Validation of absolute dose calculation of a single 125I-seed for ophthalmic brachytherapy within distance range 1–15 mm using a diode detector

BACKGROUND AND PURPOSE: Intraocular tumors can be treated with brachytherapy. Dosimetry of ophthalmic applicators is challenging due to short distance ranges and steep dose gradients. Measurements are sensitive to the structure of the dosimeter, and correction methods specific to 125I-source are needed for validation measurements of dose distribution. MATERIAL AND METHODS: We measured dose profiles of I25.S16 seed using diode detector in a water phantom at 2 and 5 mm above surface of acrylic (PMMA) and gold backing. Volume averaging effect was corrected by the ratio of dose calculated with AAPM TG-43 formalism at the center point and the dose average within the detector element area. The angular sensitivity was measured at a constant radius around the seed. The effect of gold backing was evaluated from measurements with and without gold slab behind the seed. Measurements were normalized to calculated value at the distance of 10 mm from the seed using the air-kerma rate measured with a calibrated well chamber. RESULTS: 1D/3D gamma comparison pass rates 97% with criteria 0.3 mm/3% and 60% with 0.5 mm/5% were reached for a seed attached to PMMA and gold, respectively, in the range ± 10 mm. INTERPRETATION: TG-43 calculated single seed distributions in PMMA were equivalent to measurements corrected for angular sensitivity and detector area. Gold-to-water ratio corrected calculations were not similarly comparable, and simple one-dimensional correction method is not accurate enough laterally. Diode dosimetry with these correction methods can be used for quality assurance and validation of clinical dose calculation of ophthalmic applicators.

Authors

Institutions

Publication Details

Journal
Acta Oncologica
Published
2026-05-27
DOI
https://doi.org/10.2340/1651-226x.2026.45688
Primary Topic
Ocular Oncology and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of absolute dose calculation of a single 125I-seed for ophthalmic brachytherapy within distance range 1–15 mm using a diode detector

Assi Valve, Anna Rintala, Vappu Reijonen, Mikko Tenhunen
Acta Oncologica
Ocular Oncology and Treatments
article

Validation of absolute dose calculation of a single 125I-seed for ophthalmic brachytherapy within distance range 1–15 mm using a diode detector

Assi Valve, Anna Rintala, Vappu Reijonen, Mikko Tenhunen
article en

Abstract

BACKGROUND AND PURPOSE: Intraocular tumors can be treated with brachytherapy. Dosimetry of ophthalmic applicators is challenging due to short distance ranges and steep dose gradients. Measurements are sensitive to the structure of the dosimeter, and correction methods specific to 125I-source are needed for validation measurements of dose distribution. MATERIAL AND METHODS: We measured dose profiles of I25.S16 seed using diode detector in a water phantom at 2 and 5 mm above surface of acrylic (PMMA) and gold backing. Volume averaging effect was corrected by the ratio of dose calculated with AAPM TG-43 formalism at the center point and the dose average within the detector element area. The angular sensitivity was measured at a constant radius around the seed. The effect of gold backing was evaluated from measurements with and without gold slab behind the seed. Measurements were normalized to calculated value at the distance of 10 mm from the seed using the air-kerma rate measured with a calibrated well chamber. RESULTS: 1D/3D gamma comparison pass rates 97% with criteria 0.3 mm/3% and 60% with 0.5 mm/5% were reached for a seed attached to PMMA and gold, respectively, in the range ± 10 mm. INTERPRETATION: TG-43 calculated single seed distributions in PMMA were equivalent to measurements corrected for angular sensitivity and detector area. Gold-to-water ratio corrected calculations were not similarly comparable, and simple one-dimensional correction method is not accurate enough laterally. Diode dosimetry with these correction methods can be used for quality assurance and validation of clinical dose calculation of ophthalmic applicators.

Acta OncologicaVol. 65
University of Helsinki (FI), Helsinki University Hospital (FI), Hospital District of Helsinki and Uusimaa (FI)
Openalex Percentile: Top 5%
Ocular Oncology and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.