Two immobilization techniques in pediatric total marrow and lymphoid irradiation with helical Tomotherapy: Impact on setup errors and PTV margins

Aiming to compare the interfraction setup errors and derived planning target volume (PTV) margins of two immobilization techniques for pediatric total marrow (lymphoid) irradiation (TMI/TMLI), this study evaluated a whole-body vacuum cushion (VC) versus a vacuum cushion combined with a thermoplastic mask system (VC-TM). In this retrospective study, 12 pediatric patients undergoing TMI/TMLI with helical tomotherapy were stratified into VC and VC-TM groups. Daily megavoltage CT (MVCT)-based setup errors in left-right (LR), superior-inferior (SI), and anterior-posterior (AP) directions were analyzed. Planning target volume (PTV) margins were derived from the observed setup errors. In the VC group, the setup errors were 0.65 [0.30, 1.30] mm (LR), 4.60 ± 3.61 mm (SI), and 1.15 [0.75, 4.28] mm (AP). For the VC-TM group, the setup errors were 0.70 [0.40, 1.40] mm (LR), 2.08 ± 1.36 mm (SI), and 2.06 ± 1.33 mm (AP). With daily MVCT guidance, margins calculated from random errors alone were 2 mm (LR), 4 mm (SI), and 4 mm (AP) for VC group, and 2 mm (LR), 3 mm (SI), and 2 mm (AP) for VC-TM group. Compared with the VC group, the VC-TM group demonstrated superior setup accuracy both at the whole-body level and at local anatomic landmarks, particularly in the SI direction. Both immobilization techniques are feasible for pediatric TMI/TMLI. The VC-TM system provided more stable positioning, especially for the head and neck. Daily MVCT-guided online correction is recommended to utilize this stability and enable a substantial reduction in PTV margins.

Authors

Institutions

Publication Details

Journal
Journal of Radiation Research and Applied Sciences
Published
2026-05-19
DOI
https://doi.org/10.1016/j.jrras.2026.102459
Primary Topic
Hematopoietic Stem Cell Transplantation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Two immobilization techniques in pediatric total marrow and lymphoid irradiation with helical Tomotherapy: Impact on setup errors and PTV margins

Linan Song, 孙保锦, Qianwen Li, Xiaofeng Mu et al.
Journal of Radiation Research and Applied Sciences
Hematopoietic Stem Cell Transplantation
article

Two immobilization techniques in pediatric total marrow and lymphoid irradiation with helical Tomotherapy: Impact on setup errors and PTV margins

Linan Song, 孙保锦, Qianwen Li, Xiaofeng Mu, Fei Luo, Zhao Jing, Bo Liu, Jingbo Wang
article en

Abstract

Aiming to compare the interfraction setup errors and derived planning target volume (PTV) margins of two immobilization techniques for pediatric total marrow (lymphoid) irradiation (TMI/TMLI), this study evaluated a whole-body vacuum cushion (VC) versus a vacuum cushion combined with a thermoplastic mask system (VC-TM). In this retrospective study, 12 pediatric patients undergoing TMI/TMLI with helical tomotherapy were stratified into VC and VC-TM groups. Daily megavoltage CT (MVCT)-based setup errors in left-right (LR), superior-inferior (SI), and anterior-posterior (AP) directions were analyzed. Planning target volume (PTV) margins were derived from the observed setup errors. In the VC group, the setup errors were 0.65 [0.30, 1.30] mm (LR), 4.60 ± 3.61 mm (SI), and 1.15 [0.75, 4.28] mm (AP). For the VC-TM group, the setup errors were 0.70 [0.40, 1.40] mm (LR), 2.08 ± 1.36 mm (SI), and 2.06 ± 1.33 mm (AP). With daily MVCT guidance, margins calculated from random errors alone were 2 mm (LR), 4 mm (SI), and 4 mm (AP) for VC group, and 2 mm (LR), 3 mm (SI), and 2 mm (AP) for VC-TM group. Compared with the VC group, the VC-TM group demonstrated superior setup accuracy both at the whole-body level and at local anatomic landmarks, particularly in the SI direction. Both immobilization techniques are feasible for pediatric TMI/TMLI. The VC-TM system provided more stable positioning, especially for the head and neck. Daily MVCT-guided online correction is recommended to utilize this stability and enable a substantial reduction in PTV margins.

Journal of Radiation Research and Applied SciencesVol. 19(2)
Aerospace Center Hospital (CN), Beijing Shijitan Hospital (CN)
Zero hunger
Openalex Percentile: Top 6%
Hematopoietic Stem Cell Transplantation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.