Clinical and Echocardiographic Outcomes After Implantation of the ALLEGRA Transcatheter Valve Using the Fully Repositionable IMPERIA Delivery System: One‐Year Results of the EMPIRE I Study
BACKGROUND: The EMPIRE I study assessed the early (30 days) safety and efficacy of a new fully repositionable delivery system (IMPERIA) for the commercially available ALLEGRA transcatheter aortic valve in patients with either severe native aortic stenosis or a degenerated surgical bioprosthesis. AIM: To report the final 1-year results of the EMPIRE I study. METHODS: EMPIRE I is a pre-market, prospective, multi-center, single-arm study conducted at 11 centers in 3 European countries. There were independent data safety and clinical events committees and an independent echocardiographic core laboratory. In this analysis, VARC-3 definitions of all clinical endpoints and echocardiographic parameters were employed. RESULTS: The study recruited 121 patients with severe native aortic stenosis and 16 with degenerated surgical bioprosthesis. At 1 year, the incidence of bioprosthetic valve dysfunction in the native population was 5.9%. No patients developed structural valve deterioration or endocarditis, and there were no patients with haemodynamic valve deterioration or bioprosthetic valve failure. In the valve-in-valve sub-group, 2/14 (14.3%) patients had moderate patient prosthesis mismatch at 30 days with no evidence of new bioprosthetic valve dysfunction at 1 year. In the native population, new permanent pacemaker implantation was 14.3% at 30 days and remained unchanged at 1 year. Cardiovascular mortality was 6.8% in the native population and 0% in the valve-in-valve patients. At 1-year, all-cause mortality was 10.8% (95% CI 6.4%-17.8%) in the native aortic stenosis population and 0% in the valve-in-valve patients. CONCLUSION: After 1 year of follow-up, the results of the EMPIRE 1 study show excellent haemodynamic performance, a low rate of bioprosthetic valve dysfunction, and no bioprosthetic valve failure in patients treated with the ALLEGRATM trans-catheter aortic valve implanted using the new IMPERIATM fully repositionable delivery system.
Authors
- Joan Antoni Gomez
- Samuel Copt (ORCID: https://orcid.org/0000-0001-8378-9729)
- Christof Burgdorf (ORCID: https://orcid.org/0000-0002-8130-2443)
- Keith G. Oldroyd (ORCID: https://orcid.org/0000-0002-7842-3463)
- Ben Ren (ORCID: https://orcid.org/0000-0002-8227-8094)
- Anders Jönsson
- Nicolas Van Mieghem
- Pim Tonino
- I Cruz-Gonzalez
- Ernest Spitzer
- Pieter Vriesendorp
- Nicolas Doll
- José Maria Hernândez
- José Antonio Baz
- Christian Frerker
- Pilar Jimenez
Institutions
- Radboud University Nijmegen (NL)
- Catharina Ziekenhuis (NL)
- Erasmus MC (NL)
- Bellvitge University Hospital (ES)
- University Hospital Complex Of Vigo (ES)
- Hospital Clínico San Carlos (ES)
- Maastricht University (NL)
- Universitäres Kinderwunschzentrum Lübeck (DE)
- Complejo Hospitalario de Salamanca (ES)
- Schüchtermann-Klinik (DE)
- Biosensores (Spain) (ES)
- Klinik Lüneburger Heide (DE)
- Instituto de Investigación Biomédica de Salamanca (ES)
- Hospital Clínico Universitario Virgen de la Victoria (ES)
- Cardialysis (Netherlands) (NL)
- University of Lübeck (DE)
Publication Details
- Journal
- Catheterization and Cardiovascular Interventions
- Published
- 2026-05-31
- DOI
- https://doi.org/10.1002/ccd.70675
- Primary Topic
- Cardiac Valve Diseases and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00