Image Quality and Radiation Dose Comparison for Abdominopelvic CT Studies Performed Using Photon-counting CT and Dual-energy CT: a Clinical Study

Abstract Objectives To compare radiation dose and image quality of abdominopelvic CT studies performed on dual-energy energy-integrating detector CT (EID-CT) and photo-counting CT (PCCT) for the same patient cohort with clinically aligned protocols. Materials and Methods An IRB-approved retrospective cohort study included 49 patients who underwent portal venous phase abdominopelvic CT exams using both dual-energy CT and PCCT scanners within a maximum interval of three months. Images reconstructed included virtual monoenergetic images at 50, 60, and 70 keV and mixed images. CTDIvol values were recorded to assess radiation dose. Regions of interest were drawn in the six anatomical locations for quantitative analysis. The overall image quality, contrast enhancement, image noise, and presence of artifacts were evaluated using a 5-point Likert scale. Results No significant differences were observed in image quality scores between PCCT and EID-CT for most comparisons. Calculated contrast-to-noise ratios (CNR) were significantly higher on PCCT except for 70 keV. PCCT demonstrated a significantly lower radiation dose, with an average reduction of 20 ± 15%, up to 51%. Conclusion For abdominopelvic CT exams with our institutional protocol settings, PCCT has demonstrated a significant reduction in radiation dose while maintaining similar image quality as EID-CT. Advances in knowledge With clinically aligned protocols for the same cohort of patients, the results provided robust evidence for understanding the benefits of PCCT. It also provided a possible pathway to transit from EID-CT to PCCT in the abdominopelvic imaging.

Authors

Institutions

Publication Details

Journal
BJR|Open
Published
2026-05-22
DOI
https://doi.org/10.1093/bjro/tzag012
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Image Quality and Radiation Dose Comparison for Abdominopelvic CT Studies Performed Using Photon-counting CT and Dual-energy CT: a Clinical Study

A Aljabal, N Yadu, A Goel, E Zeikus et al.
BJR|Open
Advanced X-ray and CT Imaging
article

Image Quality and Radiation Dose Comparison for Abdominopelvic CT Studies Performed Using Photon-counting CT and Dual-energy CT: a Clinical Study

A Aljabal, N Yadu, A Goel, E Zeikus, L Ren, X Duan, R Ahn, Y Xi, L Ananthakrishnan
article en

Abstract

Abstract Objectives To compare radiation dose and image quality of abdominopelvic CT studies performed on dual-energy energy-integrating detector CT (EID-CT) and photo-counting CT (PCCT) for the same patient cohort with clinically aligned protocols. Materials and Methods An IRB-approved retrospective cohort study included 49 patients who underwent portal venous phase abdominopelvic CT exams using both dual-energy CT and PCCT scanners within a maximum interval of three months. Images reconstructed included virtual monoenergetic images at 50, 60, and 70 keV and mixed images. CTDIvol values were recorded to assess radiation dose. Regions of interest were drawn in the six anatomical locations for quantitative analysis. The overall image quality, contrast enhancement, image noise, and presence of artifacts were evaluated using a 5-point Likert scale. Results No significant differences were observed in image quality scores between PCCT and EID-CT for most comparisons. Calculated contrast-to-noise ratios (CNR) were significantly higher on PCCT except for 70 keV. PCCT demonstrated a significantly lower radiation dose, with an average reduction of 20 ± 15%, up to 51%. Conclusion For abdominopelvic CT exams with our institutional protocol settings, PCCT has demonstrated a significant reduction in radiation dose while maintaining similar image quality as EID-CT. Advances in knowledge With clinically aligned protocols for the same cohort of patients, the results provided robust evidence for understanding the benefits of PCCT. It also provided a possible pathway to transit from EID-CT to PCCT in the abdominopelvic imaging.

BJR|Open
Southwestern Medical Center (US), The University of Texas Southwestern Medical Center (US)
Affordable and clean energy
Openalex Percentile: Top 9%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Image Quality and Radiation Dose Comparison for Abdominopelvic CT Studies Performed Using Photon-counting CT and Dual-energy CT: a Clinical Study — A Aljabal, N Yadu, et al. · BJR|Open (2026) | TGRS Research Map | TGRS