Analytical validation of a plasma-based cfDNA NGS assay (Guardant360 Liquid CDx) for comprehensive solid tumor profiling.

3077 Background: NGS-based profiling of circulating cell-free DNA (cfDNA) enables accurate and sensitive tumor genotyping from blood. Guardant360 Liquid CDx is intended as a companion diagnostic for comprehensive tumor mutation profiling across >700 genes in patients with malignant neoplasms. Analytical performance was evaluated per FDA IVD standards and guidelines across five variant classes: single-nucleotide variants (SNVs), insertions/deletions (indels), gene rearrangements, gene amplifications (CNA), and copy-number losses (CNLs) – with emphasis on accuracy, precision, specificity (limit of blank [LoB]), and limit of detection (LoD). Methods: Analytical Validation tested >4000 samples from nearly 2,000 unique patients across 29 cancer types, as well as contrived materials. Concordance was evaluated against FDA-approved Guardant360 CDx and an externally developed ctDNA assay, where applicable. Primary endpoints were positive percent agreement (PPA) and negative percent agreement (NPA) by variant class and category (clinically significant vs panel-wide). LoD was established at both low and high input levels. No minimum cfDNA ng input cutoff was required; low-input performance was evaluated at minimum molecule coverage (QC threshold). Precision was assessed across workflow components and blood collection tube (BCT) lots. Specificity was assessed using healthy donor cfDNA sequenced with multiple replicates at high input. Results: Clinical samples demonstrated a high QC pass rate of 99.1%. Clinically significant variants showed high concordance: SNV PPA 94.9% with NPA 99.86%, and indel PPA 96.24% with NPA 99.98%. Panel-wide specificity was near-perfect (SNV NPA 99.99966%; indel NPA 99.99991%). Rearrangements achieved PPA 94.64% and NPA 99.76%; CNA PPA 91.67% and NPA 99.97% ; CNL PPA 100% and NPA 97.56%. High-input LoD was established in silico and confirmed with clinical samples: 0.2% mutant allele frequency (MAF) for all clinically relevant SNVs, indels and most rearrangements, with selected variant LoD as low as 0.1% (KRAS G12C). At the minimum input, panel-wide LoD was 1.0% for clinically relevant SNVs and 0.9% for indels; 0.5-1.6% for rearrangements, 2.3-2.4 copies for CNAs, and 22.7% tumor fraction for CNL. Precision studies showed ≥98.6% agreement across conditions at near-LoD variant levels; within-lot and between-lot BCT precision yielded >97% analytical positive/negative agreement (targeted variants ≥1× LoD). In LoB, no false positive variants were detected in any healthy donor replicates (0% FPR per sample and per variant), validating in-silico CHIP detection methods. Conclusions: The assay demonstrates high accuracy, precision and specificity across all reportable variant classes, with LoD at or below 0.2% MAF for key variant types. These data support its use for comprehensive, pan-cancer tumor genomic profiling in routine oncology practice.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Oncology
Published
2026-05-27
DOI
https://doi.org/10.1200/jco.2026.44.16_suppl.3077
Primary Topic
Cancer Genomics and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical validation of a plasma-based cfDNA NGS assay (Guardant360 Liquid CDx) for comprehensive solid tumor profiling.

Raghu Chandramohan, Brett Kennedy, Krishnalekha Datta, Darya Chudova et al.
Journal of Clinical Oncology
Cancer Genomics and Diagnostics
article

Analytical validation of a plasma-based cfDNA NGS assay (Guardant360 Liquid CDx) for comprehensive solid tumor profiling.

Raghu Chandramohan, Brett Kennedy, Krishnalekha Datta, Darya Chudova, Ankit Jambusaria, Nageswara Alla, Soni Shukla, Paul Spurgeon, Kevin Doubleday, Sophie Weiss, Sammy Rantisi, Karen Ryall, Tianran Jia, Pooja Bhat, Xianxian Liu, Claire Olson, Jessica Kurata, Gowtham Jayakumaran, Priyanka Lakkaraju
article en

Abstract

3077 Background: NGS-based profiling of circulating cell-free DNA (cfDNA) enables accurate and sensitive tumor genotyping from blood. Guardant360 Liquid CDx is intended as a companion diagnostic for comprehensive tumor mutation profiling across >700 genes in patients with malignant neoplasms. Analytical performance was evaluated per FDA IVD standards and guidelines across five variant classes: single-nucleotide variants (SNVs), insertions/deletions (indels), gene rearrangements, gene amplifications (CNA), and copy-number losses (CNLs) – with emphasis on accuracy, precision, specificity (limit of blank [LoB]), and limit of detection (LoD). Methods: Analytical Validation tested >4000 samples from nearly 2,000 unique patients across 29 cancer types, as well as contrived materials. Concordance was evaluated against FDA-approved Guardant360 CDx and an externally developed ctDNA assay, where applicable. Primary endpoints were positive percent agreement (PPA) and negative percent agreement (NPA) by variant class and category (clinically significant vs panel-wide). LoD was established at both low and high input levels. No minimum cfDNA ng input cutoff was required; low-input performance was evaluated at minimum molecule coverage (QC threshold). Precision was assessed across workflow components and blood collection tube (BCT) lots. Specificity was assessed using healthy donor cfDNA sequenced with multiple replicates at high input. Results: Clinical samples demonstrated a high QC pass rate of 99.1%. Clinically significant variants showed high concordance: SNV PPA 94.9% with NPA 99.86%, and indel PPA 96.24% with NPA 99.98%. Panel-wide specificity was near-perfect (SNV NPA 99.99966%; indel NPA 99.99991%). Rearrangements achieved PPA 94.64% and NPA 99.76%; CNA PPA 91.67% and NPA 99.97% ; CNL PPA 100% and NPA 97.56%. High-input LoD was established in silico and confirmed with clinical samples: 0.2% mutant allele frequency (MAF) for all clinically relevant SNVs, indels and most rearrangements, with selected variant LoD as low as 0.1% (KRAS G12C). At the minimum input, panel-wide LoD was 1.0% for clinically relevant SNVs and 0.9% for indels; 0.5-1.6% for rearrangements, 2.3-2.4 copies for CNAs, and 22.7% tumor fraction for CNL. Precision studies showed ≥98.6% agreement across conditions at near-LoD variant levels; within-lot and between-lot BCT precision yielded >97% analytical positive/negative agreement (targeted variants ≥1× LoD). In LoB, no false positive variants were detected in any healthy donor replicates (0% FPR per sample and per variant), validating in-silico CHIP detection methods. Conclusions: The assay demonstrates high accuracy, precision and specificity across all reportable variant classes, with LoD at or below 0.2% MAF for key variant types. These data support its use for comprehensive, pan-cancer tumor genomic profiling in routine oncology practice.

Journal of Clinical OncologyVol. 44(16_suppl)
Guardant (United States) (US)
Openalex Percentile: Top 12%
Cancer Genomics and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.