Development and Application of a Silicon Film Chamber for Enhanced Microscope‐Type <scp>X</scp> ‐Ray Fluorescence ( <scp>μXRF</scp> ) Analysis

ABSTRACT Micro‐scale X‐ray fluorescence (μXRF) is a versatile tool for non‐destructive elemental quantification across materials science, environmental monitoring, and industrial analysis. Nevertheless, quantitative robustness can deteriorate under short acquisition times due to limited photon statistics, geometric instability, and absorption effects. A physically grounded strategy that enhances quantitative robustness without prolonging acquisition time is therefore highly desirable. Here, we systematically investigate a film‐assisted sample configuration in a microscope‐type XRF (XGT) system, focusing on acquisition time (100 s vs. 400 s) and supporting film thickness (0.1–0.3 mm) as adjustable analytical parameters. Using NaCl and DOPC‐derived phosphorus as model systems, detection success rates and concentration distributions were statistically evaluated. Under photon‐limited conditions (100 s), the supporting film significantly improved Na detection success rate ( p &lt; 0.05) and reduced dispersion for low‐fluorescence‐yield phosphorus signals ( p &lt; 0.01). Thickness‐dependent effects were element‐specific and persisted for phosphorus even at extended acquisition time. The observed trends reflect the combined influence of Beer–Lambert attenuation and counting statistics, demonstrating that absorption alone cannot explain the experimental behavior; mechanical stabilization and signal detection success rate jointly determine quantitative stability. These findings demonstrate that supporting film thickness should be regarded as an active optimization parameter in μXRF rather than a passive structural component. The proposed approach enables acquisition time reduction while preserving quantitative precision and provides broadly applicable guidance for enhancing robustness in rapid micro‐scale XRF analysis.

Authors

Institutions

Publication Details

Journal
X-Ray Spectrometry
Published
2026-05-27
DOI
https://doi.org/10.1002/xrs.70121
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and Application of a Silicon Film Chamber for Enhanced Microscope‐Type X ‐Ray Fluorescence ( μXRF ) Analysis

K. Yokohama, Tsuyoshi Yoda
X-Ray Spectrometry
X-ray Spectroscopy and Fluorescence Analysis
article

Development and Application of a Silicon Film Chamber for Enhanced Microscope‐Type X ‐Ray Fluorescence ( μXRF ) Analysis

K. Yokohama, Tsuyoshi Yoda
article en

Abstract

ABSTRACT Micro‐scale X‐ray fluorescence (μXRF) is a versatile tool for non‐destructive elemental quantification across materials science, environmental monitoring, and industrial analysis. Nevertheless, quantitative robustness can deteriorate under short acquisition times due to limited photon statistics, geometric instability, and absorption effects. A physically grounded strategy that enhances quantitative robustness without prolonging acquisition time is therefore highly desirable. Here, we systematically investigate a film‐assisted sample configuration in a microscope‐type XRF (XGT) system, focusing on acquisition time (100 s vs. 400 s) and supporting film thickness (0.1–0.3 mm) as adjustable analytical parameters. Using NaCl and DOPC‐derived phosphorus as model systems, detection success rates and concentration distributions were statistically evaluated. Under photon‐limited conditions (100 s), the supporting film significantly improved Na detection success rate ( p &lt; 0.05) and reduced dispersion for low‐fluorescence‐yield phosphorus signals ( p &lt; 0.01). Thickness‐dependent effects were element‐specific and persisted for phosphorus even at extended acquisition time. The observed trends reflect the combined influence of Beer–Lambert attenuation and counting statistics, demonstrating that absorption alone cannot explain the experimental behavior; mechanical stabilization and signal detection success rate jointly determine quantitative stability. These findings demonstrate that supporting film thickness should be regarded as an active optimization parameter in μXRF rather than a passive structural component. The proposed approach enables acquisition time reduction while preserving quantitative precision and provides broadly applicable guidance for enhancing robustness in rapid micro‐scale XRF analysis.

X-Ray Spectrometry
Iwate University (JP), Aomori Prefectural Industrial Technology Center (JP)
Openalex Percentile: Top 6%
X-ray Spectroscopy and Fluorescence Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.