Physical characterization of a CBCT scanner for maxillo-facial intra-surgical navigation

<h2>Abstract</h2><h3>Purpose</h3> To characterize the image quality and dose level of a cone-beam CT (CBCT) scanner dedicated to maxillo-facial intra-surgical navigation. <h3>Methods</h3> Acquisition protocols for head imaging of average pediatric and adult patients were investigated. The detector response curve and the modulation transfer function (MTF) were assessed. The 3D image quality was evaluated using a commercial Catphan604 phantom in terms of accuracy of Hounsfield Unit (HU) estimates and slice uniformity. The MTF and the noise power spectrum (NPS) were assessed over the 3D reconstructed images. The weighted cone-beam dose index (CBDI<sub>w</sub>) was used for dosimetric benchmarking, and TLD chips placed in two anthropomorphic phantoms provided insight into the patient dose distribution. <h3>Results</h3> The detector exhibited a response curve with a double natural logarithmic behavior and spatial resolutions up to 6.3 mm<sup>−1</sup>. The 3D spatial resolution ranged from 0.70 mm<sup>−1</sup> at the scanner isocenter, for the pediatric protocol, to 0.46 mm<sup>−1</sup> at 7 cm from the isocenter for the adult protocol. Due to beam hardening and the scatter, discrepancies between expected and measured HU were observed, in particular for positive HU values (i.e. those for acrylic, bone, delrin and teflon inserts). The CBDI<sub>w</sub> and dose distribution within anthropomorphic phantoms provided a dosimetric benchmark for the scanner. <h3>Conclusion</h3> This study characterized a CBCT scanner in maxillo-facial surgery navigation. The findings support the definition of quality assurance protocols, providing methodologies and reference values for image quality and dose, while also highlighting limitations in spatial resolution and accuracy relevant for quantitative evaluations in clinical practice.

Authors

Institutions

Publication Details

Journal
Physica Medica
Published
2026-05-29
DOI
https://doi.org/10.1016/j.ejmp.2026.105818
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Physical characterization of a CBCT scanner for maxillo-facial intra-surgical navigation

Mauro Buono, Salvatore Grasso, Umberto Crimaldi, Carolina Mainardi et al.
Physica Medica
Dental Radiography and Imaging
article

Physical characterization of a CBCT scanner for maxillo-facial intra-surgical navigation

Mauro Buono, Salvatore Grasso, Umberto Crimaldi, Carolina Mainardi, Clelia Cotena, Caterina Oliviero, Roberta Reale, Roberto Moretti, Antonio Sarno, Daniele Manzi, Stefania Clemente, Antonio Varallo, Mariagabriella Pugliese
article en

Abstract

<h2>Abstract</h2><h3>Purpose</h3> To characterize the image quality and dose level of a cone-beam CT (CBCT) scanner dedicated to maxillo-facial intra-surgical navigation. <h3>Methods</h3> Acquisition protocols for head imaging of average pediatric and adult patients were investigated. The detector response curve and the modulation transfer function (MTF) were assessed. The 3D image quality was evaluated using a commercial Catphan604 phantom in terms of accuracy of Hounsfield Unit (HU) estimates and slice uniformity. The MTF and the noise power spectrum (NPS) were assessed over the 3D reconstructed images. The weighted cone-beam dose index (CBDI<sub>w</sub>) was used for dosimetric benchmarking, and TLD chips placed in two anthropomorphic phantoms provided insight into the patient dose distribution. <h3>Results</h3> The detector exhibited a response curve with a double natural logarithmic behavior and spatial resolutions up to 6.3 mm<sup>−1</sup>. The 3D spatial resolution ranged from 0.70 mm<sup>−1</sup> at the scanner isocenter, for the pediatric protocol, to 0.46 mm<sup>−1</sup> at 7 cm from the isocenter for the adult protocol. Due to beam hardening and the scatter, discrepancies between expected and measured HU were observed, in particular for positive HU values (i.e. those for acrylic, bone, delrin and teflon inserts). The CBDI<sub>w</sub> and dose distribution within anthropomorphic phantoms provided a dosimetric benchmark for the scanner. <h3>Conclusion</h3> This study characterized a CBCT scanner in maxillo-facial surgery navigation. The findings support the definition of quality assurance protocols, providing methodologies and reference values for image quality and dose, while also highlighting limitations in spatial resolution and accuracy relevant for quantitative evaluations in clinical practice.

Physica MedicaVol. 147
University of Milan (IT), Agostino Gemelli University Polyclinic (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Napoli (IT), Italian Society of Physiotherapy (IT), Federico II University Hospital (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Milano (IT), University of Naples Federico II (IT)
Openalex Percentile: Top 5%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.