<scp>SNR</scp> ‐Enhanced Whole‐Brain Chemical Exchange Saturation Transfer Imaging by Optimized Variable Flip Angles

ABSTRACT Purpose To improve the signal‐to‐noise ratio (SNR) and mitigate T 2 ‐decay‐related blurring in whole‐brain chemical exchange saturation transfer (CEST) imaging by optimizing the refocusing flip angles of a turbo spin echo (TSE) readout, also known as the SPACE‐CEST sequence. Methods The SNR of the SPACE‐CEST sequence was modeled as a function of flip angles, and a resolution penalty term was added to control blurring. Flip angles were optimized under two modes: Resolution‐only and SNR+Resolution mode. The optimized flip angles were compared to conventional variable/constant flip angles regarding transverse magnetization usage efficiency, longitudinal magnetization storage, and specific absorption rate (SAR). The SNR and resolution benefits of the proposed method were validated in a phantom, healthy volunteers, and brain tumor patients. Results The optimized flip angles in the Resolution‐only mode generated a constant signal, validating the framework. The optimized flip angles in SNR+Resolution mode exhibited higher transverse magnetization usage efficiency and greater longitudinal storage. The SNR+Resolution mode produced signals comparable to those from conventional constant flip angles but with significantly lower flip angles. Notably, the SNR+Resolution mode demonstrated high transverse magnetization usage efficiency even with 20% B 1 inhomogeneity and across a wide range T 1 and T 2 values, along with a 32.9% decrease in SAR. On average, the SNR+Resolution mode showed a 14.2% increase in SNR in humans and a 3.6% improvement in spatial resolution. Conclusion Higher transverse magnetization usage efficiency can be achieved by optimizing refocusing flip angles of the SPACE‐CEST sequence, leading to increased SNR and decreased SAR.

Authors

Institutions

Publication Details

Journal
Magnetic Resonance in Medicine
Published
2026-05-30
DOI
https://doi.org/10.1002/mrm.70432
Primary Topic
Lanthanide and Transition Metal Complexes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

SNR ‐Enhanced Whole‐Brain Chemical Exchange Saturation Transfer Imaging by Optimized Variable Flip Angles

Tao Zu, Y Zhang, Xingwang Yong, Shanshan Lu et al.
Magnetic Resonance in Medicine
Lanthanide and Transition Metal Complexes
article

SNR ‐Enhanced Whole‐Brain Chemical Exchange Saturation Transfer Imaging by Optimized Variable Flip Angles

Tao Zu, Y Zhang, Xingwang Yong, Shanshan Lu, Chunqiu Su, Zhechuan Dai, Yi‐Cheng Hsu
article en

Abstract

ABSTRACT Purpose To improve the signal‐to‐noise ratio (SNR) and mitigate T 2 ‐decay‐related blurring in whole‐brain chemical exchange saturation transfer (CEST) imaging by optimizing the refocusing flip angles of a turbo spin echo (TSE) readout, also known as the SPACE‐CEST sequence. Methods The SNR of the SPACE‐CEST sequence was modeled as a function of flip angles, and a resolution penalty term was added to control blurring. Flip angles were optimized under two modes: Resolution‐only and SNR+Resolution mode. The optimized flip angles were compared to conventional variable/constant flip angles regarding transverse magnetization usage efficiency, longitudinal magnetization storage, and specific absorption rate (SAR). The SNR and resolution benefits of the proposed method were validated in a phantom, healthy volunteers, and brain tumor patients. Results The optimized flip angles in the Resolution‐only mode generated a constant signal, validating the framework. The optimized flip angles in SNR+Resolution mode exhibited higher transverse magnetization usage efficiency and greater longitudinal storage. The SNR+Resolution mode produced signals comparable to those from conventional constant flip angles but with significantly lower flip angles. Notably, the SNR+Resolution mode demonstrated high transverse magnetization usage efficiency even with 20% B 1 inhomogeneity and across a wide range T 1 and T 2 values, along with a 32.9% decrease in SAR. On average, the SNR+Resolution mode showed a 14.2% increase in SNR in humans and a 3.6% improvement in spatial resolution. Conclusion Higher transverse magnetization usage efficiency can be achieved by optimizing refocusing flip angles of the SPACE‐CEST sequence, leading to increased SNR and decreased SAR.

Magnetic Resonance in Medicine
Zhejiang Lab (CN), Siemens Healthcare (United States) (US), Jiangsu Province Hospital (CN), Nanjing Medical University (CN)
National Key Research and Development Program of China, Fundamental Research Funds for the Central Universities
Openalex Percentile: Top 12%
Lanthanide and Transition Metal Complexes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.