Validation of a novel three-dimensional ultrafast cardiac magnetic resonance imaging protocol in adolescents: a non-inferiority study compared with the two-dimensional gold standard

BACKGROUND: Cardiac magnetic resonance (CMR) is the gold standard for assessing cardiac anatomy and function. However, long acquisition times and multiple breath-holds pose significant challenges in pediatric imaging. OBJECTIVE: To perform a non-inferiority analysis comparing a novel 3-dimensional (D) ultrafast CMR protocol against the conventional 2-dimensional (D) gold standard for evaluating cardiac function, strain, and tissue characterization in adolescents. MATERIALS AND METHODS: Thirty-nine adolescents (mean age 12.2±2.6 years) underwent both a standard 2-D protocol and a 3-D ultrafast protocol at 3-T. The 3-D protocol comprised enhanced sensitivity encoding by static outer volume subtraction (ESSOS) cine and 3-D late gadolinium enhancement. Image quality and diagnostic confidence were compared. A pre-specified non-inferiority margin (Δ) was used to assess functional and strain parameters. Agreement was evaluated using Bland-Altman analysis and intraclass correlation coefficients (ICCs). RESULTS: The total scan time was significantly shorter for the 3-D protocol compared to the 2-D protocol (75.8±10.0 s vs. 734.4±19.6 s, P<0.05). Image quality scores were comparable between protocols (median score 4.0, P=0.74). The 3-D protocol demonstrated statistical non-inferiority for all functional and strain metrics. Bland-Altman analysis showed minimal bias for key parameters, and the 95% confidence intervals for differences met the pre-specified non-inferiority margins. Intraclass correlation coefficients (ICCs) indicated good to excellent agreement for all parameters (n=39). CONCLUSION: The novel 3-D ultrafast CMR protocol is non-inferior to the conventional 2-D gold standard for quantitative assessment of cardiac function and strain in adolescents. It offers comparable image quality with significantly reduced acquisition times, potentially improving clinical feasibility in pediatric populations.

Authors

Institutions

Publication Details

Journal
Pediatric Radiology
Published
2026-05-21
DOI
https://doi.org/10.1007/s00247-026-06651-w
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of a novel three-dimensional ultrafast cardiac magnetic resonance imaging protocol in adolescents: a non-inferiority study compared with the two-dimensional gold standard

L XU, Xinyan Tao, Shuo Liu, Rui Wang et al.
Pediatric Radiology
Cardiac Imaging and Diagnostics
article

Validation of a novel three-dimensional ultrafast cardiac magnetic resonance imaging protocol in adolescents: a non-inferiority study compared with the two-dimensional gold standard

L XU, Xinyan Tao, Shuo Liu, Rui Wang, Shuang Li, Wei Chen, Yike Zhao, Wei Li, Hui Wang, Jianxiu Lian
article en

Abstract

BACKGROUND: Cardiac magnetic resonance (CMR) is the gold standard for assessing cardiac anatomy and function. However, long acquisition times and multiple breath-holds pose significant challenges in pediatric imaging. OBJECTIVE: To perform a non-inferiority analysis comparing a novel 3-dimensional (D) ultrafast CMR protocol against the conventional 2-dimensional (D) gold standard for evaluating cardiac function, strain, and tissue characterization in adolescents. MATERIALS AND METHODS: Thirty-nine adolescents (mean age 12.2±2.6 years) underwent both a standard 2-D protocol and a 3-D ultrafast protocol at 3-T. The 3-D protocol comprised enhanced sensitivity encoding by static outer volume subtraction (ESSOS) cine and 3-D late gadolinium enhancement. Image quality and diagnostic confidence were compared. A pre-specified non-inferiority margin (Δ) was used to assess functional and strain parameters. Agreement was evaluated using Bland-Altman analysis and intraclass correlation coefficients (ICCs). RESULTS: The total scan time was significantly shorter for the 3-D protocol compared to the 2-D protocol (75.8±10.0 s vs. 734.4±19.6 s, P<0.05). Image quality scores were comparable between protocols (median score 4.0, P=0.74). The 3-D protocol demonstrated statistical non-inferiority for all functional and strain metrics. Bland-Altman analysis showed minimal bias for key parameters, and the 95% confidence intervals for differences met the pre-specified non-inferiority margins. Intraclass correlation coefficients (ICCs) indicated good to excellent agreement for all parameters (n=39). CONCLUSION: The novel 3-D ultrafast CMR protocol is non-inferior to the conventional 2-D gold standard for quantitative assessment of cardiac function and strain in adolescents. It offers comparable image quality with significantly reduced acquisition times, potentially improving clinical feasibility in pediatric populations.

Pediatric Radiology
Philips (China) (CN), Beijing Anzhen Hospital (CN), Baoding People's Hospital (CN)
Openalex Percentile: Top 5%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.