Comparison of ventilation with second-generation supraglottic airway devices in a prospective randomized cadaver study

Abstract Second-generation supraglottic airway devices (SADs) are widely used alternatives to endotracheal intubation in both anesthesia and emergency airway management. The newly developed LT®evo represents a redesigned version of the established Laryngeal Tube Suction Disposable (LTS-D). This study compared the ventilatory performance of the LT®evo with endotracheal intubation and three other second-generation SADs— LTS-D, Ambu® AuraGain™ and i-gel® Plus—using a controlled human cadaver model. In this prospective, randomized experiment, six thawed adult cadavers were sequentially ventilated via an endotracheal tube (ET) and four SADs under standardized mechanical ventilation (tidal volume 7 mL/kg, PEEP 5 cmH₂O, respiratory rate 10/min). Airway pressure and flow were continuously recorded (sampling rate 200 Hz) to determine inspiratory tidal volume, pressure, flow, and calculated resistance. Endotracheal intubation using fiberoptics was attempted via LT®evo , AuraGain™ and i-gel® Plus. Data were analyzed using Kruskal–Wallis tests with Holm-corrected post-hoc pairwise comparisons. From six human cadavers (60–85kg estimated body weight), 30 complete ventilation datasets were obtained. Initial fiberoptic control of SAD placement showed optimal positioning in 13 of 24 insertions. During ventilation cycles, inspiratory tidal volumes of the LT®evo (0.49 [0.46–0.52]) were similar to those of ET, AuraGain™ and LTS-D and superior to those of the i-gel® Plus (0.31 L [0.27–0.33], p < 0.001). Peak inspiratory pressure and inspiratory resistance were higher in ET when compared to LT®evo and all other SADs. Endotracheal intubation using fiberoptics was successful in all attempts via LT®evo , AuraGain™ and i-gel® Plus. In this controlled cadaver study, the newly designed LT®evo demonstrated ventilatory performance comparable to that of the LTS-D and the Ambu® AuraGain TM and was superior to the i-gel® Plus in delivering adequate tidal volumes. Additionally, LT®evo allowed for fiberoptic ET placement. Importantly, these results are specific to the cadaveric model and reflect technical variability in ventilation mechanics rather than clinical effectiveness. Clinical trial number German Clinical Trials Registry, DRKS00038309, on October 30 th , 2025.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-05-21
DOI
https://doi.org/10.1038/s41598-026-53005-5
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of ventilation with second-generation supraglottic airway devices in a prospective randomized cadaver study

Harald Genzwürker, Stephan Katzenschlager, Frank Weilbacher, Lisa Kaltschmidt et al.
Scientific Reports
Airway Management and Intubation Techniques
article

Comparison of ventilation with second-generation supraglottic airway devices in a prospective randomized cadaver study

Harald Genzwürker, Stephan Katzenschlager, Frank Weilbacher, Lisa Kaltschmidt, Marita Klein, Erik Popp, Nikolai Kaltschmidt
article en

Abstract

Abstract Second-generation supraglottic airway devices (SADs) are widely used alternatives to endotracheal intubation in both anesthesia and emergency airway management. The newly developed LT®evo represents a redesigned version of the established Laryngeal Tube Suction Disposable (LTS-D). This study compared the ventilatory performance of the LT®evo with endotracheal intubation and three other second-generation SADs— LTS-D, Ambu® AuraGain™ and i-gel® Plus—using a controlled human cadaver model. In this prospective, randomized experiment, six thawed adult cadavers were sequentially ventilated via an endotracheal tube (ET) and four SADs under standardized mechanical ventilation (tidal volume 7 mL/kg, PEEP 5 cmH₂O, respiratory rate 10/min). Airway pressure and flow were continuously recorded (sampling rate 200 Hz) to determine inspiratory tidal volume, pressure, flow, and calculated resistance. Endotracheal intubation using fiberoptics was attempted via LT®evo , AuraGain™ and i-gel® Plus. Data were analyzed using Kruskal–Wallis tests with Holm-corrected post-hoc pairwise comparisons. From six human cadavers (60–85kg estimated body weight), 30 complete ventilation datasets were obtained. Initial fiberoptic control of SAD placement showed optimal positioning in 13 of 24 insertions. During ventilation cycles, inspiratory tidal volumes of the LT®evo (0.49 [0.46–0.52]) were similar to those of ET, AuraGain™ and LTS-D and superior to those of the i-gel® Plus (0.31 L [0.27–0.33], p < 0.001). Peak inspiratory pressure and inspiratory resistance were higher in ET when compared to LT®evo and all other SADs. Endotracheal intubation using fiberoptics was successful in all attempts via LT®evo , AuraGain™ and i-gel® Plus. In this controlled cadaver study, the newly designed LT®evo demonstrated ventilatory performance comparable to that of the LTS-D and the Ambu® AuraGain TM and was superior to the i-gel® Plus in delivering adequate tidal volumes. Additionally, LT®evo allowed for fiberoptic ET placement. Importantly, these results are specific to the cadaveric model and reflect technical variability in ventilation mechanics rather than clinical effectiveness. Clinical trial number German Clinical Trials Registry, DRKS00038309, on October 30 th , 2025.

Scientific ReportsVol. 16(1)
Heidelberg University (DE), University Hospital Heidelberg (DE)
Universitätsklinikum Heidelberg
Openalex Percentile: Top 5%
Airway Management and Intubation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.