A Probabilistic Three-Dimensional Finite Element Model of a Cemented Hip Implant Failure Under Aseptic Loosening: A Case-Based Probabilistic Framework

Background: Hip implant fractures are rare, yet difficult to correct once they occur. For cemented implants, fracture is often associated with increased stresses at the implant stem when proximal regions of the implant have debonded. While deterministic analyses offer predictive power by using averages for model inputs, averages fail to capture the variability inherent in device manufacturing and musculoskeletal biology. This study developed a probabilistic finite element model of a debonded hip implant to better account for some of these variabilities to predict the most likely failure mode. The hypothesis was that fatigue would be more likely to occur than overloading. Methods and Materials: Monte Carlo sampling generated 1000 simulations varying the material elastic modulus (implant, cement, and bone) and loading magnitude at stance phase of the gait. The resultant distributions of maximum von Mises stress at the stem were compared to distributions for failure properties in the literature. Results: The analysis found the likelihood of the implant failing due to overloading was remote. In contrast, fatigue failure had a 99.4% chance of occurring. Fracture mechanics predicted that the debonded implant would reach critical flaw length between 1.8 and 26.4 months, with a mean of 7.2 months. Conclusions: The results show good agreement with the findings of the case study the model was based on, particularly in predicting the location of failure and fatigue life. The results of this study provide a framework for developing future decision-making tools that ultimately may assist clinicians in deciding when interventions are necessary to minimize the risk of implant or periprosthetic fracture.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-05-27
DOI
https://doi.org/10.3390/bioengineering13060623
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Probabilistic Three-Dimensional Finite Element Model of a Cemented Hip Implant Failure Under Aseptic Loosening: A Case-Based Probabilistic Framework

Lanny Griffin, Scott J. Hazelwood, Jonathan Fow, Daniel Truong
Bioengineering
Orthopaedic implants and arthroplasty
article

A Probabilistic Three-Dimensional Finite Element Model of a Cemented Hip Implant Failure Under Aseptic Loosening: A Case-Based Probabilistic Framework

Lanny Griffin, Scott J. Hazelwood, Jonathan Fow, Daniel Truong
article en

Abstract

Background: Hip implant fractures are rare, yet difficult to correct once they occur. For cemented implants, fracture is often associated with increased stresses at the implant stem when proximal regions of the implant have debonded. While deterministic analyses offer predictive power by using averages for model inputs, averages fail to capture the variability inherent in device manufacturing and musculoskeletal biology. This study developed a probabilistic finite element model of a debonded hip implant to better account for some of these variabilities to predict the most likely failure mode. The hypothesis was that fatigue would be more likely to occur than overloading. Methods and Materials: Monte Carlo sampling generated 1000 simulations varying the material elastic modulus (implant, cement, and bone) and loading magnitude at stance phase of the gait. The resultant distributions of maximum von Mises stress at the stem were compared to distributions for failure properties in the literature. Results: The analysis found the likelihood of the implant failing due to overloading was remote. In contrast, fatigue failure had a 99.4% chance of occurring. Fracture mechanics predicted that the debonded implant would reach critical flaw length between 1.8 and 26.4 months, with a mean of 7.2 months. Conclusions: The results show good agreement with the findings of the case study the model was based on, particularly in predicting the location of failure and fatigue life. The results of this study provide a framework for developing future decision-making tools that ultimately may assist clinicians in deciding when interventions are necessary to minimize the risk of implant or periprosthetic fracture.

BioengineeringVol. 13(6)
California Polytechnic State University (US), Central Coast Local Health District (AU)
Openalex Percentile: Top 5%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies