OBJECTIVES: This study evaluated the effects of restorative materials and intraoral scanners on the mesh data quality of dental models and the accuracy of 3D-printed clear aligners. MATERIALS AND METHODS: Reference scan data were obtained with a tabletop scanner, while test data were collected using PrimeScan, Trios 3, and i600 scanners. Typodonts with metallic (gold) and non-metallic (zirconia) crowns were scanned. Data from each scanner were assessed for mesh quality and 3D accuracy. 3D printed aligners were subsequently printed based on each scanner's mesh data and evaluated. RESULTS: The i600 scanner had the highest mesh quality (0.81), followed by PrimeScan (0.76) and Trios 3 (0.74) scanners, whereas the restorative material had no significant effect on mesh quality (p < 0.05). Resin and zirconia crowns were generally measured larger than actual sizes, while gold crowns tended to be similarly measured or smaller. PrimeScan showed the highest accuracy for full-arch scans, while Trios 3 measured significantly smaller than the other two scanners p < 0.05). For gold crowns, i600 displayed the highest RMS value at 59 μm p < 0.05). Aligners printed from i600 scans showed the largest RMS value on the inner surface near gold crowns, averaging 94 μm. CONCLUSIONS: Types of restorative material and scanner can significantly affect the accuracy of 3D printed aligners and mesh quality. Metallic restorations can decrease accuracy with triangulation-based scanners. CLINICAL RELEVANCE: Clinical application of intraoral scanners that are less influenced by restorative materials, particularly metallic restorations, may improve the predictability of 3D-printed aligner treatment and help optimize clinical outcomes.

Authors

Institutions

Publication Details

Journal
BMC Oral Health
Published
2026-05-29
DOI
https://doi.org/10.1186/s12903-026-08734-4
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of restorative material types on 3D printed aligner accuracy using intraoral scanners

Hyung‐Seog Yu, Ji-Hoi Kim, Cha Jk, Jae-Hun Yu et al.
BMC Oral Health
Dental materials and restorations
article

Influence of restorative material types on 3D printed aligner accuracy using intraoral scanners

Hyung‐Seog Yu, Ji-Hoi Kim, Cha Jk, Jae-Hun Yu, Sung-Hwan Choi, Kee-Joon Lee, Utkarsh Mangal, Jing Liu, Jae-Sung Kwon
article en

Abstract

OBJECTIVES: This study evaluated the effects of restorative materials and intraoral scanners on the mesh data quality of dental models and the accuracy of 3D-printed clear aligners. MATERIALS AND METHODS: Reference scan data were obtained with a tabletop scanner, while test data were collected using PrimeScan, Trios 3, and i600 scanners. Typodonts with metallic (gold) and non-metallic (zirconia) crowns were scanned. Data from each scanner were assessed for mesh quality and 3D accuracy. 3D printed aligners were subsequently printed based on each scanner's mesh data and evaluated. RESULTS: The i600 scanner had the highest mesh quality (0.81), followed by PrimeScan (0.76) and Trios 3 (0.74) scanners, whereas the restorative material had no significant effect on mesh quality (p < 0.05). Resin and zirconia crowns were generally measured larger than actual sizes, while gold crowns tended to be similarly measured or smaller. PrimeScan showed the highest accuracy for full-arch scans, while Trios 3 measured significantly smaller than the other two scanners p < 0.05). For gold crowns, i600 displayed the highest RMS value at 59 μm p < 0.05). Aligners printed from i600 scans showed the largest RMS value on the inner surface near gold crowns, averaging 94 μm. CONCLUSIONS: Types of restorative material and scanner can significantly affect the accuracy of 3D printed aligners and mesh quality. Metallic restorations can decrease accuracy with triangulation-based scanners. CLINICAL RELEVANCE: Clinical application of intraoral scanners that are less influenced by restorative materials, particularly metallic restorations, may improve the predictability of 3D-printed aligner treatment and help optimize clinical outcomes.

BMC Oral Health
Yonsei University (KR), Yonsei University Dental Hospital (KR), Center for Systems Biology (US), Yonsei University Health System (KR), Yonsei University College of Dentistry, Bucheon University (KR)
Korea Evaluation Institute of Industrial Technology
Openalex Percentile: Top 5%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.