Glycemic Concordance of Simplera Sync <b>™</b> and Instinct <b>™</b> Sensors in MiniMed <b>™</b> 780G System Users: A Randomized Crossover Feasibility Study

Introduction: This study evaluated the performance of Simplera Sync™ and Instinct™ sensors when paired with the MiniMed™ 780G (MM780G) system and assessed whether switching between sensors affected clinical outcomes of people with type 1 diabetes (PwT1D). Methods: A single-center, two-arm, randomized crossover study in adult PwT1D started with either the Simplera Sync sensor or Instinct sensor for a 6-week period, followed by crossover to the alternate sensor for an additional 6 weeks. The primary end point was noninferiority in time in range (TIR; 70–180 mg/dL), with a noninferiority margin of 7.5%. Secondary end points included HbA1c and other continuous glucose monitoring-derived metrics. Results: Twenty-five participants were assessed for eligibility, 24 were randomized, and 23 completed the study. TIR was 80.6 (9.8)% during the Simplera Sync sensor period and 79.1 (9.5)% during the Instinct sensor period. The mean difference was 1.5% (95% confidence interval: 0.18–2.79), and the noninferiority criteria were met. HbA1c at the end of the Simplera Sync and Instinct sensor periods was 6.8 (0.7)% and 6.7 (0.7)%, respectively. No serious adverse events related to sensor use were reported. Conclusion: This randomized controlled crossover trial demonstrates that both Simplera Sync and Instinct sensors meet international consensus targets for glycemic control when used with the MM780G system, and that the sensors can be interchanged without meaningful differences in glycemic control. This study supports the concept that the automated insulin delivery algorithm, rather than the sensor, is the primary determinant of clinical outcomes in MM780G users.

Authors

Institutions

Publication Details

Journal
Diabetes Technology & Therapeutics
Published
2026-05-21
DOI
https://doi.org/10.1177/15209156261449884
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Glycemic Concordance of Simplera Sync ™ and Instinct ™ Sensors in MiniMed ™ 780G System Users: A Randomized Crossover Feasibility Study

Benyamin Grossman, Maya Laron-Hirsh, Anirban Roy, George Mallas et al.
Diabetes Technology & Therapeutics
Diabetes Management and Research
article

Glycemic Concordance of Simplera Sync ™ and Instinct ™ Sensors in MiniMed ™ 780G System Users: A Randomized Crossover Feasibility Study

Benyamin Grossman, Maya Laron-Hirsh, Anirban Roy, George Mallas, Robert A. Vigersky, Ohad Cohen, Sapir Barel, Amir Tirosh
article en

Abstract

Introduction: This study evaluated the performance of Simplera Sync™ and Instinct™ sensors when paired with the MiniMed™ 780G (MM780G) system and assessed whether switching between sensors affected clinical outcomes of people with type 1 diabetes (PwT1D). Methods: A single-center, two-arm, randomized crossover study in adult PwT1D started with either the Simplera Sync sensor or Instinct sensor for a 6-week period, followed by crossover to the alternate sensor for an additional 6 weeks. The primary end point was noninferiority in time in range (TIR; 70–180 mg/dL), with a noninferiority margin of 7.5%. Secondary end points included HbA1c and other continuous glucose monitoring-derived metrics. Results: Twenty-five participants were assessed for eligibility, 24 were randomized, and 23 completed the study. TIR was 80.6 (9.8)% during the Simplera Sync sensor period and 79.1 (9.5)% during the Instinct sensor period. The mean difference was 1.5% (95% confidence interval: 0.18–2.79), and the noninferiority criteria were met. HbA1c at the end of the Simplera Sync and Instinct sensor periods was 6.8 (0.7)% and 6.7 (0.7)%, respectively. No serious adverse events related to sensor use were reported. Conclusion: This randomized controlled crossover trial demonstrates that both Simplera Sync and Instinct sensors meet international consensus targets for glycemic control when used with the MM780G system, and that the sensors can be interchanged without meaningful differences in glycemic control. This study supports the concept that the automated insulin delivery algorithm, rather than the sensor, is the primary determinant of clinical outcomes in MM780G users.

Diabetes Technology & Therapeutics
California State University, Northridge (US), Sheba Medical Center (IL), Diabetes Foundation (IN)
Openalex Percentile: Top 6%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.