Optimizing line blot assays: impact of band intensity adjustment on diagnostic performance in myositis autoantibodies testing

INTRODUCTION: Line blot assays (LBA) have several limitations regarding diagnostic validity, as sensitivity and specificity vary considerably among antigens. OBJECTIVE: The aim of this study was to assess whether the adjustment of each specific band intensity (BI) relative to the positive control band (PCB), combined with increasing the cut-off value from 15 to 20, improves the diagnostic performance of the LBA for detecting autoantibodies related to idiopathic inflammatory myopathies in our patient cohort. METHODS: We included all serum samples tested for myositis‑specific antibodies (MSA) and myositis‑associated antibodies (MAA) between January 2022 and April 2023 at the Immunology Department of Hospital Clínic of Barcelona. Two analyses were performed: Analysis A, using the manufacturer's recommended cut-off (BI ≥ 15), and Analysis B, applying the proposed BI adjustment relative to the PCB and raising the cut-off to 20 in samples that were positive in Analysis A. RESULTS: A total of 939 patients were included. Using Analysis A, 280/939 (29.8%) patients had at least one positive result, whereas 659/939 (70.2%) were negative for all MSA and MAA. After applying Analysis B, 151/280 (53.9%) remained positive, while 129/280 (46.1%) became negative. With Analysis A, sensitivity was 76%, specificity 75%, positive predictive value (PPV) 24%, and negative predictive value (NPV) 96%. Under Analysis B, sensitivity was 70%, specificity 89%, PPV 41%, and NPV 96%. CONCLUSIONS: Applying the proposed BI adjustment effectively removes weakly positive results with low specificity in an objective and reproducible manner, thereby increasing the overall specificity of the LBA. Key Points • Applying the proposed band intensity removes weakly positive results. • This approach enhances the specificity, and positive predictive value of a commercial line blot assay.

Authors

Institutions

Publication Details

Journal
Clinical Rheumatology
Published
2026-05-27
DOI
https://doi.org/10.1007/s10067-026-08153-4
Primary Topic
Inflammatory Myopathies and Dermatomyositis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing line blot assays: impact of band intensity adjustment on diagnostic performance in myositis autoantibodies testing

Albert Pérez-Isidro, Noemí de Moner, Maria Torradeflot, S Prieto-Gonzalez et al.
Clinical Rheumatology
Inflammatory Myopathies and Dermatomyositis
article

Optimizing line blot assays: impact of band intensity adjustment on diagnostic performance in myositis autoantibodies testing

Albert Pérez-Isidro, Noemí de Moner, Maria Torradeflot, S Prieto-Gonzalez, F Hernandez-Gonzalez, Estíbaliz Ruiz-Ortiz, Francisco Javier Hüttmann, María Ortiz-Fernández
article en

Abstract

INTRODUCTION: Line blot assays (LBA) have several limitations regarding diagnostic validity, as sensitivity and specificity vary considerably among antigens. OBJECTIVE: The aim of this study was to assess whether the adjustment of each specific band intensity (BI) relative to the positive control band (PCB), combined with increasing the cut-off value from 15 to 20, improves the diagnostic performance of the LBA for detecting autoantibodies related to idiopathic inflammatory myopathies in our patient cohort. METHODS: We included all serum samples tested for myositis‑specific antibodies (MSA) and myositis‑associated antibodies (MAA) between January 2022 and April 2023 at the Immunology Department of Hospital Clínic of Barcelona. Two analyses were performed: Analysis A, using the manufacturer's recommended cut-off (BI ≥ 15), and Analysis B, applying the proposed BI adjustment relative to the PCB and raising the cut-off to 20 in samples that were positive in Analysis A. RESULTS: A total of 939 patients were included. Using Analysis A, 280/939 (29.8%) patients had at least one positive result, whereas 659/939 (70.2%) were negative for all MSA and MAA. After applying Analysis B, 151/280 (53.9%) remained positive, while 129/280 (46.1%) became negative. With Analysis A, sensitivity was 76%, specificity 75%, positive predictive value (PPV) 24%, and negative predictive value (NPV) 96%. Under Analysis B, sensitivity was 70%, specificity 89%, PPV 41%, and NPV 96%. CONCLUSIONS: Applying the proposed BI adjustment effectively removes weakly positive results with low specificity in an objective and reproducible manner, thereby increasing the overall specificity of the LBA. Key Points • Applying the proposed band intensity removes weakly positive results. • This approach enhances the specificity, and positive predictive value of a commercial line blot assay.

Clinical Rheumatology
Hospital Clínic de Barcelona (ES), Consorci Institut D'Investigacions Biomediques August Pi I Sunyer (ES), Universitat de Barcelona (ES)
Openalex Percentile: Top 6%
Inflammatory Myopathies and Dermatomyositis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.