Samsung Medical Center (KR), Ansan University (KR), Biosolution (South Korea) (KR), Sungkyunkwan University (KR)
Openalex Percentile: Top 7%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.