Safety and Feasibility of Pulsed Field Ablation With a Pentaspline Catheter in Patients With Extravascular ICD

Pulsed-field ablation (PFA) in patients with an implanted extravascular implantable cardioverter-defibrillator (EV-ICD) poses theoretical risks of electromagnetic interference or lead displacement. We report two cases of atrial fibrillation (AF) ablation using the FARAPULSE pentaspline catheter after EV-ICD implantation. In both patients, PFA was successfully performed without significant changes of the parameters/functions of the EV-ICDs. No lead dislodgement was observed on a fluoroscopy examination. These cases suggest that AF ablation using PFA can be safely performed in patients with EV-ICDs.

Authors

Institutions

Publication Details

Journal
Pacing and Clinical Electrophysiology
Published
2026-05-31
DOI
https://doi.org/10.1111/pace.70312
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Safety and Feasibility of Pulsed Field Ablation With a Pentaspline Catheter in Patients With Extravascular ICD

Masato Fukunaga, Rei Kuji, Jun Hirokami, Tomonori Katsuki et al.
Pacing and Clinical Electrophysiology
Cardiac Arrhythmias and Treatments
article

Safety and Feasibility of Pulsed Field Ablation With a Pentaspline Catheter in Patients With Extravascular ICD

Masato Fukunaga, Rei Kuji, Jun Hirokami, Tomonori Katsuki, Kengo Korai, Norihito Oyanagi, Keigo Misonou, Kenichi Hiroshima, Koumei Onuki, M Nagashima, Hidetaka Chinen, Hiroki Nagai, Maiko Kuroda, Hiroyuki Kono, Kenji Ando
article en

Abstract

Pulsed-field ablation (PFA) in patients with an implanted extravascular implantable cardioverter-defibrillator (EV-ICD) poses theoretical risks of electromagnetic interference or lead displacement. We report two cases of atrial fibrillation (AF) ablation using the FARAPULSE pentaspline catheter after EV-ICD implantation. In both patients, PFA was successfully performed without significant changes of the parameters/functions of the EV-ICDs. No lead dislodgement was observed on a fluoroscopy examination. These cases suggest that AF ablation using PFA can be safely performed in patients with EV-ICDs.

Pacing and Clinical Electrophysiology
Kokura Memorial Hospital (JP)
Good health and well-being
Openalex Percentile: Top 6%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.