Diagnostic Performance of a Novel Wireless Direct-Conversion Intraoral Sensor Versus CMOS and PSP Systems for Proximal Caries Detection: An Ex Vivo Study

OBJECTIVES: To compare the diagnostic accuracy of a new wireless direct-conversion digital intraoral sensor to current indirect conversion, complementary metal-oxide semiconductor (CMOS) and photostimulable phosphor (PSP) sensors for detecting proximal caries. METHODS: The sample in this ex vivo study were 28 human premolars with 21 micro-CT verified non-cavitated proximal dental carious lesions. Images were obtained using three digital intraoral systems: a wireless direct-conversion sensor, a wired indirect conversion CMOS sensor, and a PSP sensor. A custom phantom was constructed providing a limited dental arch segment comprising a molar, first premolar, and canine. The premolar was substituted and the phantom imaged using each intraoral system with (filtered) and without (raw) the application of a manufacturer-recommended enhancements, providing 168 images. Eight oral radiologists assessed the images using a five-point scale. Receiver operating characteristic analysis was performed to calculate sensitivity, specificity and the area under the curve (AUC) and compared across modalities using two-way analysis of variance (ANOVA) with post hoc Tukey's test at an a priori level of significance of 0.05. RESULTS: The AUC ranged from 0.58 to 0.73 (sensitivity range: 0.21-0.68, specificity range: 0.69-0.92). The wireless direct-conversion sensor (0.72 ± 0.06) showed diagnostic performance comparable to CMOS (0.72 ± 0.06) and superior to PSP (0.58 ± 0.07) for raw images (p < 0.001). The direct-conversion sensor without image enhancement (0.68 ± 0.18) demonstrated the highest sensitivity, whereas the PSP without image enhancement (0.21 ± 0.09) had the lowest sensitivity. Application of image enhancement improved PSP performance (0.48 ± 0.16), approximating that of the other systems. CONCLUSIONS: Based on this ex vivo study, the wireless direct-conversion and CMOS sensors demonstrated similar diagnostic performance, superior to PSP in unfiltered conditions. These findings should be interpreted in light of the ex vivo design and modest sample size.

Authors

Institutions

Publication Details

Journal
Dentomaxillofacial Radiology
Published
2026-05-21
DOI
https://doi.org/10.1093/dmfr/twag031
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic Performance of a Novel Wireless Direct-Conversion Intraoral Sensor Versus CMOS and PSP Systems for Proximal Caries Detection: An Ex Vivo Study

Matheus Barros‐Costa, Gustavo Machado Santaella, Gabriela Gavilán Hadid, Deborah Queiroz Freitas et al.
Dentomaxillofacial Radiology
Dental Radiography and Imaging
article

Diagnostic Performance of a Novel Wireless Direct-Conversion Intraoral Sensor Versus CMOS and PSP Systems for Proximal Caries Detection: An Ex Vivo Study

Matheus Barros‐Costa, Gustavo Machado Santaella, Gabriela Gavilán Hadid, Deborah Queiroz Freitas, Francisco Carlos Groppo, William C Scarfe, Christiano Oliveira-Santos
article en

Abstract

OBJECTIVES: To compare the diagnostic accuracy of a new wireless direct-conversion digital intraoral sensor to current indirect conversion, complementary metal-oxide semiconductor (CMOS) and photostimulable phosphor (PSP) sensors for detecting proximal caries. METHODS: The sample in this ex vivo study were 28 human premolars with 21 micro-CT verified non-cavitated proximal dental carious lesions. Images were obtained using three digital intraoral systems: a wireless direct-conversion sensor, a wired indirect conversion CMOS sensor, and a PSP sensor. A custom phantom was constructed providing a limited dental arch segment comprising a molar, first premolar, and canine. The premolar was substituted and the phantom imaged using each intraoral system with (filtered) and without (raw) the application of a manufacturer-recommended enhancements, providing 168 images. Eight oral radiologists assessed the images using a five-point scale. Receiver operating characteristic analysis was performed to calculate sensitivity, specificity and the area under the curve (AUC) and compared across modalities using two-way analysis of variance (ANOVA) with post hoc Tukey's test at an a priori level of significance of 0.05. RESULTS: The AUC ranged from 0.58 to 0.73 (sensitivity range: 0.21-0.68, specificity range: 0.69-0.92). The wireless direct-conversion sensor (0.72 ± 0.06) showed diagnostic performance comparable to CMOS (0.72 ± 0.06) and superior to PSP (0.58 ± 0.07) for raw images (p < 0.001). The direct-conversion sensor without image enhancement (0.68 ± 0.18) demonstrated the highest sensitivity, whereas the PSP without image enhancement (0.21 ± 0.09) had the lowest sensitivity. Application of image enhancement improved PSP performance (0.48 ± 0.16), approximating that of the other systems. CONCLUSIONS: Based on this ex vivo study, the wireless direct-conversion and CMOS sensors demonstrated similar diagnostic performance, superior to PSP in unfiltered conditions. These findings should be interpreted in light of the ex vivo design and modest sample size.

Dentomaxillofacial Radiology
University of Louisville (US), Universidade Estadual de Campinas (UNICAMP) (BR), Methodist University of Piracicaba (BR), Forest Science and Research Institute (BR)
Openalex Percentile: Top 4%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.