The impact of a novel deep learning reconstruction algorithm on image quality in ultralow-dose CT: a quantitative phantom study

OBJECTIVE: The aim of this study is to evaluate the performance of a novel deep learning image reconstruction (DLIR) algorithm in noise reduction, contrast-to-noise ratio (CNR), and low iodine concentration detection for ultralow-dose computed tomography (CT) imaging. METHODS: A nine-hole phantom with iodine concentrations (0-40 mg/mL) was scanned at various tube voltages (60-120 kVp). Images were reconstructed using filtered back projection (FBP), iterative reconstruction (IR), and DLIR at different weight levels (10%-90%). Objective metrics (noise, CNR, CT value accuracy via Bland-Altman analysis) and subjective image quality were assessed. RESULTS: At all tube voltages (60-120 kVp), DLIR with medium-to-high weight levels (50%-90%) reduced background noise and increased CNR compared with FBP and IR (p < 0.001). The CNR at a low iodine concentration (1.25 mg/mL) was enhanced, and the DLIR algorithm (weight levels 30%-90%) was able to continuously detect an iodine concentration of 1.25 mg/mL (CNR ≥ 3) at all tube voltages. Under fixed ultralow-dose conditions, DLIR preserved image quality and low-contrast detectability. DLIR (weight levels 90%) reduced background noise by 84.7% compared with FBP and improved CNR (p < 0.001). Bland-Altman analysis confirmed excellent quantitative accuracy for DLIR. The exploratory subjective evaluation was consistent with objective metrics. CONCLUSION: The DLIR algorithm can enhance image quality in low-dose CT imaging and improve the ability to detect low concentrations of iodine. These findings demonstrate that DLIR maintains image quality and CNR at low iodine concentrations in phantom studies. Clinical implications require further validation. RELEVANCE STATEMENT: This phantom study shows that the deep learning reconstruction algorithm can still maintain the diagnostic image quality and low-contrast detectability even under ultralow-dose CT (94% dose reduction). These findings support further clinical research to optimize the dosage regimens and potentially reduce the use of iodine contrast agents. KEY POINTS: Under ultralow-dose conditions (60 kV), DLIR preserved image quality metrics and detectability thresholds in a phantom under ultralow-dose conditions. It significantly suppressed image noise and improved the CNR. The algorithm reliably detected low iodine concentrations (1.25 mg/mL) at all dose levels.

Authors

Institutions

Publication Details

Journal
European Radiology Experimental
Published
2026-06-08
DOI
https://doi.org/10.1186/s41747-026-00751-w
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The impact of a novel deep learning reconstruction algorithm on image quality in ultralow-dose CT: a quantitative phantom study

Huairong Zhang, Yun Shen, Yongjun Jia, Tong Su
European Radiology Experimental
Radiation Dose and Imaging
article

The impact of a novel deep learning reconstruction algorithm on image quality in ultralow-dose CT: a quantitative phantom study

Huairong Zhang, Yun Shen, Yongjun Jia, Tong Su
article en

Abstract

OBJECTIVE: The aim of this study is to evaluate the performance of a novel deep learning image reconstruction (DLIR) algorithm in noise reduction, contrast-to-noise ratio (CNR), and low iodine concentration detection for ultralow-dose computed tomography (CT) imaging. METHODS: A nine-hole phantom with iodine concentrations (0-40 mg/mL) was scanned at various tube voltages (60-120 kVp). Images were reconstructed using filtered back projection (FBP), iterative reconstruction (IR), and DLIR at different weight levels (10%-90%). Objective metrics (noise, CNR, CT value accuracy via Bland-Altman analysis) and subjective image quality were assessed. RESULTS: At all tube voltages (60-120 kVp), DLIR with medium-to-high weight levels (50%-90%) reduced background noise and increased CNR compared with FBP and IR (p < 0.001). The CNR at a low iodine concentration (1.25 mg/mL) was enhanced, and the DLIR algorithm (weight levels 30%-90%) was able to continuously detect an iodine concentration of 1.25 mg/mL (CNR ≥ 3) at all tube voltages. Under fixed ultralow-dose conditions, DLIR preserved image quality and low-contrast detectability. DLIR (weight levels 90%) reduced background noise by 84.7% compared with FBP and improved CNR (p < 0.001). Bland-Altman analysis confirmed excellent quantitative accuracy for DLIR. The exploratory subjective evaluation was consistent with objective metrics. CONCLUSION: The DLIR algorithm can enhance image quality in low-dose CT imaging and improve the ability to detect low concentrations of iodine. These findings demonstrate that DLIR maintains image quality and CNR at low iodine concentrations in phantom studies. Clinical implications require further validation. RELEVANCE STATEMENT: This phantom study shows that the deep learning reconstruction algorithm can still maintain the diagnostic image quality and low-contrast detectability even under ultralow-dose CT (94% dose reduction). These findings support further clinical research to optimize the dosage regimens and potentially reduce the use of iodine contrast agents. KEY POINTS: Under ultralow-dose conditions (60 kV), DLIR preserved image quality metrics and detectability thresholds in a phantom under ultralow-dose conditions. It significantly suppressed image noise and improved the CNR. The algorithm reliably detected low iodine concentrations (1.25 mg/mL) at all dose levels.

European Radiology ExperimentalVol. 10(1)
Ningxia Medical University (CN), Neusoft (China) (CN), Affiliated Hospital of Shaanxi University of Chinese Medicine (CN), Ningxia Medical University General Hospital (CN)
Openalex Percentile: Top 6%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.