Occupational dose attenuation and dosimetric performance of autonomous radiation protection systems in fluoroscopy: a scoping review with implications for the role of conventional lead aprons

Abstract

Background: Interventional cardiology procedures chronically expose healthcare personnel to ionizing radiation, raising concerns about long-term health effects such as cataracts and cancer. Consequently, growing attention has been directed toward advanced shielding technologies capable of substantially reducing occupational dose while improving ergonomic conditions." 
Objective: To synthesize the available evidence on the technical performance and radiation attenuation capabilities of Autonomous Radiation Protection Systems (Protego™, Rampart™, and EggNest XR™), used in interventional cardiology settings.
Methods: A scoping review was conducted following the PRISMA-ScR framework. Relevant studies were identified through a systematic search of PubMed, EBSCO, Scopus, WOS and Cochrane. Eligible studies included clinical and phantom-based assessments reporting quantitative dosimetric outcomes.
Results: Fourteen studies met the inclusion criteria. Protego™ was associated with consistently high reported attenuation, typically ≥95% in studies providing quantifiable measurements; in several investigations, post-shielding doses fell below the detection limits of standard personal dosimeters, whereas Rampart™ demonstrated wider reported attenuation ranges (approximately 16.7-100%) across heterogeneous study designs and measurement protocols and EggNest XR™ exhibited attenuation values ranging from 91.5% to 99.0% in a single phantom-based study. Device performance varied by anatomical site and professional role, with higher attenuation typically observed among primary operators (often exceeding 95% at head and torso levels) compared with more variable reductions in peripheral personnel such as nurses and anesthesiologists, where attenuation occasionally decreased to the 80-90% range depending on positioning and projection.
Conclusion: The attenuation results for autonomous shielding systems are promising, particularly for fixed systems, which demonstrate consistently high performance. Given their potential for ergonomic improvement and radiation reduction, Autonomous Radiation Protection Systems (ARPS) are emerging as valuable tools. However, given the limited clinical validation and the heterogeneity of existing study, the implementation of these systems should be supported by site-specific validation and standardized dosimetry protocols."
 &#xD.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2026-05-21
DOI
https://doi.org/10.1088/1361-6498/ae7122
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Occupational dose attenuation and dosimetric performance of autonomous radiation protection systems in fluoroscopy: a scoping review with implications for the role of conventional lead aprons

Ricardo Ampuero, Adlin López Díaz, Sergio Ramos-Avasola, E Honorio da Silva et al.
Journal of Radiological Protection
Radiation Dose and Imaging
article

Occupational dose attenuation and dosimetric performance of autonomous radiation protection systems in fluoroscopy: a scoping review with implications for the role of conventional lead aprons

Ricardo Ampuero, Adlin López Díaz, Sergio Ramos-Avasola, E Honorio da Silva, Constanza Vira, Mario Araya, Nitza Espinosa-Rojo
article en

Abstract

Abstract

Background: Interventional cardiology procedures chronically expose healthcare personnel to ionizing radiation, raising concerns about long-term health effects such as cataracts and cancer. Consequently, growing attention has been directed toward advanced shielding technologies capable of substantially reducing occupational dose while improving ergonomic conditions." 
Objective: To synthesize the available evidence on the technical performance and radiation attenuation capabilities of Autonomous Radiation Protection Systems (Protego™, Rampart™, and EggNest XR™), used in interventional cardiology settings.
Methods: A scoping review was conducted following the PRISMA-ScR framework. Relevant studies were identified through a systematic search of PubMed, EBSCO, Scopus, WOS and Cochrane. Eligible studies included clinical and phantom-based assessments reporting quantitative dosimetric outcomes.
Results: Fourteen studies met the inclusion criteria. Protego™ was associated with consistently high reported attenuation, typically ≥95% in studies providing quantifiable measurements; in several investigations, post-shielding doses fell below the detection limits of standard personal dosimeters, whereas Rampart™ demonstrated wider reported attenuation ranges (approximately 16.7-100%) across heterogeneous study designs and measurement protocols and EggNest XR™ exhibited attenuation values ranging from 91.5% to 99.0% in a single phantom-based study. Device performance varied by anatomical site and professional role, with higher attenuation typically observed among primary operators (often exceeding 95% at head and torso levels) compared with more variable reductions in peripheral personnel such as nurses and anesthesiologists, where attenuation occasionally decreased to the 80-90% range depending on positioning and projection.
Conclusion: The attenuation results for autonomous shielding systems are promising, particularly for fixed systems, which demonstrate consistently high performance. Given their potential for ergonomic improvement and radiation reduction, Autonomous Radiation Protection Systems (ARPS) are emerging as valuable tools. However, given the limited clinical validation and the heterogeneity of existing study, the implementation of these systems should be supported by site-specific validation and standardized dosimetry protocols."
 &#xD.

Journal of Radiological Protection
Universidad de La Frontera (CL), Universidade de São Paulo (BR), Clínica Alemana (CL), San Sebastián University (CL), Hospital Regional de Arica Doctor Juan Noé Crevani (CL)
Openalex Percentile: Top 5%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.