Two Doppler blood pressure devices and 2 oscillometric blood pressure devices failed American College of Veterinary Internal Medicine validation in conscious dogs

Objective: To compare the accuracy of 2 Doppler and 2 oscillometric devices to invasive blood pressure (IBP) measurements in conscious dogs at 2 cuff locations using the American College of Veterinary Internal Medicine (ACVIM) validation criteria. Methods: 8 Beagles had dorsal pedal arterial catheters placed under anesthesia on November 13 and 18, 2024. Once conscious, paired non-IBP and IBP measurements were collected at the tail and forelimb for all devices. Bias, SD, percentage of values within 10 and 20 mm Hg of IBP, correlation coefficients, and limits of agreement were calculated for each paired measurement. Results: No device met ACVIM validation criteria at either cuff location. Devices performed better at the tail compared to the forelimb. Doppler devices performed better than oscillometric devices for systolic blood pressure. Oscillometric devices performed best for mean blood pressure (MBP). The MBP measured by the multiparameter monitor at the tail showed the most favorable overall performance, including a bias of 1.75 mm Hg (SD, 8.98); 82% and 96% of measurements within 10 and 20 mm Hg of IBP, respectively; and limits of agreement of -15 to 19. The correlation coefficient threshold was not met by any device. Conclusions: None of the devices were validated against IBP in conscious adult dogs using ACVIM criteria. Tail placement improved performance across devices. Oscillometric devices appeared more reliable for MBP, whereas Doppler devices provided more reliable systolic blood pressure measurements. Clinical Relevance: Clinicians should interpret NIBP cautiously in conscious dogs.

Authors

Institutions

Publication Details

Journal
American Journal of Veterinary Research
Published
2026-05-27
DOI
https://doi.org/10.2460/ajvr.26.03.0108
Primary Topic
Cardiovascular Conditions and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Two Doppler blood pressure devices and 2 oscillometric blood pressure devices failed American College of Veterinary Internal Medicine validation in conscious dogs

Kaylee A. Quinn, Kristin M. Zersen, Julien Guillaumin, Casey Lorge
American Journal of Veterinary Research
Cardiovascular Conditions and Treatments
article

Two Doppler blood pressure devices and 2 oscillometric blood pressure devices failed American College of Veterinary Internal Medicine validation in conscious dogs

Kaylee A. Quinn, Kristin M. Zersen, Julien Guillaumin, Casey Lorge
article en

Abstract

Objective: To compare the accuracy of 2 Doppler and 2 oscillometric devices to invasive blood pressure (IBP) measurements in conscious dogs at 2 cuff locations using the American College of Veterinary Internal Medicine (ACVIM) validation criteria. Methods: 8 Beagles had dorsal pedal arterial catheters placed under anesthesia on November 13 and 18, 2024. Once conscious, paired non-IBP and IBP measurements were collected at the tail and forelimb for all devices. Bias, SD, percentage of values within 10 and 20 mm Hg of IBP, correlation coefficients, and limits of agreement were calculated for each paired measurement. Results: No device met ACVIM validation criteria at either cuff location. Devices performed better at the tail compared to the forelimb. Doppler devices performed better than oscillometric devices for systolic blood pressure. Oscillometric devices performed best for mean blood pressure (MBP). The MBP measured by the multiparameter monitor at the tail showed the most favorable overall performance, including a bias of 1.75 mm Hg (SD, 8.98); 82% and 96% of measurements within 10 and 20 mm Hg of IBP, respectively; and limits of agreement of -15 to 19. The correlation coefficient threshold was not met by any device. Conclusions: None of the devices were validated against IBP in conscious adult dogs using ACVIM criteria. Tail placement improved performance across devices. Oscillometric devices appeared more reliable for MBP, whereas Doppler devices provided more reliable systolic blood pressure measurements. Clinical Relevance: Clinicians should interpret NIBP cautiously in conscious dogs.

American Journal of Veterinary Research
Colorado State University (US)
Good health and well-being
Openalex Percentile: Top 6%
Cardiovascular Conditions and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.