Optimization of brain PET markerless head motion correction in a large human cohort

PURPOSE: Uncorrected head motion in brain PET degrades image resolution and biases quantification. Previously, we evaluated the United Imaging Healthcare markerless motion tracking (UMT) system for brain PET (Baseline UMT). This study provides an expanded validation in a large cohort compared to the Vicra with improved methodology (Optimized UMT). METHODS: We developed automatic processing to define the upper face region and added a 2-step Iterative Closest Points registration to reduce non-rigid motion effects. Optimized and Baseline UMT were first compared with simulations using measured motion and intentional facial expressions. In the human study, we evaluated 51 dynamic human studies (Siemens mCT) using 5 radiopharmaceuticals. Vicra data were used as the gold standard, after assessment of motion data quality using motion-corrected centroid-of-distribution. Comparisons between Optimized UMT, no motion correction (NMC), Post-Reconstruction Registration (PRR), and Baseline UMT were performed. Static and dynamic analyses assessed the differences of region-of-interest standardized uptake values (SUV) compared to Vicra. Time activity curve (TAC) analysis was performed to quantify uncorrected motion from the residual standard deviation (SD) around fitted TACs. RESULTS: In simulation studies, optimized UMT reduced error by 20% relative to NMC and by 9% relative to Baseline UMT under intentional facial expressions, reflecting the impact of the updated UMT registration. In the human static analysis of SUV, optimized UMT outperformed PRR (+ 1.6%) and NMC (+ 6.5%) on average grey matter difference. Similar results were found in the dynamic analysis. In the TAC analysis, optimized UMT consistently yielded lower residual variability than NMC and PRR across tracers and ROIs, with a narrower range of median %SD values (0.3-1.8%) compared with NMC (1.6-6.5%) and PRR (0.5-1.9%), while demonstrating performance comparable to the reference Vicra-based correction. CONCLUSION: Optimized UMT achieved comparable performance to Vicra and significantly improved quantification over PRR.

Authors

Institutions

Publication Details

Journal
EJNMMI Physics
Published
2026-05-29
DOI
https://doi.org/10.1186/s40658-026-00885-1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of brain PET markerless head motion correction in a large human cohort

J Zhang, Tianyi Zeng, Mika Naganawa, Tim Mulnix et al.
EJNMMI Physics
Medical Imaging Techniques and Applications
article

Optimization of brain PET markerless head motion correction in a large human cohort

J Zhang, Tianyi Zeng, Mika Naganawa, Tim Mulnix, Xishan Sun, Paul Gravel, Weize Jiang, Xinyue Zhang, Zhaohui Yang, Jean-Dominique Gallezot, Kathryn Fontaine, Richard E. Carson, Lingzhi Hu, Su Wang
article en

Abstract

PURPOSE: Uncorrected head motion in brain PET degrades image resolution and biases quantification. Previously, we evaluated the United Imaging Healthcare markerless motion tracking (UMT) system for brain PET (Baseline UMT). This study provides an expanded validation in a large cohort compared to the Vicra with improved methodology (Optimized UMT). METHODS: We developed automatic processing to define the upper face region and added a 2-step Iterative Closest Points registration to reduce non-rigid motion effects. Optimized and Baseline UMT were first compared with simulations using measured motion and intentional facial expressions. In the human study, we evaluated 51 dynamic human studies (Siemens mCT) using 5 radiopharmaceuticals. Vicra data were used as the gold standard, after assessment of motion data quality using motion-corrected centroid-of-distribution. Comparisons between Optimized UMT, no motion correction (NMC), Post-Reconstruction Registration (PRR), and Baseline UMT were performed. Static and dynamic analyses assessed the differences of region-of-interest standardized uptake values (SUV) compared to Vicra. Time activity curve (TAC) analysis was performed to quantify uncorrected motion from the residual standard deviation (SD) around fitted TACs. RESULTS: In simulation studies, optimized UMT reduced error by 20% relative to NMC and by 9% relative to Baseline UMT under intentional facial expressions, reflecting the impact of the updated UMT registration. In the human static analysis of SUV, optimized UMT outperformed PRR (+ 1.6%) and NMC (+ 6.5%) on average grey matter difference. Similar results were found in the dynamic analysis. In the TAC analysis, optimized UMT consistently yielded lower residual variability than NMC and PRR across tracers and ROIs, with a narrower range of median %SD values (0.3-1.8%) compared with NMC (1.6-6.5%) and PRR (0.5-1.9%), while demonstrating performance comparable to the reference Vicra-based correction. CONCLUSION: Optimized UMT achieved comparable performance to Vicra and significantly improved quantification over PRR.

EJNMMI Physics
Yale University (US), United Imaging Healthcare (China) (CN)
National Institutes of Health
Openalex Percentile: Top 5%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.