The Effect of <scp>PTFE</scp> Coating and <scp>PTFE</scp> Tape Wrapping Abutment Screw After Cyclic Loading: An In Vitro Study

OBJECTIVE: This study aimed to compare the effect of PTFE tape wrapping and PTFE-coated titanium alloy screws on the removal torque after cyclic loading and on preload values under repetitive tightening. METHODS: A total of 60 titanium abutment screws were allocated to three surface-treatment groups (n = 20 each): untreated (Control), PTFE tape-wrapped (Wrap), and PTFE-coated (Coat). Within each group, 10 specimens underwent cyclic loading, whereas 10 were assigned to preload evaluation. Cyclic loading followed ISO 14801:2016. Screws were tightened to 30 N·cm, retightened after 10 min, and subjected to 300 000 cycles at 200 N using a fatigue testing machine (DYNA-MESS Prüfsysteme GmbH, Aachen, Germany). Removal torque was measured with a digital torque gauge (BTGE50CN, Tohnichi, Japan). For preload assessment, each specimen completed 10 tightening-loosening cycles at the same torque during which tensile force was recorded using a customized device. One representative screw per group was analyzed by SEM and micro-CT before and after loading to evaluate surface and structural changes. RESULTS: After cyclic loading, the removal torque did not differ significantly between the Control group (17.58 ± 1.85 N·cm) and the Wrap group (17.50 ± 1.77 N·cm), whereas the Coat group demonstrated significantly lower removal torque (14.71 ± 2.35 N·cm; p < 0.05). In the preload test, preload progressively decreased in the Control (-3.13 N per cycle) and Coat groups (-0.51 N per cycle), while the Wrap group showed a consistent increase with repeated tightening (+8.22 N per cycle). SEM revealed surface wear and partial loss of PTFE material after loading, particularly in the coated specimens. Micro-CT analysis showed no detectable differences in internal adaptation among groups before or after cyclic loading. CONCLUSIONS: PTFE tape wrapping increased preload during repeated tightening but did not improve removal torque after cyclic loading compared with those of untreated screws. Although PTFE-coated screws produced the highest initial preload, they were associated with reduced removal torque after loading. SEM analysis confirmed coating deterioration, whereas internal fit remained comparable across groups.

Authors

Institutions

Publication Details

Journal
Clinical Implant Dentistry and Related Research
Published
2026-05-22
DOI
https://doi.org/10.1111/cid.70162
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Effect of PTFE Coating and PTFE Tape Wrapping Abutment Screw After Cyclic Loading: An In Vitro Study

Pathawee Khongkhunthian, Weerapan Aunmeungtong, Poonnarat Kongkhuntian
Clinical Implant Dentistry and Related Research
Orthopaedic implants and arthroplasty
article

The Effect of PTFE Coating and PTFE Tape Wrapping Abutment Screw After Cyclic Loading: An In Vitro Study

Pathawee Khongkhunthian, Weerapan Aunmeungtong, Poonnarat Kongkhuntian
article en

Abstract

OBJECTIVE: This study aimed to compare the effect of PTFE tape wrapping and PTFE-coated titanium alloy screws on the removal torque after cyclic loading and on preload values under repetitive tightening. METHODS: A total of 60 titanium abutment screws were allocated to three surface-treatment groups (n = 20 each): untreated (Control), PTFE tape-wrapped (Wrap), and PTFE-coated (Coat). Within each group, 10 specimens underwent cyclic loading, whereas 10 were assigned to preload evaluation. Cyclic loading followed ISO 14801:2016. Screws were tightened to 30 N·cm, retightened after 10 min, and subjected to 300 000 cycles at 200 N using a fatigue testing machine (DYNA-MESS Prüfsysteme GmbH, Aachen, Germany). Removal torque was measured with a digital torque gauge (BTGE50CN, Tohnichi, Japan). For preload assessment, each specimen completed 10 tightening-loosening cycles at the same torque during which tensile force was recorded using a customized device. One representative screw per group was analyzed by SEM and micro-CT before and after loading to evaluate surface and structural changes. RESULTS: After cyclic loading, the removal torque did not differ significantly between the Control group (17.58 ± 1.85 N·cm) and the Wrap group (17.50 ± 1.77 N·cm), whereas the Coat group demonstrated significantly lower removal torque (14.71 ± 2.35 N·cm; p < 0.05). In the preload test, preload progressively decreased in the Control (-3.13 N per cycle) and Coat groups (-0.51 N per cycle), while the Wrap group showed a consistent increase with repeated tightening (+8.22 N per cycle). SEM revealed surface wear and partial loss of PTFE material after loading, particularly in the coated specimens. Micro-CT analysis showed no detectable differences in internal adaptation among groups before or after cyclic loading. CONCLUSIONS: PTFE tape wrapping increased preload during repeated tightening but did not improve removal torque after cyclic loading compared with those of untreated screws. Although PTFE-coated screws produced the highest initial preload, they were associated with reduced removal torque after loading. SEM analysis confirmed coating deterioration, whereas internal fit remained comparable across groups.

Clinical Implant Dentistry and Related ResearchVol. 28(3)
Chiang Mai University (TH)
Chiang Mai University
Openalex Percentile: Top 4%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.