Why catheter size matters more than viscosity in ethylene–vinyl alcohol copolymer reflux: an analytical study

Abstract Purpose Proximal reflux of liquid embolic agents remains a primary safety concern during endovascular embolization, yet the quantitative relationships governing reflux mechanics are poorly characterized. An analytical model was developed to determine the relative contributions of catheter geometry, agent viscosity, polymerization kinetics, and injection technique to reflux behavior of ethylene–vinyl alcohol copolymer (EVOH). Methods An annular Poiseuille flow model was constructed for a 1 mm vessel with coaxial microcatheters (1.3–2.8 French; catheter-to-vessel diameter ratio D* = 0.43–0.92). Carreau rheology was applied to EVOH-18 and EVOH-34; blood and iodinated contrast were modeled as Newtonian. Extensions incorporated vessel compliance, first-order EVOH polymerization kinetics, and intermittent injection parameterized by duty cycle δ. Results The annular geometry function φ(D*) spanned three orders of magnitude across the catheter range (0.179 to 0.0006), surpassing the maximum 2.4-fold inter-agent viscosity difference. At injection shear rates, EVOH-18 thinned to near-contrast viscosity (6.4 vs 6.1 mPa·s). EVOH polymerization reduced reflux up to 68% at low D* but was negligible at D* > 0.9. Under intermittent injection (δ = 0.20), effective reflux lengths fell up to tenfold, and the protective advantage of EVOH-34 over contrast increased from 1.35-fold to 4.1-fold. Conclusions Catheter-to-vessel geometry dominates reflux by orders of magnitude over viscosity; catheter selection is the primary modifiable safety determinant. EVOH polymerization provides geometrically selective secondary protection maximized by intermittent injection. The duty cycle δ is proposed as a standardizable parameter for reporting injection protocols.

Authors

Publication Details

Journal
CVIR Endovascular
Published
2026-05-20
DOI
https://doi.org/10.1186/s42155-026-00693-9
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Why catheter size matters more than viscosity in ethylene–vinyl alcohol copolymer reflux: an analytical study

Ruben Geevarghese, Arthur A. Cornelis, Stephen B. Solomon, Francois H. Cornelis
CVIR Endovascular
Central Venous Catheters and Hemodialysis
article

Why catheter size matters more than viscosity in ethylene–vinyl alcohol copolymer reflux: an analytical study

Ruben Geevarghese, Arthur A. Cornelis, Stephen B. Solomon, Francois H. Cornelis
article en

Abstract

Abstract Purpose Proximal reflux of liquid embolic agents remains a primary safety concern during endovascular embolization, yet the quantitative relationships governing reflux mechanics are poorly characterized. An analytical model was developed to determine the relative contributions of catheter geometry, agent viscosity, polymerization kinetics, and injection technique to reflux behavior of ethylene–vinyl alcohol copolymer (EVOH). Methods An annular Poiseuille flow model was constructed for a 1 mm vessel with coaxial microcatheters (1.3–2.8 French; catheter-to-vessel diameter ratio D* = 0.43–0.92). Carreau rheology was applied to EVOH-18 and EVOH-34; blood and iodinated contrast were modeled as Newtonian. Extensions incorporated vessel compliance, first-order EVOH polymerization kinetics, and intermittent injection parameterized by duty cycle δ. Results The annular geometry function φ(D*) spanned three orders of magnitude across the catheter range (0.179 to 0.0006), surpassing the maximum 2.4-fold inter-agent viscosity difference. At injection shear rates, EVOH-18 thinned to near-contrast viscosity (6.4 vs 6.1 mPa·s). EVOH polymerization reduced reflux up to 68% at low D* but was negligible at D* > 0.9. Under intermittent injection (δ = 0.20), effective reflux lengths fell up to tenfold, and the protective advantage of EVOH-34 over contrast increased from 1.35-fold to 4.1-fold. Conclusions Catheter-to-vessel geometry dominates reflux by orders of magnitude over viscosity; catheter selection is the primary modifiable safety determinant. EVOH polymerization provides geometrically selective secondary protection maximized by intermittent injection. The duty cycle δ is proposed as a standardizable parameter for reporting injection protocols.

CVIR EndovascularVol. 9(1)
Openalex Percentile: Top 3%
Central Venous Catheters and Hemodialysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Why catheter size matters more than viscosity in ethylene–vinyl alcohol copolymer reflux: an analytical study — Ruben Geevarghese, Arthur A. Cornelis, et al. · CVIR Endovascular (2026) | TGRS Research Map | TGRS