Accuracy of a Continuous Glucose Monitoring System Under Hyperglycemic Challenge in Chinese Hospitalized Patients With Diabetes

BACKGROUND: To assess the accuracy of a continuous glucose monitoring (CGM) system under hyperglycemic conditions, we conducted a steamed bread meal tolerance test (SBMTT) in hospitalized patients aged 18 years or older with type 1and 2 diabetes mellitus. METHODS: Hospitalized adults with diabetes wore a Sibionics CGM (GS1) sensor, which was placed at least 48 hours prior to the test. Participants subsequently underwent an SBMTT containing 75 g of carbohydrates. Plasma glucose (PG) was measured at 0, 30, 60, 120, and 180 minutes. Accuracy was evaluated by calculating the mean absolute relative difference (MARD), %15/15, %20/20, %30/30, and %40/40 agreement rates. Clinical accuracy was assessed using the Diabetes Technology Society (DTS) error grid analysis. The time lag between CGM and PG was calculated. RESULTS: A total of 580 matched CGM-PG pairs from 116 participants were analyzed. The overall MARD was 11.83%, with agreement rates of 69.5% (%15/15), 86.2% (%20/20), 98.3% (%30/30), and 99.7% (%40/40). Diabetes Technology Society error grid analysis showed 100% of the CGM points falling into zones A and B, with 86% in zone A and 14% in zone B. After minimizing MARD, the overall CGM lag time was 5 minutes, but it increased during rapid glucose rise (0-60 minutes), reaching 10 minutes at 30 minutes and 20 minutes at 60 minutes. CONCLUSIONS: Continuous glucose monitoring maintained good accuracy during hyperglycemic challenges in hospitalized patients with diabetes. The time lag was influenced by the rate of glucose change, showing prolongation during the rapid glucose rise phase (0-60 minutes).

Authors

Institutions

Publication Details

Journal
Journal of Diabetes Science and Technology
Published
2026-05-23
DOI
https://doi.org/10.1177/19322968261442573
Primary Topic
Hyperglycemia and glycemic control in critically ill and hospitalized patients
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of a Continuous Glucose Monitoring System Under Hyperglycemic Challenge in Chinese Hospitalized Patients With Diabetes

Xiaolong Jin, Zhenhua Tian, F-Y Xu, Dongmei Zheng et al.
Journal of Diabetes Science and Technology
Hyperglycemia and glycemic control in critically ill and hospitalized patients
article

Accuracy of a Continuous Glucose Monitoring System Under Hyperglycemic Challenge in Chinese Hospitalized Patients With Diabetes

Xiaolong Jin, Zhenhua Tian, F-Y Xu, Dongmei Zheng, Xingmei Liu, Anqi Dai, XiujuanZhang MD, ChunxiaoYu PhD, Yachao Yang, Qingbo Guan, Chao Xu
article en

Abstract

BACKGROUND: To assess the accuracy of a continuous glucose monitoring (CGM) system under hyperglycemic conditions, we conducted a steamed bread meal tolerance test (SBMTT) in hospitalized patients aged 18 years or older with type 1and 2 diabetes mellitus. METHODS: Hospitalized adults with diabetes wore a Sibionics CGM (GS1) sensor, which was placed at least 48 hours prior to the test. Participants subsequently underwent an SBMTT containing 75 g of carbohydrates. Plasma glucose (PG) was measured at 0, 30, 60, 120, and 180 minutes. Accuracy was evaluated by calculating the mean absolute relative difference (MARD), %15/15, %20/20, %30/30, and %40/40 agreement rates. Clinical accuracy was assessed using the Diabetes Technology Society (DTS) error grid analysis. The time lag between CGM and PG was calculated. RESULTS: A total of 580 matched CGM-PG pairs from 116 participants were analyzed. The overall MARD was 11.83%, with agreement rates of 69.5% (%15/15), 86.2% (%20/20), 98.3% (%30/30), and 99.7% (%40/40). Diabetes Technology Society error grid analysis showed 100% of the CGM points falling into zones A and B, with 86% in zone A and 14% in zone B. After minimizing MARD, the overall CGM lag time was 5 minutes, but it increased during rapid glucose rise (0-60 minutes), reaching 10 minutes at 30 minutes and 20 minutes at 60 minutes. CONCLUSIONS: Continuous glucose monitoring maintained good accuracy during hyperglycemic challenges in hospitalized patients with diabetes. The time lag was influenced by the rate of glucose change, showing prolongation during the rapid glucose rise phase (0-60 minutes).

Journal of Diabetes Science and Technology
Shandong University (CN), Shandong Provincial Hospital (CN), Ministry of Education (TH), Weihai Municipal Hospital (CN), Tengzhou Central People's Hospital (CN), Shanghai Clinical Research Center (CN), Shandong First Medical University (CN), Shandong Academy of Chinese Medicine (CN)
National Natural Science Foundation of China
Openalex Percentile: Top 5%
Hyperglycemia and glycemic control in critically ill and hospitalized patients
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.