The Determination of potentially toxic elements (Arsenic, Cadmium, Mercury, and Lead) in Hemp (Cannabis sativa L.) from using Inductively Coupled Plasma Optical Emission Spectroscopy (ICP OES)

BACKGROUND: Hemp (Cannabis sativa L.) readily accumulates potentially toxic elements (PTEs) from soil. PTE exposure is linked to cancer, nephrotoxicity, and neurotoxicity. Non-intoxicating hemp products in New Jersey are not subject to mandatory PTE testing, posing a consumer safety risk. OBJECTIVE: This study presents a single-laboratory validation of an ICP OES method for the determination of As, Cd, Pb, and Hg in hemp, evaluated against AOAC SMPR 2020.001 performance criteria. METHODS: Ground hemp (0.125 g) was digested using HNO3/HCl/H2O2 (190 °C, 20 min) and diluted to 50 mL. Samples were analyzed on an Agilent 5800 ICP OES with a seven-point calibration (1-100 μg/L), Au/Sc internal standards, and KHP carbon for matrix matching. RESULTS: Spike recoveries for Cd (92.7%) and Pb (96.2%) met AOAC SMPR 2020.001 criteria (80-115%). Mercury recovery (47%) fell below acceptance limits due to matrix suppression and volatility losses. The As LOD (0.807 μg/g) exceeded sample concentrations in most specimens. LOQ values for all four elements exceeded the SMPR 2020.001 threshold of 10 µg/kg (0.01 µg/g). Calibration linearity was confirmed for all analytes (r ≥ 0.995). CONCLUSION: Partial validation was achieved: Cd and Pb demonstrated acceptable accuracy and repeatability, but LOQ criteria were not met for any analyte. ICP OES is not suitable for Hg or As at regulatory trace levels; ICP MS is recommended for these elements. This method offers a cost-effective screening tool for Cd and Pb in hemp. HIGHLIGHTS: This study presents the first single-laboratory ICP OES validation for PTEs in hemp under AOAC SMPR 2020.001. Cd and Pb met accuracy and repeatability criteria; As and Hg did not meet sensitivity requirements. ICP MS is recommended for regulatory As and Hg testing in hemp.

Authors

Institutions

Publication Details

Journal
Journal of AOAC International
Published
2026-05-29
DOI
https://doi.org/10.1093/jaoacint/qsag044
Primary Topic
Heavy Metal Exposure and Toxicity
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Determination of potentially toxic elements (Arsenic, Cadmium, Mercury, and Lead) in Hemp (Cannabis sativa L.) from using Inductively Coupled Plasma Optical Emission Spectroscopy (ICP OES)

Ghazal A Khan, Esha G Sachdev
Journal of AOAC International
Heavy Metal Exposure and Toxicity
article

The Determination of potentially toxic elements (Arsenic, Cadmium, Mercury, and Lead) in Hemp (Cannabis sativa L.) from using Inductively Coupled Plasma Optical Emission Spectroscopy (ICP OES)

Ghazal A Khan, Esha G Sachdev
article en

Abstract

BACKGROUND: Hemp (Cannabis sativa L.) readily accumulates potentially toxic elements (PTEs) from soil. PTE exposure is linked to cancer, nephrotoxicity, and neurotoxicity. Non-intoxicating hemp products in New Jersey are not subject to mandatory PTE testing, posing a consumer safety risk. OBJECTIVE: This study presents a single-laboratory validation of an ICP OES method for the determination of As, Cd, Pb, and Hg in hemp, evaluated against AOAC SMPR 2020.001 performance criteria. METHODS: Ground hemp (0.125 g) was digested using HNO3/HCl/H2O2 (190 °C, 20 min) and diluted to 50 mL. Samples were analyzed on an Agilent 5800 ICP OES with a seven-point calibration (1-100 μg/L), Au/Sc internal standards, and KHP carbon for matrix matching. RESULTS: Spike recoveries for Cd (92.7%) and Pb (96.2%) met AOAC SMPR 2020.001 criteria (80-115%). Mercury recovery (47%) fell below acceptance limits due to matrix suppression and volatility losses. The As LOD (0.807 μg/g) exceeded sample concentrations in most specimens. LOQ values for all four elements exceeded the SMPR 2020.001 threshold of 10 µg/kg (0.01 µg/g). Calibration linearity was confirmed for all analytes (r ≥ 0.995). CONCLUSION: Partial validation was achieved: Cd and Pb demonstrated acceptable accuracy and repeatability, but LOQ criteria were not met for any analyte. ICP OES is not suitable for Hg or As at regulatory trace levels; ICP MS is recommended for these elements. This method offers a cost-effective screening tool for Cd and Pb in hemp. HIGHLIGHTS: This study presents the first single-laboratory ICP OES validation for PTEs in hemp under AOAC SMPR 2020.001. Cd and Pb met accuracy and repeatability criteria; As and Hg did not meet sensitivity requirements. ICP MS is recommended for regulatory As and Hg testing in hemp.

Journal of AOAC International
Bureau of Plant Industry (PH)
Openalex Percentile: Top 7%
Heavy Metal Exposure and Toxicity
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.