Oxygenator failures during cardiopulmonary bypass lead to associated injury patterns

Background Cardiopulmonary bypass (CPB) oxygenators provide essential intra-operative gas exchange. Although rare, CPB oxygenator failures can cause catastrophic morbidity and mortality, and published analyses of CPB-specific oxygenator failures are limited. We aimed to study reports of CPB oxygenator failures to observe patterns to inform preparation and device improvement. Methods We retrospectively reviewed Manufacturer and User Facility Device Experience (MAUDE) reports (April 2015–April 2025) of CPB oxygenator failures resulting in injury or death. ECMO events, malfunctions without harm, and duplicates were excluded. Manufacturer, device problem, and patient problem were compared between injury and death cohorts using chi-square or Fisher’s exact tests. Narrative reports underwent supervised natural language processing (NLP) to identify clinical themes. Results A total of 388 events met criteria: 279 injuries (71.9%) and 109 deaths (28.1%). Events were commonly related to six manufacturers. Manufacturer B and Manufacturer D devices were disproportionately represented in deaths (43.1% vs 11.1% of injuries, p < 0.001 and 12.84% vs 5.38% of injuries, p = 0.017), while Manufacturer A was more frequently linked to injuries (60.6% vs 22.9%, p < 0.001), although the incidence of failure for each manufacturer could not be calculated. The remaining manufacturers were evenly distributed among report types. Device problem categories differed by outcome: “Unexpected Device Operation” was significantly more common in deaths (7.3% vs 1.4%, p = 0.0054), whereas “Human Factors” were more common in injuries (11.8% vs 4.6%, p = 0.036). NLP identified intra-operative performance/monitoring failures, hemolysis and thromboembolic events, and post-procedural evaluation/quality management as common descriptions of device failure situations. Conclusion Using MAUDE reports, CPB oxygenator failures causing harm show distinct manufacturer, device problem, and outcome patterns but incidence could not be determined. Reports of unexpected operational failures and thromboembolic complications are more often fatal. NLP enhances detection of clinically relevant patterns, supporting targeted safety strategies and device improvements.

Authors

Institutions

Publication Details

Journal
Perfusion
Published
2026-05-21
DOI
https://doi.org/10.1177/02676591261452056
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Oxygenator failures during cardiopulmonary bypass lead to associated injury patterns

Jace C. Bradshaw, AlleaBelle Bradshaw, Chad Wierschke, Jennifer S. Lawton et al.
Perfusion
Mechanical Circulatory Support Devices
article

Oxygenator failures during cardiopulmonary bypass lead to associated injury patterns

Jace C. Bradshaw, AlleaBelle Bradshaw, Chad Wierschke, Jennifer S. Lawton, Rishab Agarwal, Shivani Shirodkar, Caroline Tran, Vishnu Dontu, Chetan Pasrija, Laeben Lester
article en

Abstract

Background Cardiopulmonary bypass (CPB) oxygenators provide essential intra-operative gas exchange. Although rare, CPB oxygenator failures can cause catastrophic morbidity and mortality, and published analyses of CPB-specific oxygenator failures are limited. We aimed to study reports of CPB oxygenator failures to observe patterns to inform preparation and device improvement. Methods We retrospectively reviewed Manufacturer and User Facility Device Experience (MAUDE) reports (April 2015–April 2025) of CPB oxygenator failures resulting in injury or death. ECMO events, malfunctions without harm, and duplicates were excluded. Manufacturer, device problem, and patient problem were compared between injury and death cohorts using chi-square or Fisher’s exact tests. Narrative reports underwent supervised natural language processing (NLP) to identify clinical themes. Results A total of 388 events met criteria: 279 injuries (71.9%) and 109 deaths (28.1%). Events were commonly related to six manufacturers. Manufacturer B and Manufacturer D devices were disproportionately represented in deaths (43.1% vs 11.1% of injuries, p < 0.001 and 12.84% vs 5.38% of injuries, p = 0.017), while Manufacturer A was more frequently linked to injuries (60.6% vs 22.9%, p < 0.001), although the incidence of failure for each manufacturer could not be calculated. The remaining manufacturers were evenly distributed among report types. Device problem categories differed by outcome: “Unexpected Device Operation” was significantly more common in deaths (7.3% vs 1.4%, p = 0.0054), whereas “Human Factors” were more common in injuries (11.8% vs 4.6%, p = 0.036). NLP identified intra-operative performance/monitoring failures, hemolysis and thromboembolic events, and post-procedural evaluation/quality management as common descriptions of device failure situations. Conclusion Using MAUDE reports, CPB oxygenator failures causing harm show distinct manufacturer, device problem, and outcome patterns but incidence could not be determined. Reports of unexpected operational failures and thromboembolic complications are more often fatal. NLP enhances detection of clinically relevant patterns, supporting targeted safety strategies and device improvements.

Perfusion
Johns Hopkins University (US), Johns Hopkins Medicine (US)
Good health and well-being
Openalex Percentile: Top 9%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.