The stability of diffusion measurements with simultaneous multi-slice imaging: A multi-scanner, multi-site, test-retest study using the NIST/QIBA CaliberMRI diffusion phantom

This study evaluated how simultaneous multi-slice (SMS) acceleration affects the accuracy and repeatability of apparent diffusion coefficient (ADC) measurements across multiple MRI platforms using a traceable diffusion phantom. A multi-centre, multi-scanner test-retest study was conducted across seven MRI systems from three major vendors using a CaliberMRI diffusion phantom. Diffusion-weighted images with SMS acceleration factors of 0, 2, 3, and 4 were acquired. Measured ADC values were compared to temperature matched NIST reference standards. The signal-to-noise ratio (SNR) was calculated for each acquisition. Bias was measured using percentage change from reference values and multivariable regression. Lin's coefficient was used to summarise the concordance between repeat acquisitions. On average, scanners overestimated ADC by 4%. SMS factor 2 did not significantly bias measurements compared to non-SMS acquisitions. However, SMS factors 3 and 4 introduced systematic biases, with mean ADC overestimations of 4.62% and 9.22%, respectively. Biases were most pronounced in media with ADC values outside the physiological range. Despite reduced SNR at higher SMS factors, intra- and inter-session repeatability remained high, suggesting intrinsic bias rather than random noise. These findings demonstrate that higher SMS factors can bias ADC measures and degrade image quality highlighting the need for caution when absolute quantification is required. In contrast, low SMS acceleration appears suitable for reducing scan times in clinical and research protocols without materially compromising ADC accuracy across platforms. These findings have implications for the adoption of SMS in quantitative diffusion imaging protocols across platforms.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-05-21
DOI
https://doi.org/10.1088/2057-1976/ae7116
Primary Topic
Advanced Neuroimaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The stability of diffusion measurements with simultaneous multi-slice imaging: A multi-scanner, multi-site, test-retest study using the NIST/QIBA CaliberMRI diffusion phantom

Matthew Marzetti, Gráinne Bourke, Hamied Haroon, Yue Wang et al.
Biomedical Physics & Engineering Express
Advanced Neuroimaging Techniques and Applications
article

The stability of diffusion measurements with simultaneous multi-slice imaging: A multi-scanner, multi-site, test-retest study using the NIST/QIBA CaliberMRI diffusion phantom

Matthew Marzetti, Gráinne Bourke, Hamied Haroon, Yue Wang, William Lloyd, I R V I N Teh, Caroline Lea‐Carnall, Ryckie Wade, Laura Parkes, Todor Karaulanov
article en

Abstract

This study evaluated how simultaneous multi-slice (SMS) acceleration affects the accuracy and repeatability of apparent diffusion coefficient (ADC) measurements across multiple MRI platforms using a traceable diffusion phantom. A multi-centre, multi-scanner test-retest study was conducted across seven MRI systems from three major vendors using a CaliberMRI diffusion phantom. Diffusion-weighted images with SMS acceleration factors of 0, 2, 3, and 4 were acquired. Measured ADC values were compared to temperature matched NIST reference standards. The signal-to-noise ratio (SNR) was calculated for each acquisition. Bias was measured using percentage change from reference values and multivariable regression. Lin's coefficient was used to summarise the concordance between repeat acquisitions. On average, scanners overestimated ADC by 4%. SMS factor 2 did not significantly bias measurements compared to non-SMS acquisitions. However, SMS factors 3 and 4 introduced systematic biases, with mean ADC overestimations of 4.62% and 9.22%, respectively. Biases were most pronounced in media with ADC values outside the physiological range. Despite reduced SNR at higher SMS factors, intra- and inter-session repeatability remained high, suggesting intrinsic bias rather than random noise. These findings demonstrate that higher SMS factors can bias ADC measures and degrade image quality highlighting the need for caution when absolute quantification is required. In contrast, low SMS acceleration appears suitable for reducing scan times in clinical and research protocols without materially compromising ADC accuracy across platforms. These findings have implications for the adoption of SMS in quantitative diffusion imaging protocols across platforms.

Biomedical Physics & Engineering Express
University of Leeds (GB), Leeds Teaching Hospitals NHS Trust (GB), University of Manchester (GB), Caliber Imaging and Diagnostics (United States) (US), Tianjin Hospital (CN)
National Institute for Health and Care Research, Department of Health and Social Care, Invention for Innovation Programme, Invention for Innovation, Leeds Biomedical Research Centre
Openalex Percentile: Top 5%
Advanced Neuroimaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.