Human factors validation study of a disposable autoinjector (Bmab 1200 autoinjector) for subcutaneous delivery of Ustekinumab
= 7), none causing device damage or failed delivery. The knowledge task success rate was 99.6% (2,511/2,520 measures). Nine use errors (study artifacts or inattention to instructions) occurred only among caregivers. All participants expressed confidence in safe dosing and found the device user-friendly. The Bmab 1200 autoinjector demonstrated high usability and safety across diverse users for ustekinumab biosimilar delivery in chronic autoimmune conditions.