SMUPhantom: A 3D-printable modular CT perfusion phantom for quantitative evaluation of tissue-mimicking dynamic contrast behavior

PURPOSE: Physical computed tomography (CT) perfusion (CTP) phantoms are useful for controlled protocol evaluation, but a modular 3D-printable platform that can generate distinguishable compartment-specific dynamic contrast behavior and support repeatability assessment remains needed. We developed the Separate Module Uniting Phantom (SMUPhantom) for protocol sensitivity studies and workflow benchmarking. METHODS: The SMUPhantom consists of three modules representing inflow, exchange, and outflow regions. Polyoxymethylene (POM) beads, soft wax, and acrylonitrile-butadiene-styrene (ABS) inserts were used to create healthy, penumbra-mimicking, and infarct-core-mimicking pathways, and a detachable bypass-flow pathway was added for flow-routing experiments. Two experimental series were performed on a NeuViz Epoch Elite CT scanner: contrast volume and injection rate were varied separately, each with and without bypass flow. Time-density curves (TDCs) and workstation-derived cerebral blood flow (CBF), cerebral blood volume (CBV), mean transit time (MTT), and time to peak (TTP) were evaluated. Five repeat measurements per condition assessed intra-scanner repeatability. RESULTS: The three compartments showed separable dynamic responses, with the healthy pathway exhibiting earlier and higher enhancement and the infarct-core-mimicking pathway showing lower and delayed responses. Increasing contrast volume broadened the bolus response and raised peak enhancement from 220.8 to 454.2 HU in the healthy pathway and from 149.4 to 354.4 HU in the infarct-core-mimicking pathway, whereas increasing injection rate shortened TTP from 22-25 s to 12-15 s and MTT from 18-21 s to 10-14 s. Opening bypass flow modified the compartment curves and map appearance, particularly in the infarct-core-mimicking pathway, while preserving interpretable trends across repeated scans. Repeatability analysis showed stable peak enhancement and cerebral blood volume measurements, with CVs not exceeding 3.54% and 2.81%, respectively. CONCLUSION: SMUPhantom provides a 3D-printable and quantitatively characterized platform for CTP protocol sensitivity studies, repeatability assessment, and workflow benchmarking under controlled flow-routing conditions.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-05-21
DOI
https://doi.org/10.1088/2057-1976/ae711c
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

SMUPhantom: A 3D-printable modular CT perfusion phantom for quantitative evaluation of tissue-mimicking dynamic contrast behavior

Mengwei Ma, H Wang, Hanxiao Song, Ge Wen et al.
Biomedical Physics & Engineering Express
Cardiac Imaging and Diagnostics
article

SMUPhantom: A 3D-printable modular CT perfusion phantom for quantitative evaluation of tissue-mimicking dynamic contrast behavior

Mengwei Ma, H Wang, Hanxiao Song, Ge Wen, Jianwei Zhou, Boxuan Cao, Basheng Hu, Dong Zeng, Jianhua Ma
article en

Abstract

PURPOSE: Physical computed tomography (CT) perfusion (CTP) phantoms are useful for controlled protocol evaluation, but a modular 3D-printable platform that can generate distinguishable compartment-specific dynamic contrast behavior and support repeatability assessment remains needed. We developed the Separate Module Uniting Phantom (SMUPhantom) for protocol sensitivity studies and workflow benchmarking. METHODS: The SMUPhantom consists of three modules representing inflow, exchange, and outflow regions. Polyoxymethylene (POM) beads, soft wax, and acrylonitrile-butadiene-styrene (ABS) inserts were used to create healthy, penumbra-mimicking, and infarct-core-mimicking pathways, and a detachable bypass-flow pathway was added for flow-routing experiments. Two experimental series were performed on a NeuViz Epoch Elite CT scanner: contrast volume and injection rate were varied separately, each with and without bypass flow. Time-density curves (TDCs) and workstation-derived cerebral blood flow (CBF), cerebral blood volume (CBV), mean transit time (MTT), and time to peak (TTP) were evaluated. Five repeat measurements per condition assessed intra-scanner repeatability. RESULTS: The three compartments showed separable dynamic responses, with the healthy pathway exhibiting earlier and higher enhancement and the infarct-core-mimicking pathway showing lower and delayed responses. Increasing contrast volume broadened the bolus response and raised peak enhancement from 220.8 to 454.2 HU in the healthy pathway and from 149.4 to 354.4 HU in the infarct-core-mimicking pathway, whereas increasing injection rate shortened TTP from 22-25 s to 12-15 s and MTT from 18-21 s to 10-14 s. Opening bypass flow modified the compartment curves and map appearance, particularly in the infarct-core-mimicking pathway, while preserving interpretable trends across repeated scans. Repeatability analysis showed stable peak enhancement and cerebral blood volume measurements, with CVs not exceeding 3.54% and 2.81%, respectively. CONCLUSION: SMUPhantom provides a 3D-printable and quantitatively characterized platform for CTP protocol sensitivity studies, repeatability assessment, and workflow benchmarking under controlled flow-routing conditions.

Biomedical Physics & Engineering Express
Nanfang Hospital (CN), Southern Medical University (CN)
National Natural Science Foundation of China, Beijing Municipal Natural Science Foundation, National Key Research and Development Program of China, Basic and Applied Basic Research Foundation of Guangdong Province
Openalex Percentile: Top 6%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.