Physical characterization of a CBCT scanner for maxillo-facial intra-surgical navigation
<h2>Abstract</h2><h3>Purpose</h3> To characterize the image quality and dose level of a cone-beam CT (CBCT) scanner dedicated to maxillo-facial intra-surgical navigation. <h3>Methods</h3> Acquisition protocols for head imaging of average pediatric and adult patients were investigated. The detector response curve and the modulation transfer function (MTF) were assessed. The 3D image quality was evaluated using a commercial Catphan604 phantom in terms of accuracy of Hounsfield Unit (HU) estimates and slice uniformity. The MTF and the noise power spectrum (NPS) were assessed over the 3D reconstructed images. The weighted cone-beam dose index (CBDI<sub>w</sub>) was used for dosimetric benchmarking, and TLD chips placed in two anthropomorphic phantoms provided insight into the patient dose distribution. <h3>Results</h3> The detector exhibited a response curve with a double natural logarithmic behavior and spatial resolutions up to 6.3 mm<sup>−1</sup>. The 3D spatial resolution ranged from 0.70 mm<sup>−1</sup> at the scanner isocenter, for the pediatric protocol, to 0.46 mm<sup>−1</sup> at 7 cm from the isocenter for the adult protocol. Due to beam hardening and the scatter, discrepancies between expected and measured HU were observed, in particular for positive HU values (i.e. those for acrylic, bone, delrin and teflon inserts). The CBDI<sub>w</sub> and dose distribution within anthropomorphic phantoms provided a dosimetric benchmark for the scanner. <h3>Conclusion</h3> This study characterized a CBCT scanner in maxillo-facial surgery navigation. The findings support the definition of quality assurance protocols, providing methodologies and reference values for image quality and dose, while also highlighting limitations in spatial resolution and accuracy relevant for quantitative evaluations in clinical practice.
Authors
- Mauro Buono (ORCID: https://orcid.org/0000-0003-1712-5849)
- Salvatore Grasso (ORCID: https://orcid.org/0000-0002-2753-4008)
- Umberto Crimaldi
- Carolina Mainardi
- Clelia Cotena (ORCID: https://orcid.org/0009-0001-4287-5644)
- Caterina Oliviero (ORCID: https://orcid.org/0009-0009-5024-8067)
- Roberta Reale
- Roberto Moretti (ORCID: https://orcid.org/0000-0002-5374-9045)
- Antonio Sarno (ORCID: https://orcid.org/0000-0002-3034-7166)
- Daniele Manzi
- Stefania Clemente
- Antonio Varallo (ORCID: https://orcid.org/0000-0001-6279-6719)
- Mariagabriella Pugliese
Institutions
- University of Milan (IT)
- Agostino Gemelli University Polyclinic (IT)
- Istituto Nazionale di Fisica Nucleare, Sezione di Napoli (IT)
- Italian Society of Physiotherapy (IT)
- Federico II University Hospital (IT)
- Istituto Nazionale di Fisica Nucleare, Sezione di Milano (IT)
- University of Naples Federico II (IT)
Publication Details
- Journal
- Physica Medica
- Published
- 2026-05-29
- DOI
- https://doi.org/10.1016/j.ejmp.2026.105818
- Primary Topic
- Dental Radiography and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00