Effect of a Spinal Orthosis With Leaf Spring and Programmable Local Vibration on Kyphosis Angle and Trunk Muscle Strength in Hyperkyphosis: Development and Feasibility Study.

Background: The application of spinal orthoses and local vibration in the treatment of hyperkyphosis is a developing concept. Further research is needed to determine the impact of leaf springs and local vibration techniques on the neuromusculoskeletal system. Objective: This study aimed to evaluate the effect of the simultaneous application of a leaf spring and a programmed local vibration system in a spinal orthosis on thoracic kyphosis angle (TKA) and the function of trunk flexor and extensor muscles. Methods: We used a soft thoracolumbar orthosis as a base for 2 parallel leaf spring bars made from AISI (American Iron and Steel Institute) 1075 steel (45 cm × 2 cm × 2 mm each). We added 4 solenoid actuators (60 g each) housed in 2 units positioned 12 cm apart, generating 20 N vertical force at 8 Hz frequency with 10 mm free displacement. A programmable Arduino microcontroller delivered 10-second vibration bursts followed by 5-minute rest intervals, repeated for 12 cycles per session, applied every other day. One participant with hyperkyphosis (baseline TKA=47.24°) used this device for 4 weeks. Outcomes included TKA (measured by photogrammetry) and trunk muscle function (assessed under isometric, isotonic, and isokinetic conditions). Feasibility outcomes included adherence, wear time, and adverse events. Results: Bench testing confirmed an 8 Hz vibration frequency, a 10 mm free displacement, and an on-body contact pressure of 9.5 N/cm² (95 kPa). In the short-term assessment (N=10), mean acceptability scores were 7.95 (SD 0.83) for appearance and 7.6 (SD 0.07) for comfort, and no skin-related adverse events occurred. In the 4-week single-participant pilot (N=1), adherence was 100% (12/12 sessions completed), mean daily wear time was 1.4 (SD 0.3) hours, and no adverse events were reported. TKA decreased from 47.24° to 45.81° (change: -1.43°), which did not exceed the minimal detectable change (4.62°). Pre-post changes in muscle function outcomes are reported descriptively. Conclusions: A semirigid thoracolumbar orthosis with integrated programmable vibration (8 Hz, 20 N free force) is technically feasible, showing reliable bench performance. Short-term assessment in 10 participants indicated acceptable comfort and appearance with no immediate safety concerns. The 4-week single-participant pilot demonstrated 100% adherence and no adverse events. The observed 1.43° kyphosis reduction was not clinically meaningful (below minimal detectable change). Larger controlled studies are needed to determine whether this device improves muscle function or spinal alignment.

Authors

Institutions

Publication Details

Journal
PubMed
Published
2026-06-25
DOI
https://doi.org/10.2196/82013
Primary Topic
Scoliosis diagnosis and treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of a Spinal Orthosis With Leaf Spring and Programmable Local Vibration on Kyphosis Angle and Trunk Muscle Strength in Hyperkyphosis: Development and Feasibility Study.

Saeed Behzadipour, Fatemeh Keshavarzi, Mokhtar Arazpour
PubMed
Scoliosis diagnosis and treatment
article

Effect of a Spinal Orthosis With Leaf Spring and Programmable Local Vibration on Kyphosis Angle and Trunk Muscle Strength in Hyperkyphosis: Development and Feasibility Study.

Saeed Behzadipour, Fatemeh Keshavarzi, Mokhtar Arazpour
article en

Abstract

Background: The application of spinal orthoses and local vibration in the treatment of hyperkyphosis is a developing concept. Further research is needed to determine the impact of leaf springs and local vibration techniques on the neuromusculoskeletal system. Objective: This study aimed to evaluate the effect of the simultaneous application of a leaf spring and a programmed local vibration system in a spinal orthosis on thoracic kyphosis angle (TKA) and the function of trunk flexor and extensor muscles. Methods: We used a soft thoracolumbar orthosis as a base for 2 parallel leaf spring bars made from AISI (American Iron and Steel Institute) 1075 steel (45 cm × 2 cm × 2 mm each). We added 4 solenoid actuators (60 g each) housed in 2 units positioned 12 cm apart, generating 20 N vertical force at 8 Hz frequency with 10 mm free displacement. A programmable Arduino microcontroller delivered 10-second vibration bursts followed by 5-minute rest intervals, repeated for 12 cycles per session, applied every other day. One participant with hyperkyphosis (baseline TKA=47.24°) used this device for 4 weeks. Outcomes included TKA (measured by photogrammetry) and trunk muscle function (assessed under isometric, isotonic, and isokinetic conditions). Feasibility outcomes included adherence, wear time, and adverse events. Results: Bench testing confirmed an 8 Hz vibration frequency, a 10 mm free displacement, and an on-body contact pressure of 9.5 N/cm² (95 kPa). In the short-term assessment (N=10), mean acceptability scores were 7.95 (SD 0.83) for appearance and 7.6 (SD 0.07) for comfort, and no skin-related adverse events occurred. In the 4-week single-participant pilot (N=1), adherence was 100% (12/12 sessions completed), mean daily wear time was 1.4 (SD 0.3) hours, and no adverse events were reported. TKA decreased from 47.24° to 45.81° (change: -1.43°), which did not exceed the minimal detectable change (4.62°). Pre-post changes in muscle function outcomes are reported descriptively. Conclusions: A semirigid thoracolumbar orthosis with integrated programmable vibration (8 Hz, 20 N free force) is technically feasible, showing reliable bench performance. Short-term assessment in 10 participants indicated acceptable comfort and appearance with no immediate safety concerns. The 4-week single-participant pilot demonstrated 100% adherence and no adverse events. The observed 1.43° kyphosis reduction was not clinically meaningful (below minimal detectable change). Larger controlled studies are needed to determine whether this device improves muscle function or spinal alignment.

PubMedVol. 11
Sharif University of Technology (IR), University of Social Welfare and Rehabilitation Sciences (IR)
Openalex Percentile: Top 25%
Scoliosis diagnosis and treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.